Scanning electron microscopy
Tabletop Fei Quanta Hitachi
Scanning electron microscopy Fei Quanta Tabletop Hitachi Example: - - PowerPoint PPT Presentation
Scanning electron microscopy Fei Quanta Tabletop Hitachi Example: Tin soldier Pb M Sn L EDS analysis Average composition Secondary electrons Backscatter electrons Learning goals: Understanding the principle
Tabletop Fei Quanta Hitachi
Pb M Sn L Secondary electrons Backscatter electrons EDS analysis – Average composition
– Of the instrument – Of type of signals – detectors
change influences the picture and chemical analysis.
– Table top SEM – Environmental SEM – High resolution SEM
1930 1970 2018
Unaided eye ~ 0.1 mm Light microscope ~ 0.2 mm Scanning EM ~ 1.0 nm Transmission EM ~ 0.1 nm
The higher the accelerating voltage, the smaller the wavelength of the electrons and the higher the possible achievable resolution.
In SEM, there are several electromagnetic lenses, including condenser lenses and one objective lens. Electromagnetic lenses are for electron probe formation, not for image formation directly, as in
diameter of the electron beam. The objective lens further reduces the cross-section of the electron beam and focuses the electron beam as probe on the specimen surface.
Objective lens Cross section
The final probe-forming lens has to
distance (WD= 10 mm), that is the distance between the specimen and lower pole-piece. This is necessary so that the emitted radiation can be collected and detected with desired efficiency. The long working distance increases the spherical aberration of the probe- forming lens, which increases the size of the smallest attainable electron-beam spot. Final Lens
Electron beam SED Virtual lens Specimen
Out lens In lens Semi-in-lens ☆Easy to observe magnetic sample ☆Possible to observe bigger sample ☆Topographical imaging ☆Deep depth of Focus
☆Ultra high resolution ☆High throughput observation at sample exchange position ☆Variety of signal detecting system to optimize the contrast
Features Features Features
☆Ultra high resolution ☆ Possible to observe bigger sample at short WD ☆Variety of signal detecting system to optimize the contrast
SED Electron beam Virtual lens Specimen Specimen Virtual lens Electron beam SED SED SED
Compare the three different designs of objective lens : 1:Conventional lens Here the magnetic field of the lens is located inside the lens and the specimen sits away from the field. This leads to that the beam travels “unprotected” between lens and sample and is more vulnerable to EM disturbance. On the other hand the distance between lens and specimen only has limited effect on the signal strength. 2: Snorkel lens, semi-in-lens or immersion lens Here the magnetic field is projected down below the lens to enclose the specimen if it is located at short working distance. At the same time as the beam is protected from EM disturbance the electrons are effectively captured and led up through the lens to be detected by an in-lens detector. For longer working distance or when a side illumination effect is wanted, a classical SE detector is mounted in the chamber. 3: In-lens Here the sample sits on a TEM-type holder and has a maximum size of ca. 4x4x9 mm. Benefits here are excellent mechanical stability, high electron collection efficiency and very high EDS count rates.
> 5 years 18 months 6 months 1 month Life time 0.2 eV 0.5 eV 1.5 eV 2.0 eV Energy spread, E 1000 2x109 500 5x108 10 1x107 1 1x106 Brightness [A/cm2sr] Room temp 1500 C 1500 C 2300 C Temperature 3 - 5 nm 10 – 25 nm 1 – 2 mm 1 – 2 mm Source size Cold FE Schottky FE LaB6 W 10-11 Torr 10-9 Torr 10-7 Torr 10-5 Torr Gun vacuum 20-30 nA >100 nA 50 nA -1 mA 50 nA -1 mA Probecurrent 2-3 % 0.2 % 0.2 % 0.1 % Stability,%/h
to the linear size of the specimen area being scanned. The linear magnification is given by
Electronbeam is scanned across the specimen and the procedure is known as Raster scanning. Raster scanning causes the beam to sequentially cover a rectangular area on the specimen. The signal electrons emitted from the specimen are collected by the detector, amplified and used to reconstruct the image according to one-to-one correlation between scanning points on the specimen and picture points on the screen of cathode ray tube (CRT). CRT converts the electronic signals to a visual display.
The current in the condenser lens changes the spot size or diameter of the beam of electrons that scans the sample. More detailed information will be collected when the electron beam scans the same area with a smaller spot size. An increased current or a higher number for the condenser lens (CL) setting, will produce a smaller spot size and in general will result in a better resolution (A).
