basic prinicpals of high
play

BASIC PRINICPALS OF HIGH PERFORMANCE LIQUID CHROMATOGRAPHY Violeta - PowerPoint PPT Presentation

BASIC PRINICPALS OF HIGH PERFORMANCE LIQUID CHROMATOGRAPHY Violeta Ivanova-Petropulos Faculty of Agriculture, University Goce Delev tip, R. Macedonia What is liquid chromatography? Liquid chromatography (LC) is an analytical


  1. BASIC PRINICPALS OF HIGH PERFORMANCE LIQUID CHROMATOGRAPHY Violeta Ivanova-Petropulos Faculty of Agriculture, University “Goce Delčev” – Štip, R. Macedonia

  2. What is liquid chromatography?  Liquid chromatography (LC) is an analytical technique based on the separation of molecules due to differences in their structure and/or composition.  Liquid chromatography was defined in the early 1900s by Mikhail S. Tswett. - Separation of compounds (leaf pigments) extracted from plants using a solvent, in a column packed with particles. Page  2

  3. Tswett's Experiment Ether Chromato graphy Colors Plant extract CaCO 3 Page  3

  4. Chromatographic methods are applied for: - SEPARATION OF COMPOUNDS in a mixture - Identification and determination - QUALITATIVE ANALYSIS (retention time, UV-Vis spectra, MS spectra) - QUANTITAVIE ANALYSIS (peak area or peak height) - Separation is performed between two phases, mobile and stationary . - Compounds which are longer retained at the stationary phase will elute later, compared to those which are distributed into the mobile phase. Page  4

  5. Chromatography Types Mobile phase Gas Liquid Solid Gas Stationary Liquid phase Liquid Gas chromatography chromatography Solid Page  5

  6. High performance liquid chromatography (HPLC) system  HPLC is a form of liquid chromatography used to separate compounds that are dissolved in solution.  HPLC instruments consist of a reservoir of mobile phase, a pump, an injector, a separation column, detector and data processor. Detector Column Column oven Pump (thermostatic column chamber) Eluent Sample injection unit Drain (mobile phase) (injector) Data processor Degasser Page  6

  7. HPLC instruments Page  7

  8. HPLC columns  The column is the “ core” of any chromatographic system  One of the most important components where the separation of the sample components is achieved  Columns are commercially available in different lengths, bore sizes and packing materials. Page  8

  9.  The most widely used packing materials for HPLC separations are silica-based.  The most popular material is octadecyl-silica (ODS-silica), which contains C18 coating - materials with C1, C2, C4, C6, C8 and C22 coatings  The column life can be prolonged with proper maintenance: - flushing a column with mobile phase of high elution strength following sample runs is essential. - When a column is not in use, it should be capped to prevent it from drying out. - Particulate samples need to be filtered and when possible a guard column should be utilized. Page  9

  10. Column types  Normal phase  Reverse phase  Size exclusion  Ion exchange Page  10

  11. Normal phase  Stationary phase: High polarity - Silica or organic moieties with cyano and amino functional groups  Mobile phase: Low polarity – Hexan or heptan Page  11

  12. Reverse phase  Stationary phase: Low polarity – Octadecyl group-bonded silical gel (ODS)  Mobile phase: High polarity – Water, methanol, acetonitrile – Salt or acid is sometimes added.  Typical stationary phases are nonpolar hydrocarbons (such as C18, C8, etc.) and the solvents are polar aqueous-organic mixtures such as methanol-water or acetonitrile-water. CH 2 CH 2 CH 2 CH 2 CH 2 CH 2 CH 2 CH 2 CH 2 Si -O-Si CH 2 CH 2 CH 2 CH 2 CH 2 CH 2 CH 2 CH 2 CH 3 C 18 (ODS) Page  12

  13. Reverse phase Page  13

  14. Normal Phase/Reversed Phase Type Stationary phase Mobile phase Normal High polarity Low polarity phase (hydrophilic) (hydrophobic) Reversed Low polarity High polarity phase (hydrophobic) (hydrophilic) • The polarities of stationary phase and mobile phase have to be different! Page  14

  15. Elution  Isocratic  Constant eluent composition, same eluent: for example 50 % methanol  Gradient – Varying eluent composition • HPGE (High Pressure Gradient): High gradient accuracy, complex system configuration (multiple pumps required) • LPGE (Low Pressure Gradient): Simple system configuration, degasser required Page  15

  16.  In isocratic mode CH 3 OH/H 2 O = 6/4 Long analysis time!! Poor CH 3 OH/H 2 O = 8/2 separation!! (Column: ODS type) Page  16

  17.  In gradient mode Concentration of methanol in eluent 95% 30% Page  17

  18. Detector requirements  Sensitivity – The detector must have the appropriate level of sensitivity.  Selectivity – The detector must be able to detect the target substance without, if possible, detecting other substances.  Adaptability to separation conditions  Operability , etc. Page  18

