Ghost Peaks: How to Fix a Haunting Problem Jacob A. Rebholz - - PowerPoint PPT Presentation

ghost peaks
SMART_READER_LITE
LIVE PREVIEW

Ghost Peaks: How to Fix a Haunting Problem Jacob A. Rebholz - - PowerPoint PPT Presentation

Ghost Peaks: How to Fix a Haunting Problem Jacob A. Rebholz Teledyne Tekmar VOC Product Line Manager Outline of Topics What is a ghost peak? How to identify ghost peaks and their source How to correct various sources of


slide-1
SLIDE 1

Ghost Peaks:

How to Fix a Haunting Problem

Jacob A. Rebholz Teledyne Tekmar VOC Product Line Manager

slide-2
SLIDE 2

Outline of Topics ■ What is a “ghost” peak? ■ How to identify ghost peaks and their source ■ How to correct various sources of contamination ■ How to prevent future “hauntings”!

2

slide-3
SLIDE 3

What is a “Ghost” Peak? ■ Simply put: any peak that does not belong in your chromatogram.

3

slide-4
SLIDE 4

What is a “Ghost” Peak? ■ Simply put: any peak that does not belong in your chromatogram.

4

slide-5
SLIDE 5

What is a “Ghost” Peak? ■ Less simply put: contamination can present in several ways

■ Elevated baseline ■ Contribution to target peaks ■ Tailing/ “splitting”

5

slide-6
SLIDE 6

What is a “Ghost” Peak?

6

■ Less simply put: contamination can present in several ways

■ Elevated baseline ■ Contribution to target peaks ■ Tailing/ “splitting”

slide-7
SLIDE 7

What is a “Ghost” Peak? ■ Less simply put: contamination can present in several ways

■ Elevated baseline ■ Contribution to target peaks ■ Tailing/ “splitting”

7

Contamination Contribution Clean Peak

slide-8
SLIDE 8

What is a “Ghost” Peak? ■ Less simply put: contamination can present in several ways

■ Elevated baseline ■ Contribution to target peaks ■ Tailing/ “splitting”

8

slide-9
SLIDE 9

How to Identify a “Ghost” Peak

9

■ Visually detect stray peaks in chromatogram

■ Helps to have an example of a clean chromatogram, or old chromatograms from when the system was clean ■ Helps identify extra peaks, elevated baselines and potentially tailing/split peaks

■ Poor calibration results

■ Can help ID contribution to peaks and split/tailing peaks

slide-10
SLIDE 10

Troubleshooting Tips ■ Take a moment to review manuals ■ Examine the facts and use valid reasoning ■ Identify the “root cause” of the problem ■ Avoid quick fixes and shotgun approach

BE PATIENT!

10

slide-11
SLIDE 11

Exorcising the Demon ■ Step 1: locate the source of the contamination

■ Break system down into major components:

■ Autosampler components, Detector, GC inlet, GC column

11

slide-12
SLIDE 12

A Word on Method Parameters ■ GC cycle time

■ Heavy compounds missing from the end of chromatogram ■ Show up in following blank ■ Fix: increase final temperature and/or hold time

12

slide-13
SLIDE 13

A Word on Method Parameters ■ Inlet temperature

■ Typically heavy compound carryover, potentially others ■ Check temperature of inlet and/or any transfer lines ■ Fix: ensure temperature is adequate to prevent condensation

  • f compounds, increase if necessary

13

slide-14
SLIDE 14

Exorcising the Demon ■ System blanks

■ Run a full system blank ■ Run GC only blank (no autosampler) ■ Run Detector only Blank (no column connected if possible)

14

slide-15
SLIDE 15

Solvents/Standards ■ Solvent

■ Methanol, acetone, etc. ■ Direct inject of solvent used to prepare standards ■ Compare results to a second source of solvent ■ Don’t forget that syringe could contribute ■ Second GC system to confirm

15

slide-16
SLIDE 16

Solvents/Standards ■ Solvent ruled out ■ Standards

■ Crack new ampules of internal and surrogate standards ■ Direct inject ■ Crack new ampules of calibration standard ■ Direct inject ■ Do not forget syringes as source of contamination

16

slide-17
SLIDE 17

Carrier Gas ■ Gas filters ■ Gas supply tubing ■ Gas regulators ■ Carrier gas supply

17

http://www.restek.com/catalog/view/3445/22019

slide-18
SLIDE 18

Carrier Gas ■ Allow GC oven to sit at or below starting temperature for up to several hours

■ Run GC blank (no injection) ■ If carrier gas contamination is focusing on cold column, peaks will be larger ■ If no change in peak size, continue with trouble shooting

18

slide-19
SLIDE 19

Carrier Gas

■ Verify filter is not expired (color change indicator)

■ Check for contamination ■ Replace if uncertain

■ Gas supply

■ Tank tubing and regulator ■ Replace all, verify issue resolved, then isolate issue to source ■ If gas contaminated, other components may be contaminated as well

19

slide-20
SLIDE 20

Detectors ■ Mass Spectrometers

■ Check the tune report, and/or air-water report ■ Is the detector working properly? ■ Is there sufficient vacuum? ■ Is the source clean?