Diatome Courtesy of Gokhan Taken by Quanta SEM microscope Magnification: 30000x Sample: Diatome Detector: ETD Voltage: 4.0kV Vacuum: 4.01e-4Pa Horizontal Field Width: 9.93μm Working Distance: 7.7 Spot: 2.0 https://www.fei.com/image-gallery/
Same picture - different size
“Inelastic” scattering refers to a variety of physical processes that act to progressively reduce the energy of the beam electron by transferring that energy to the specimen atoms through interactions with tightly bound inner-shell atomic electrons and loosely bound valence electrons. Although the various inelastic scattering energy loss processes are discrete and independent, Bethe (1930) was able to summarize their collective effects into a “continuous energy where E is the beam energy (keV), Z is the atomic number, ρ is the density (g/cm3), A is the atomic weight (g/mol), and J is the “mean ionization potential” (keV) given by
Images are formed because of beam interaction with the sample This happens in a volume, not in a point The size of this volume varies with beam energy...
1μm Vacc : 10kV Vacc : 1kV
Sample = Si
The elastic scattering crossection, can be used to estimate how far the beam electron must travel on average to experience an elastic scattering event, a distance called the “mean free path,” λ Simultaneously with inelastic scattering, “elastic scattering” events
an atom (the positive nuclear charge as partially shielded by the negative charge of the atom’s orbital electrons), causing the beam electron to deviate from its previous path onto a new trajectory, as illustrated schematically in the figure.
The mean free path is of the order of nm. Elastic scattering is thus likely to occur hundreds to thousands of times along a Bethe range of several hundred to several thousand nanometers.
一 一 一 一 一 SE Surface information SE I SE Inner information Electron beam Z SE escape depth Low energy BSE (II) sample SE II 一 一 一 一 一 一 一 一 一 一 SE 一 一 一 High energy BSE (I) Z BSE escape depth
1 10 100 1000 10000 Energy of signal electron (eV) Electron amount Secondary electron Back scattered electron (SE) (BSE) 50eV
When a sample is hit by an electron beam a variety of electron emissions are available.
(Notice that the scale is logarithmic)
“Huskelapp” The group of secondary electrons (SE) can be divided into 4 groups. There are SEI, SEII, SEIII, and SEIV. These electrons are classed according to how they are generated. A SEI is an electron is an electron that is generated at the point of primary beam impingement in surface of the
A SEII is an electron that is generated when a backscattered electron leaves the surface of the specimen. Due to the energy of backscattered electrons, this SEII could leave the surface of the specimen microns away from the primary beam impingement site. SEIIs hurt the resolution of the image, but add greatly to overall image brightness. A SEIII is an electron released when an energetic backscattered electron strikes the interior of the specimen chamber, causing a SE to be released. SEIVs are formed when the primary beam strikes an aperature within the electron column. SEIIIs and SEIVs contribute noise to the image. By understanding signal formation, the specimen can be properly prepared for analysis.
electron beam range causes a large, characteristic variation in the SE yield
at 30keV to in excess of 1 at around 1keV, and as high as 100 for some materials
Experimental SE yield data for Ag
too deep to escape so the SE yield d is low
is so small that most of the SE generated can escape so the SE yield rises rapidly
produced because less energy is available so the SE yield falls again interaction volumes
low voltage
high voltage
compared to the high energy
28
Occurrence of edge effect in fine structured surfaces. Change in SE yield at different incident e-beam angles.
a volume 3-5nm in diameter, but the SE2 signal from a volume of 1-3µm in diameter. But at low energy the SE1 and SE2 electrons emerge from the same volume because of the reduction in the size of the interaction volume
limited by diffraction
500eV
30eV
100nm
一 一 一 一 一 SE Surface information SE I SE Inner information Electron beam Z SE escape depth Low energy BSE (II) sample SE II 一 一 一 一 一 一 一 一 一 一 SE 一 一 一 High energy BSE (I) Z BSE escape depth
Sample : Polyvinyl Alcohol Accelerating Voltage : 3kV, Vacuum: 60Pa,
Sample : Polyvinyl Alcohol Accelerating Voltage : 3kV, Vacuum: 60Pa,
BSE detector SE
33
The count ratio of BSE:s depends on the mean atomic number of the specimen.
