Methods of Photosynthesis Spectrometry Christophe Six Dfinitions - - PowerPoint PPT Presentation
Methods of Photosynthesis Spectrometry Christophe Six Dfinitions - - PowerPoint PPT Presentation
Master 2 me anne : Sciences et Technologies Mention : Sciences de lUnivers, Environnement, Ecologie Spcialit : Ocanographie et Environnement Marin Spcialit : Ocanographie et Environnement Marin Unit dEnseignement : Evolution
Définitions
Spectrometry = spectroscopy : Methods of spectral analysis allowing to understand the composition the structure of matter and/or the study the composition, the structure of matter and/or the study
- f systems transferring energy
Qualitative and quantitative studies of spectra derived from the interaction Qualitative and quantitative studies of spectra derived from the interaction between the matter and the wavy radiations of different frequences . Spectrophotometry is an analytic, quantitative method that consists in measuring the absorbance (= absorption = optical density) of a given chemical substance (or of a whole unicell organism) in solution, function of the light wavelength. (or of a whole unicell organism) in solution, function of the light wavelength. Spectrofluorimetry is an analytic, quantitative method that consists in measuring p y y , q g the emission and excitation levels of fluorescence of a given chemical substance (or of a whole unicell organism) in solution, function of the light wavelength.
Energy and wavelength
E = (h . c) /λ
E : Photon energy
800 nm
E : Photon energy h : Plank constant factor c : Light celerity l : Photon wavelength
Wavelength in nanometers 400 nm
X-rays U.V. Visible Infrared Radio wavelengths
Absorbance of molecules and molecular complexes
.Understanding photophysics and photobiology Very useful for assays .Very useful for assays
Using colorimetric assays Using colorimetric assays
Concept of absorbance and measurement
Sample Photomultiplicator Light Source
I0 I
>
Absorbance Transmittance
A = log (I0 / I) T = I / I0 A = -log T
Spectrophotometers
Components : .One or several light source(s)
Extended Visible (350-900 nm) : Tungsten, Halogen
Components :
( ) g g UV (<400 nm) : Deuterium
. One monochromator : Selection of wavelengths . One sample compartment . One detector : photomultiplicator or photodiode detector . A result display system
Single beam spectrophotometers
- r monochromator
D.O.
- r monochromator
λ (
)
400 500
λ (nm)
400 500
Single beam spectrophotometers
. A simple compartment for a single sample cuvette . The simplest system . The reference = blank is measured before the samples for zeroing the device
Blank : all chemical components (buffer, solvant, etc) except the absorbing substance that you want to measure. It is actually rare to be able to use a perfect true blank but one should approach it as much as possible. approach it as much as possible.
. Instrument useful for simple routine applications (single or few wavelengths)
V i l i t i ( t i l i id i t t ) Various colorimetric assays (proteins, nucleic acids, pigments, etc.)
. Main problems
The decrease of lamp intensity is not compensed In single wavelength mode, one cannot check for artefacts
I0 I
(fix) (measured) The making of these instruments is usually less careful
Double beam spectrophotometers
M h t
Reference Cuvette
Monochromator
I0
Chopper Sample Cuvette Chopper Cuvette
I
Double beam spectrophotometers
. For each wavelength, one mesures the absorbance of the sample AND the absorbance of the reference (blank) . Good reliability of the measurements, ideal for absorption spectra
(Eli i ti f l t b ti ) (Elimination of solvent absorption)
. Correction of the variations of the light sources
Artefacts
Refraction : déviation of a wave when its speed changes (interface between 2 media)
Diopter (surface of the cuvette and surface of the sample)
=> Aλ
. Other optical phenomenons linked to diffusion, reflexion and diffraction
- f light may also distort the measurement.
. Other optical phenomenons linked to diffusion, reflexion and diffraction
- f light may also distort the measurement.
