Detecting Next to Nothing: Spectroscopy in Optical Cavities Kevin - - PowerPoint PPT Presentation
Detecting Next to Nothing: Spectroscopy in Optical Cavities Kevin - - PowerPoint PPT Presentation
Detecting Next to Nothing: Spectroscopy in Optical Cavities Kevin Lehmann Kevin Lehmann Departments of Chemistry & Physics University of Virginia Cavity Ring Down Spectroscopy Collaborators Daniele Romanini Peter Tarsa
Collaborators
- Daniele Romanini
- Joan Gambogi
- John Dudek
- Peter Tarsa
- Iris Scheele
- Haifeng Huang (UVa)
Cavity Ring Down Spectroscopy
- Greg Engel
- Wilton Virgo
- Paul Johnston (UVa)
- Paul Rabinowitz
Wen-Bin Yan, Calvin Kruzen, Bob Augustine, Chris Wu, Yu Chen, Lisa Bergson
Generic Absorption Spectroscopy Instrument
Lock-In
L Cavity Ring Down Spectroscopy Source Filter Sample Chopper Detector
Lock-In
DAS L
Beer’s Law
- Iout(ν) = Iin(ν) exp ( - σ(ν) N L)
− σ(ν): Absorption cross section – N: Number density of absorber
Cavity Ring Down Spectroscopy
– N: Number density of absorber (Concentration) – L: Optical pathlength through sample
- Minimum Detectable Concentration:
Nmin = (∆I/I)min / (σ(ν) L)
Review of Optical Cavities (aka etalons)
Cavity Ring Down Spectroscopy
(aka etalons)
Radii of Curvature: R1, R2 Length of Cavity: L detector Simple Linear Optical Cavity Cavity Ring Down Spectroscopy Length of Cavity: L Mirror Transmission: T Mirror Reflectivity: R Mirror Loss: A = 1 - R - T
Stable Optical Cavities
- For R = 1, modes exist which exactly reproduce
themselves upon round trip.
- 0 < L < R1 or R2 < L < R1+R2 (R1 < R2)
– If R2 - R1, << L, then R1 < L < R2 only weakly unstable
- Optic axis defined by line through centers of curvature of
Cavity Ring Down Spectroscopy
- Optic axis defined by line through centers of curvature of
mirrors
- Light rays will oscillate around optic axis.
- Spot size of mirror
- Cavity Transmission as function of Mirror Reflectivity
Cavity Ring Down Spectroscopy
- Ring-down Time
Peak Transmission of Cavity
- Peak Intracavity Gain
- Cavity Ring Down Spectroscopy
Mirrors with T,A ~ 5 ppm are available in near IR and red Intracavity power gain of ~105 can be realized
Transverse resonance modes of Cavities
- Cavity Ring Down Spectroscopy
If δg = M/N, then we have rational cavity with periodic transmission spectrum. Arbitrary pulse inside Cavity will exactly reshape after N round trips
- - such cavities are used for Herriott Cells.
Some Uses for Optical Cavities in Spectroscopy
- Control frequency and linewidth of lasers
- Monitor laser scan for calibration
- Laser linewidth is reduced by locking on to the
transmission peak of a cavity
Cavity Ring Down Spectroscopy
transmission peak of a cavity
- Cavities are used to build up intensity
– External c.w. second harmonic generators – Pump extremely weak transitions
- Used to enhance optical absorption
Cavity Ring-Down Absorption Spectroscopy
R>99.99% R>99.99%
Absorption Cavity Ring Down Spectroscopy
Time Intensity
Where: c : speed of light L : length of cavity R : mirror reflectivity σ : absorption cross section N : number density (concentration) Leff: effective pathlength
Absorption I(t) = I0 exp[−t( c Lln R+ c⋅σ(λ)⋅ N)] Leff=L/(1-R)
Ring Down Cavity Technique
First Developed by O’Keefe and Deacon
- Rev. Sci. Instr. 59, 2544 (1988)
Theory: Romanini and Lehmann
- J. Chem. Phys. 99, 6287 (1993)
Cavity Ring Down Spectroscopy
- Use a passive optical cavity formed from two
high reflective mirrors (T~1-100 ppm)
- Excite cavity with a pulsed laser to ‘fill’ with photons
- Detect exponential decay of light intensity inside resonator
- Decay rate reflects:
– Loss due to mirrors (slowly changing with wavelengths) – Absorption of gas between mirrors
Advantages of CRDS Method
- Allows much longer pathlengths than traditional multipass cells
- Only sensitive to absorption and scattering between mirrors
- Beer’s Law holds for all pathlengths; pathlengths determined by time
– if resolution exceeds width of absorption lines – Calibration samples are not needed
Cavity Ring Down Spectroscopy
– Calibration samples are not needed
- Cell is very compact; light contained in narrow spot of ~ 1 mm2
- Cell insensitive to vibration since it is a stable optical cavity
- Amplitude noise of laser not important
- Can use low power optical sources
Continue growth- publications per year
1980 1 1992 1 2001 60 1993 5 2002 77 1984 2 1994 12 2003 69 1985 1 1995 18 2004 99 1996 20 2005 95
Cavity Ring Down Spectroscopy
1996 20 2005 95 1988 3 1997 30 2006 144 1998 40 2007 130 1990 3 1999 68 1991 2 2000 48
Based upon searches of Web of Science data base for CRDS, CRLAS, CEAS, ICOS, NICE-OHMS
Early CRDS Work @ Princeton
- Used to detect high overtone transitions
- f HCN and other molecules
– Provided way to determine absolute
Cavity Ring Down Spectroscopy
– Provided way to determine absolute absorption strength of extremely weak transitions. – Could be done with very simple experimental set-up compared to intracavity photoaccoustic spectroscopy.