SU8000Series standard optics
Top Upper Lower
STEM Detector
Signal type Signal name Detector information BSE HA- BSE T
Composition, crystal BSE LA- BSE Upper Composition + T
(Charge suppression) SE SE Upper Surface information (Including voltage contrast) SE Lower Lower T
STEM
BF- STEM
※1
STEM Sample internal information + Crystal STEM
DF- STEM
※1
Lower Sample internal information + Composition SE BSE STEM Electrode
V ariety of signal detection system in SU8200-series
37
Upper
Electrode
Sample EXB
SE BSE
Top
Topographical image with shadow
Lower
Sample : Photocatalyst Vacc : 3.0V Mag. : x 50k courtesy of : Nagaoka University of Technology, Faculty of Engineering,
Model :SU8020
Lower detector
LOWER(SE-L)
Upper
EXB
Lower Top
SE BSE UPPER(SE)
Sample : Photocatalyst Vacc : 3.0V Mag. : x 50k courtesy of : Nagaoka University of Technology, Faculty of Engineering,
Model :SU8020
Surface information (incl. Voltage contrast)
Control electrode Electrode
Upper detector: Pure SE
Sample
Upper
EXB
Lower Top
SE BSE
Sample : Photocatalyst Vacc : 3.0V Mag. : x 50k courtesy of : Nagaoka University of Technology, Faculty of Engineering,
Model :SU8020
Topographical + Compositional information
UPPER(LA-BSE) Electrode
Upper detector: SE Filtering, LA-BSE signal
Control electrode
Sample
Upper
EXB
Top Lower
SE BSE
Sample : Photocatalyst Vacc : 3.0V Mag. : x 50k courtesy of : Nagaoka University of Technology, Faculty of Engineering,
Model :SU8020
Compositional + Crystal information (Less topographical information)
TOP(HA-BSE) Electrode
Top detector: High-angle BSE
Control electrode
Sample
LOWER (SE-L) UPPER (SE) UPPER (LA-BSE) TOP (HA-BSE)
Vi = Vo - VR Objective lens
Specimen
Primary beam VR = 0 V Vi = 0.5kV VR = 1.5kV Vi = 0.5kV In deceleration, a negative voltage is applied to the specimen to decelerate primary electrons before arriving at specimen surface. Deceleration Landing voltage: Vi
Expands Optimum αi
Improved resolution
Normal mode
Top Upper
Accelerated BSE SE Accelerated SE
Vd 1 Both of the SE and BSE are accelerated by the deceleration voltage(negative bias) 2 Makes it more difficult to only select the low energy electrons 3 Sample bias and the accelerated BSE:s increase the signal and contrast
Top : Low energy signal Upper : High energy signal
But if we accelerate the SE:s ?
Backscattered Electrons (BSE) Cathodoluminescence Auger Electrons Characteristic X-rays Secondary Electrons (SE) Transmitted Electrons Absorbed Electrons Electron Beam Heat
We can change
We want: High resolution We have choosen low Accelerating voltage Ask questions in lab: We will repeat inlens detectors Next week
PbM Sn L
Transitions have different probabilities Lines have different intensities
1 keV = 1.602677·10-16 J
continuum or Bremsstrahlung (breaking radiation)
field of the atom core
energy
Observerte data fra SEM + EDS
shell electron replaces it
2 4 6 8 10 12 14 keV 1 2 3 4 5 cps/eV
C Si Cr Cr Mn Mn Fe Fe Ni Ni
If beam energy E > EK then a K-electron may be excited
Energy of emitted photon can be calculated:
EPhot = E1 – E2
e.g.: Fe L → K
E1 = EK = 7.11 keV E2 = EL = 0.71 keV EKa = 6.40 keV
X-ray energy is the difference between two energy levels !
0.71 keV 7.11 keV 6.40 keV
X-ray and AUGER generation process
Emission of Auger electron Emission of X-ray Auger and X-ray yield are competing processes
C Ge
characteristic X-rays (rest produce Auger electrons)
+ A = 1
0.005 for C K 0.5 for Ge K
L-familie
Kα Kβ
Fe K-familie
S (16) K1,2 2308 eV K 2464 eV (K - K1,2) 156 eV S (16) Ca (20) K1,2 2308 eV 3692 eV K 2464 eV 4013 eV (K - K1,2) 156 eV 319 eV S (16) Ca (20) Mn (26) K1,2 2308 eV 3692 eV 5900 eV K 2464 eV 4013 eV 6492 eV (K - K1,2) 156 eV 319 eV 592 eV
Mo (42) Ce (58) L1 2292 eV 4839 eV L 2394 eV 5262 eV Ll 2014 eV 4287 eV Mo (42) L1 2292 eV L 2394 eV Ll 2014 eV
Lα1 Lβ1 Lβ2 Lγ1 Ll Lγ2/3 Lα2
Spectrum Barium L series, 15 kV α1 : β1 : γ1 = 100 : 52 : 10
Spectrum Barium L-series, 15 kV α1 : β1 : γ1 = 100 : 52 : 10
Ti-Kα1
Ti-Kβ1
BaTiO3 Barium
Spectrum BaTiO3, 15 kV Overlapped Ba L-series and Ti K-series
Lα1 (100) Lβ1 (31) Lβ2 Lγ1 (5) Ll Lγ2/3 Lα2 Lα1 Lβ1 Lβ2 Lγ1 Ll Lγ2/3 Lα2
Ionisation cross section for electrons
Important for the selection
Fe Cr Ni
Cr: 33% Fe: 33% Ni: 33% U = 10 keV
:
bind exc
E E
Cr Fe Ni 1,847 1,561 1,337
A sample with equal amount of Cr,Fe and Ni.