Artefacts : Light diffusion
Turbid solutions cell suspensions . Turbid solutions, cell suspensions
=> Aλ
Diffusion occurs when some light is deflected by particules and therefore does not reach the detector
σS =
F( d, n)
λ4
=
Diffusion of Rayleigh
λ
d : Diameter of particules n : Refraction index
λ : Wavelength Diffusion also depends on :
- Particule concentration
- Particule shape
Impact of diffusion on absorption spectra
4,0E-11 5,0E-11
1
1 0E-11 2,0E-11 3,0E-11
y = 1 x4 Diffusion is λ-dependent
0,0E+00 1,0E-11
350 450 550 650 750 850 Longueur d'onde (nm) g ( )
=> Aλ ok => Aλ
Impact of diffusion on absorption spectra
Example : absorption spectrum of a phycoerythrin I
Spectrum with diffusion Fitting a correction curve Final spectrum
ce sorbanc Abs Wavelength (nm)
Measuring absorbance in a diffusing sample
=> Aλ
Bringing the detector Bringing the detector nearer to the cuvette Increasing the surface g
- f the detector
Measuring absorbance in a diffusing sample
Echantillon homogène Détecteur du photomultiplicateur
Homogeneous sample Light detector
DO (nm) DO (nm)
A
(nm) (nm) DO DO
B
Suspension de cellules
(nm) (nm)
Cell suspension
Source Rayon lumineux
DO (nm)
C
Light beam Light source and
lumineuse et monochromateur Sphère d’intégration
Integration sphere Light source and monochromator
If the absorbance of a sample is not stable…
. Sample much colder than the atmosphere of the compartment
Condensation on the cuvette Condensation on the cuvette Gaz formation (diffusion)
Sample drops on the outside of the cuvette . Sample drops on the outside of the cuvette . The sample contains absorbing particules that sink in the cuvette . There’s not enough sample in the cuvette and the beam passes through the meniscus . Cuvettes not adapted (micro-cuvettes)
The Beer-Lambert law
At a given wavelength the absorbance of a solution is proportional to the At a given wavelength, the absorbance of a solution is proportional to the concentration of the absorbing chemical species that are present in this solution, and to the optical path A
Aλ = ελ . l . C
A : Absorbance (no unit) λ : Wavelength (nm) l : Optical path (cm) C : C
t ti ( l L 1)
λ λ
C : Concentration (mol L-1)
ελ : Extinction coefficient (L mol-1 cm-1)
. The Beer-Lambert law is additive. Pour n chemical species :
A = ε l C + ε l C + ε l C + + ε l C Aλ = ελ,1 . l . C1 + ελ,2 . l . C2 + ελ3 . l . C3 + … + ελ,n . l . Cn
. For l = 1 cm :
Aλ = ελ . C => C = A / ελ A = ε C + ε C + ε C + + ε C Aλ = ελ,1 . C1 + ελ,2 . C2 + ελ3 . C3 + … + ελ,n . Cn
Fluorescence: what is it ?
Stokes shift
Intensity of fluorescence emission
. With fluorescence, there’s no general relation such as the absorbance Beer-lambert law
The measurement depends strongly on :
- The nature of the fluorescent system that is studied
y
- The device used to quantify fluorescence (light source intensity, optics configuration, etc.)
Need to use standard curves to quantify molecules by fluorescence Need to use standard curves to quantify molecules by fluorescence
. It is possible to quantify the fluorescence energy when a fluorescence quantum yield Qf :
Energy of fluorescence emitted (If) = Absorbed energy (Ia) x Qf Qf = f (λ, T°C, pH, ions, etc.)
Spectrofluorimeters
. None photon from the excitation light must be detected by the detector excitation at 90°
On average there is 106 times less photons that hit the detector of a spectrofluorimeter On average, there is 10 times less photons that hit the detector of a spectrofluorimeter than in a spectrophotometer
- A light source : Mercury or xenon lamp
Main components : A light source : Mercury or xenon lamp
- Two monochromators selecting either the emission or excitation precise wavelengths
- A dark compartment with the cuvette in a 90° excitation/emission cuvette holder
- A photomultiplicator
Diagrammic representation of a spectrofluorimeter
Photomultiplicator Xenon lamp Lens Entrance Slit Exit slit Photomultiplicator Monochromator shutter Monochromator Monochromator Slit Mirror Lens Lens Sample
Emission and Excitation spectra of fluorescence
Monochromator scanning
Emission spectrum
Fix monochromator : Monochromator scanning all wavelengths Quantification of the fluorescence emitted ∼ 15 nm Fix monochromator : One given λ Emission fluorescence emitted by the excitation of a given λ Excitation Sample At which λ is the maximum of fluorescence emission of the compound ? 600 500 700 400 p Fix monochromator : One given λ
Excitation spectrum
Emission Monochromateur scanning all wavelengths Quantification of the fluorescence emitted many wavelengths λ Excitation Sample 600 500 700 400 600 500 700 400 Which λ gives rise to the fluorescence emission at a given λ ? (Excitation spectra are often similar to absorption spectra)
Fluorescence of marine picocyanobacteria : Synechococcus spp.
Marine phycoerythrins & spectrofluorimetry
. There are several types of phycoerythrins (PE) Excitation In the Emission Variable Excitation Emission In the blue-green region, at 500nm between 560-580 nm depending on the between 400 and 550 nm Emission at 580 nm
(for instance)
Excitation spectrum Emission spectrum
(for instance)
type of PE
One or two major maxima
495 545 600 500 700 400 600 500 700 400
Phycoerythrin structure and excitation spectra
Phycobiliprotein = Apoprotein + pigment Phycobiliprotein = Apoprotein + pigment Pigment = chromophore phycobilin One or two types of phycobilin are bound to marine phycoerythrins
Excitation spectrum
545
One or two major maxima
495 600 500 700 400