  19. Types of Detectors  UV-Vis absorbance detector  Photodiode array-type UV-VIS absorbance detector (DAD)  Fluorescence detector  Refractive index detector  Electrical conductivity detector  Electrochemical detector  Mass spectrometer Page  19

  20. - UV-Vis detector has only one sample-side light-receiving section - DAD has multiple (1024 for L-2455/2455U) photodiode arrays to obtain information over a wide range of wavelengths at one time Page  20

  21. UV-Vis spectra of anthocyanin monoglucosides 528.0 0.20 Mv-Glc UV max = 528.0 nm Dp-Glc UV max = 525.6 nm 0.18 Cy-Glc UV max = 520.7 nm 0.16 Pt-Glc UV max = 525.6 nm Pn-Glc UV max = 515.9 nm 0.14 276.5 0.12 AU 0.10 243.4 0.08 0.06 0.04 525.6 515.9 348.0 276.5 0.02 357.4 290.8 345.7 0.00 250.00 300.00 350.00 400.00 450.00 500.00 550.00 nm Page  21

  22. UV-Vis spectra of vitisin A and vitisin B Page  22

  23. Fluorescence detector  The most sensitive among the existing modern HPLC detectors.  Typically, fluorescence sensitivity is 10 -1000 times higher than that of the UV detectors  Fluorescence detectors are very specific and selective among the others optical detectors.  Roughly about 15% of all compounds have a natural fluorescence - derivatization is necessary Page  23

  24. Refractive index detector  Measures the refractive index of an analyte relative to the solvent  They can detect anything with a refractive index different from the solvent, but they have low sensitivity  Very sensitive to slight changesd of the mobile phase, not compatible for gradient elution Page  24

  25. Mass spectrometer  Mass spectrometry (MS) is an analytical technique that ionizes chemical species and sorts the ions based on their mass to charge ratio.  Mass spectrum measures the masses within a sample.  Mass spectrometry is used in many different fields and is applied to pure samples as well as complex mixtures.  Used for: • characterization of complex structures of compounds • detection of new compounds in different matrices • ……… Page  25

  26. UV and visible chromatograms of Extracted ion chromatograms at different m/z polyphenols: (a) 280 nm, (b) 320 nm, (c) 360 values, which correspond to the M + signals of nm, (d) 520 nm the anthocyanins Intens Intens. x10 8 . (a) (a) [mAU] 4 5 150 100 2 3 1 1.0 5 50 3 4 0 4 200 (b) 1 2 150 0.0 x10 7 (b) 6 100 5’ 4 50 0 3 4 11 (c) 2 9 40 3’ 4’ 13 7 10 1 8 1’ 2’ 20 12 0 x10 7 0 (c) 5’’ 0 (d) Anthocyanin- 3 40 monoglucosides 30 Anthocyanin- acetylglucosides 2 4’’ Anthocyanin- 20 p -coumaroylglucosides 3’’ 10 1 1’’ 2’' 0 0 10 20 30 40 50 Time [min] 0 0 10 20 30 40 50 Time [min] Page  26

  27. Mass spectrum of catechin ( m/z 291) obtained under positive mode Page  27

  28. Mass spectrum of procyanidin ( m/z 577) obtained under negative mode m/z 577 559, 451, 425, 289 425 [ M-H] - = 577 100 [M-H-152] - 95 Dimer 90 85 152 OH 80 75 OH 70 65 O HO Intensity 60 -H 2 O OH 55 OH 126 50 OH [M-H-170] - OH 45 407 40 289 O OH 35 30 OH 25 451 [M-H-126] - 20 OH 289 15 559 [M-H-H 2 O] - 10 5 0 Page  28 200 300 400 500 600 700 800 900 1000 1100 m/z

  29. Quantitative analysis  Quantitation performed with peak area or height.  Calibration curve created beforehand using a standard. – External standard method – Internal standard method – Standard addition method Page  29

  30. External standard method  The simplest method  The accuracy of this method is dependent on the reproducibility of the injection volume.  Standard solutions of known concentrations of the compound of interest are prepared with one standard that is similar in concentration to the unknown.  A fixed amount of sample is injected.  Peak height or area is then plotted versus the concentration for each compound. The plot should be linear and go through the origin.  The concentration of the unknown is then determined according to the following formula: Area unknown Conc. unknown = conc. known Area known Page  30

Download Presentation
Download Policy: The content available on the website is offered to you 'AS IS' for your personal information and use only. It cannot be commercialized, licensed, or distributed on other websites without prior consent from the author. To download a presentation, simply click this link. If you encounter any difficulties during the download process, it's possible that the publisher has removed the file from their server.

Recommend


More recommend