20

slide-21
SLIDE 21

Detectors ■ Other detectors (ECD, TCD)

■ Remove column and cap detector ■ Run a blank and compare detector response to that of blank with column installed ■ Improved: contamination from GC or autosampler ■ Same: contamination from make up gas, detector itself

21

slide-22
SLIDE 22

Inlet Maintenance ■ Detector has been ruled out ■ Replace inlet liner, septum ■ Inspect seals for damage/discoloration and replace ■ Clip the column ■ Verify correct column depth and re-install ■ Re-run system blank

22

slide-23
SLIDE 23

Other Inlet Issues

■ Split ratio

■ Verify that split ratio valve is functioning properly ■ Sticking or non-opening valve can cause overloading of inlet, column and detector

■ Septum purge

■ Faulty septum purge can allow septum off-gassing to reach detector. ■ Tubing can collect contamination

23

slide-24
SLIDE 24

Column Issues ■ Detector and inlet issues ruled out ■ Column clipped during inlet maintenance ■ Swap column with a new one or other column of similar phase

■ Issue resolved, replace column

24

slide-25
SLIDE 25

Autosamplers ■ Liquid Injection ■ Headspace ■ Purge and Trap

25

slide-26
SLIDE 26

Liquid Injection Autosamplers ■ Syringe

■ Replace/clean- verify no change

■ Cleaning solvent

■ Verify adequate number of cleans between samples with uncontaminated solvent

■ Vials

■ Verify contamination-free sample vials

26

slide-27
SLIDE 27

Headspace Autosamplers ■ Needle ■ Transfer line ■ Sample loop ■ Vials ■ 6-port valve

27

slide-28
SLIDE 28

Headspace Autosamplers ■ Vials

■ Verify clean, compare to a different lot/manufacturer

■ Needle

■ Often the source of contamination ■ Clean/replace and re-run

■ Trap

■ In dynamic systems, replace trap and re-run to verify

28

slide-29
SLIDE 29

Headspace Autosamplers ■ Transfer line, valve oven, sample loop

■ Verify temperatures sufficient to prevent condensation of compounds ■ If condition persists, active site or more persistent contamination

29

slide-30
SLIDE 30

Purge and Trap Systems ■ Two main components

■ Concentrator

■ Sparger, sample mount, analytical trap, condensate trap, 6-port valve, transfer line

■ Autosampler

■ Sample needle, syringe/sample loop, transfer tubing, standards injector

30

slide-31
SLIDE 31

Concentrator ■ Typically carryover is just late eluters, but contamination can be anything ■ Check temps during Desorb and Bake

■ Ensure desorb of trap and cleanup between samples

31

slide-32
SLIDE 32

Concentrator ■ Desorb only

■ Rules out gas pathway ■ If contaminated, change trap and re-run ■ Still have issues, likely an active site in gas pathway, including tubing, 6 port, transfer line, and/or condensate trap

32

slide-33
SLIDE 33

Concentrator ■ Disconnect autosampler/bypass autosampler functions ■ Liquid pathway

■ Sparger, sample introduction valve, drain valve

■ Sparger

■ Check for discoloration. Clean or replace, then run blank ■ Check drain valve for proper function

33

slide-34
SLIDE 34

Autosampler ■ Check hot water and/or MeOH Rinse

■ Ensure cleanup of liquid path between samples

■ Possible sources of contamination

■ Sample vials ■ Multi-stage needle ■ Syringe/sample loop ■ Standard addition system

34

slide-35
SLIDE 35

Autosampler ■ Soil autosampler

■ Contamination in water mode or soil mode? Run each to verify ■ Soil mode only: needle, soil valve, transfer line ■ Water mode only: liquid handling pathway ■ Both: rinse water, syringe, needle

35

slide-36
SLIDE 36

Autosampler ■ Syringe based system

■ Bypass needle by placing transfer tubing directly into vial

  • f water

■ Clean/replace syringe ■ Bypass standard addition system ■ Still have issues, multi-port valve

36

slide-37
SLIDE 37

Autosampler ■ Loop and valve system

■ Verify loop rinse times, ensure adequate to rinse loop between samples ■ Clean/replace loop ■ Bypass standard addition system ■ Clean/replace transfer tubing

37

slide-38
SLIDE 38

Preventative Actions ■ Maintain proper methods

■ Before making method changes, as self “Why?” ■ Retain records of changes made and why ■ Keep a copy of last known “good” method ■ If its not broke, don’t fix it!

38

slide-39
SLIDE 39

Preventative Actions ■ Regular preventative maintenance

■ GC maintenance

■ Inlet maintenance, clip column, regular detector cleaning/tuning

■ Column life

■ Condition columns properly and replace on a regular basis as needed ■ Don’t wait until problems arise, waste valuable time troubleshooting!

39

slide-40
SLIDE 40

Preventative Actions ■ Autosampler system maintenance

■ Clean/replace syringes, rinse solvents/water and other consumables on a regular basis ■ Analytical traps

■ Same as columns, condition properly and replace regularly as needed ■ Monitor purge and bake pressures, spike in pressure may signal time to replace

40

slide-41
SLIDE 41

Preventative Actions ■ Consumables

■ Purchase standards, solvents and vials from reputable manufacturers ■ Refresh standards and solvents regularly to prevent contamination ■ Use good practice when cleaning syringes, flasks and other glassware

41

slide-42
SLIDE 42

Preventative Actions ■ Know your samples

■ Consider pre-screening for P&T samples

■ Headspace, prevents contamination from high conc. samples

■ Consider in-vial purge for difficult matrices ■ Consider headspace to prevent column fouling ■ Increase maintenance frequency if/when necessary

42

slide-43
SLIDE 43

As Always, We’re Here to Help! ■ Don’t let troubleshooting scare you, we’re here to help!

■ Tech Support 1-800-874-2004 ■ Tekmar_IntlTech@Teledyne.com ■ tekmarsupport@teledyne.com ■ www.teledynetekmar.com ■ Jaocb.Rebholz@Teledyne.com

43