Instrumentation (and
Process Control)
Fall 1393 Bonab University
Instrumentation - Course Information Department of Electrical - - PowerPoint PPT Presentation
Instrumentation (and Fundamentals & Process Control) Characteristics Fall 1393 Bonab University Instrumentation - Course Information Department of Electrical Engineering Instructor: Fariborz Rahimi Lectures: 13-14:30pm (Sat), 8-10am
Fall 1393 Bonab University
Department of Electrical Engineering
Instructor: Fariborz Rahimi Lectures: 13-14:30pm (Sat), 8-10am (Sun*) Prerequisites/Useful courses: Signals-Systems, Electrical measurements, Linear control systems Evaluation:
20%
References:
2
References:
3
instrumentation & control
systems
proportional/integral/derivative (PID) control laws, PLC controllers (or Micro-controllers)
4
5
Automation field
Sensor / Transducer Signal Conditioning Amplifier Recorder Data Analysis Controller Control Command Power Supply
Introduction
Parameter adjustment)
automatic control
temperature, pressure and flow)
Controller, Interface
6
Introduction
Refining, combining, handling, manipulating inputs (raw) to profitably produce end products is:
Small changes in a process = a large impact on the end result. Variations in:
must be carefully and consistently controlled to produce the desired end product with a minimum of raw materials/energy Process control more precise operations =profitability, quality & safety
7
Introduction
8
Measurement & Sensors
Errors in measurement:
Error Sources: 1. Gross errors (هدمع یاطخ)
2. Systematic errors (دنمشور)
9
Measurem ent System Input: True value of a variable
Output: Measured value of a variable
Pressure, speed, flow,…
May
Measurement & Sensors
2. Systematic errors
external electrostatic or magnetic field)
3. Random/Residual errors (یفداصت)
due to a multitude of small factors which change or fluctuate from one measurement to another. The happenings
Error Categories:
10
Measurement & Sensors
standards of known accuracy
1. Sensitivity 2. Accuracy (= تحص)
Closeness to the true value
Identifyability, clearity, freedom from Random error
11
𝑓𝑑 = 𝑓𝑢2 + 𝑓𝑡𝑑2 + 𝑓𝑏2 + 𝑓𝑆2 Input quantity Qi Output quantity Qo Qi
R Zero shift Allowable deviation
𝑇 = lim 𝑅𝑗 → 0 ∆𝑅𝑝 ∆𝑅𝑗
Measurement & Sensors
(Root Mean Square, RMS)
measured
12
Measurement & Sensors
external power source
13
Measurement & Sensors
downward force balances the fluid pressure. (Pressure measurement in terms of weight)
more accurate than deflection types
14
Measurement & Sensors
(output can have an infinite number of values within the range, the deflection- type of pressure gauge)
values (needed for Microprocessor/computer)
15
Measurement & Sensors
physical quantity measured
magnitude is proportional to the measured quantity (commonly as part of automatic control systems)
16
Measurement & Sensors
between those that do incorporate a microprocessor (smart/intelligent) and those that don’t
17
Measurement & Sensors
±1.0% f.s. (±1% of full-scale reading).
pressure gauge is measuring a pressure of 1 bar?
the instrument. Thus, when measuring a pressure of 1 bar, the maximum possible error of 0.1 bar is 10% of the measurement value
18
Measurement & Sensors
instrument spread of readings = very small
19
Measurement & Sensors
robots programmed to place components at a particular point on a table
accuracy of some instruments is sometimes quoted as a tolerance value)
Example: A packet of resistors gives the nominal resistance value as 1000 Ohm and the manufacturing tolerance as ±5%. If one resistor is chosen at random from the packet, what is the minimum and maximum resistance value that this particular resistor is likely to have? Solution: 950 & 1050 Ohm
20
Measurement & Sensors
the quantity being measured
readings marked X from this straight line
full-scale reading.
21
Measurement & Sensors
measured changes by a given amount
measurement sensitivity of the instrument in ohms/oC Resistance (V) Temperature (oC) 307 200 314 230 321 260 328 290 Solution: 7/30 = 0.233 Ohm/oC
22
Measurement & Sensors
before the change in the instrument output reading is of a large enough magnitude to be detectable
magnitude of the change in the input measured quantity that produces an observable change
Resolution = 5km/h
23
Measurement & Sensors
conditions of temperature, pressure, …
disturbance is a measure of the magnitude of this change affect instrument:
a constant error (bathroom scale)
(a thumbwheel for the scale)
sensitivity of measurement varies as ambient conditions change
24
Measurement & Sensors
The following table shows output measurements of a voltmeter under two sets of conditions: (a) Use in an environment kept at 20oC which is the temperature that it was calibrated at (b) Use in an environment at a temperature of 50oC
Voltage readings at calibration temperature
10.2 10.5 20.3 20.6 30.7 40.0 40.8 50.1
Determine the zero drift when it is used in the 50oC environment, assuming that the measurement values when it was used in the 20oC environment are correct. Also calculate the zero drift coefficient.
25
Measurement & Sensors
Voltage readings at temperature of 50oC
Zero drift at the temperature of 50oC is the constant difference between the pairs of output readings, that is, 0.3 volts.
temperature change causing the drift (30oC). Thus the zero drift coefficient is 0.3/30 = 0.01 volts/oC.
deflection/load characteristic: Load (kg) 0 1 2 3 Deflection (mm) 0 20 40 60
Load (kg) 0 1 2 3 Deflection (mm) 5 27 49 71
26
Measurement & Sensors
At 20oC, deflection/load characteristic is a straight line. Sensitivity = 20 mm/kg. At 30oC, deflection/load characteristic is still a straight line. Sensitivity = 22 mm/kg. Zero drift (bias) = 5 mm (the no-load deflection) Sensitivity drift = 2 mm/kg Zero drift/oC = 5/10 ¼ 0.5 mm/oC Sensitivity drift/oC = 2/10 = 0.2 (mm/kg)/oC
27
Measurement & Sensors
1. the input measured quantity to the instrument is increased steadily from a negative value Output: curve-A 2. the input variable is then decreased steadily Output: curve-B
these loading and unloading curves
friction (mechanical flyball), magnetic hyst. in iron cores (LVDT-RVDT)
28
Measurement & Sensors
there is no change in output value
29
Measurement & Sensors
A typical cause of dead space: Backlash in gears Example: converting between translational and rotational motion
instrument settles down to (e.g. accuracy)
measured quantity changes value and the time when the instrument output attains a steady value in response
conditions: variation in Dyn. Char.
30
Measurement & Sensors
the instrument output moves immediately to a new value
31
Measurement & Sensors
taken for the output quantity q0 to reach 63% of final
32
Measurement & Sensors
A balloon:
(a) balloon released time = zero , upward velocity = 5 meters/second
draw a table showing the temperature and altitude measurements (intervals of 10s) 50s Show also in the table the error in each temperature reading
(b) What temperature does the balloon report at an altitude of 5000 meters?
33
Measurement & Sensors
a) temperature reported at general time t be Tr
x=5t
r=10 C= -0.75
b) 5000m, t=1000
34
Measurement & Sensors
_ _ _
35
Measurement & Sensors
Standard format 2nd order instrument
A. No-damping = oscillation B. Light damping = oscillatory C. Critically damped (the most practical) reduces osc. and overshoot D. Damped E. Over-damped
36
Measurement & Sensors