Fe/Cr=0,8 Ni/Cr=0,7 Fe/Cr=0,5 Ni/Cr=0,15
where ns is the number of electrons in the shell or subshell (e.g., nK = 2), and bs and cs are constants for a given shell (e.g.,bK = 0.35 and cK = 1) E= beam energy Ec= ionization energy
http://www.med.harvard.edu/jpnm/physics/refs/xrayemis.html
Different excitation ranges for:
Bremsstrahlung,
sample surface electron beam (E0) secondary electrons
back-scattered electrons bremsstrahlung X-rays
MEF4010 Scanning electron microscopy
WDX wavelength dispersive EDX Energy dispersive Energy resolution is 100 times better with the WDX. 125-150 eV EDX 10 eV WDX
function of primary beam energy With higher primary electron energy penetration depth is increasing
Rd ≈ 2,5 µm Rd ≈ 1,3 µm Rd ≈ 0,4 µm
EHT = 10 kV EHT = 20 kV EHT = 30 kV
http://www.gel.usherbrooke.ca/casino/What.html
atomic number (EHT = 15 kV) With higher density penetration depth is decreasing
Carbon Rd ≈ 2 µm Iron Rd ≈ 0,6 µm Gold Rd ≈ 0,2 µm
Ichar increases with increasing E0/Ec X-ray signal improves Rx increases with increasing E0/Ec X-ray spatial resolution degrades
C
E U 2... 2,5 E
Optimum overvoltage
Without sufficient overvoltage, x-ray production is dramatically lowered.
QE (cm2 keV2)
U
U = Eo/Ec
Optimum overvoltage U is around 2-2.5 The ionization cross-section describes the probability that a particular event will take place:
For X-rays: Q = 6.5x10-20 nsbs UEc ln(csU)
ns = number of electrons in a shell Ec = critical excitation voltage bs & cs = constants related to the electron shell U = overvoltage (Eo/Ec)
Can't calculate ZAFs unless concentration is known. But we don’t know the concentration?? Use k values (I/I° = k) to estimate compositions of each element. Then calculate ZAFs, and refine by iteration Z – at. no. correction Function of electron backscattering factor & electron stopping power - depend upon the average atomic number of unknown and standard Dependent on ionization cross section Varies with composition and accelerating voltage A – absorption correction Varies with m, takeoff angle, accelerating voltage F – fluorescence correction primary fluorescent x-rays ––> secondary fluorescent x-rays Varies with composition and accelerating voltage
http://www.probesoftware.com/Technical.html
http://www.probesoftware.com/download/PROBEWIN.pdf
kV=15 vinkel = 40
Mass absorption coefficients Mass absorption coefficients are stored as a matrix of numbers of absorption of a particular X-ray line (the emitter) by an absorber: For example, a portion of the MAC matrix for Kα X-rays for Z = 23 to 29 is shown below. http://www.ammrf.org.au/myscope/analysis/e ds/quantitative/
Mass absorption coefficients Emitter V 4952 eV Cr 5415 eV Mn 5899 eV Fe 6403 eV Co 6930 eV Ni 7478 eV Cu 8048 eV V 94.6 73.8 498.4 403.4 328.2 268.5 220.8 Cr 111.1 86.7 68.4 454.8 370.8 303.9 250.4 Mn 125 97.6 76.9 61.3 401.9 330.1 272.4 Fe 145 113.2 89.3 71.1 57.1 370.2 306 Co 160.9 125.7 99.1 79 63.5 51.4 329.4 Ni 187.9 146.8 115.8 92.3 74.1 60 49 Cu 200.7 156.8 123.8 98.7 79.3 64.2 52.4
Note that the MAC for absorption of Fe Kα by Co (79.0) is different from the MAC for absorption of Co Kα by Fe (57.1).
rough surface polished surface tilted polished surface X-rays electron beam take-off angle negative tilt angle
SDD d1 d2 d1: distance to (imaginary) polished surface d2: actual distance to specimen surface