Cavity Ring Down Spectroscopy
- D. Romanini and KKL, J. Chem. Phys. 99, 6287-301 (1993)
HCN (106) overtone band L(eff) = 24 km Cavity Ring Down Spectroscopy This spectrum is now on the cover of a Spectroscopy text by Hollis
Diode Laser Advantages
- Low cost, compact, all solid state
- Low power requirements
Cavity Ring Down Spectroscopy
- Wide electronic frequency tuning
- Single mode diodes in the near-IR are
becoming available for sensing apps.
– H2O, C2H2, CH4, CO2, NO2, NH3, etc.
Faraday Isolator Mirror Acousto-Optical Modulator Diode Laser HR Mirror
Experimental Setup
Mode Matching Optics
Cavity Ring Down Spectroscopy
Cavity Ring-Down Absorption Cell HR Mirror HR Mirror
3 PZTs
Trigger Computer Photodiode Mirror
- J. B. Dudek et al., Analytical Chemistry 75, 4599-4609 (2003).
0.3 0.4 0.5
Cavity Ring-Down Decay
τ = 295.82+0.20 µsec
nal (Volts) Cavity Ring Down Spectroscopy
200 400 600 800 1000 1200 1400 1600 0.0 0.1 0.2 0.3
τ = 295.82 0.20 µsec data points fit
Signa
Time (µsec)
125 130 135 140
7161.5 cm-1
Water Scan with Lorentz fit
Cavity Ring Down Spectroscopy
7160.8 7161.0 7161.2 7161.4 7161.6 7161.8 7162.0 105 110 115 120
7161.5 cm-1 S=1.5×10-20 ~70 ppb
Wavenumber
How stable is the ring down time?
Laser power upon entering the cavity: 1.397 µm ~ 3 - 5 mW Ensemble Standard Deviation:
Cavity Ring Down Spectroscopy
Ensemble Standard Deviation: 10 pts: 201.384 ±0.139 µs 0.069% 100 pts: 201.378 ±0.165 µs 0.082% 1000 pts: 201.304 ±0.146 µs 0.073%
Allan variance and the detection limit in CRDS
In CRDS the accurate measurement of decay time constant can be limited by slow drift of setup. Allan variance can be used to analyze the stability of instrument. For a N time-series data xi the Allan variance is given by:
- k is the subgroup size and m+1 is the number of subgroups. The integration time T equals to k/f, where f is sample
- rate. When white noise is dominant in the system (uncorrelated decays), Allan variance is proportional to 1/T and
averaging data can improve the signal to noise ratio. When the drift appears Allan variance will become larger. The longest T during which the instrument can be regarded stable is determined by the drift of the system. The minimum
- f Allan variance gives the smallest detectable change during the longest integration time period.
Applied Physics B 57, 131- 139 (1993)
Cavity Ring Down Spectroscopy
- f Allan variance gives the smallest detectable change during the longest integration time period.
What does this mean for water detection?
- Noise equivalent to 68 pptv divided by
the square root of the number of ring
Cavity Ring Down Spectroscopy
the square root of the number of ring down events averaged to get signal, i.e. ~2 pptv for averaging 1000 decays.
CRDS in Practice
Cavity Ring Down Spectroscopy
MTO-1000-H2O
Size (14”x 19”x 26”) Weight (45 kg)
MTO Response to Moisture Addition
Data used with permission from Air Products and Chemicals Inc., Allentown, PA
1.5 2 2.5
ing (ppb)
Cavity Ring Down Spectroscopy
0.5 1 6/15/02 3:36 6/15/02 6:00 6/15/02 8:24 6/15/02 10:48 6/15/02 13:12 6/15/02 15:36 6/15/02 18:00
Time Readin
Variations on CRDS Method
- CRDS (a.k.a. CRLAS, RDCS, cavity leak-out
spectroscopy)
– pulsed CRDS – cw CRDS – phase shift CRDS – Fourier Transform CRDS
Cavity Ring Down Spectroscopy
– Fourier Transform CRDS – broad band CRDS – evanescent wave CRDS – fiber optic CRDS, fiber loop CRDS – Cavity Ring-down polarimetry – Optical feedback CRDS
- Cavity Enhanced Absorption Spectroscopy
(CEAS)-Engleln, Meijer, et al.
– a.k.a Integrated cavity output spectroscopy (ICOS) - O’Keefe – Frequency chirped CEAS
- Noise Immune Intracavity optical
Cavity Ring Down Spectroscopy
- Noise Immune Intracavity optical
heterodyne method (NICE-OHMS)
- Intracavity laser absorption spectroscopy
(ICLAS)
- Intracavity photoaccoustic spectroscopy
– attractive with optical locking!
Excellent Review: C. Vallance, New J. of Chem. 29, 869 (2005)
SUPERCONTINUUM BASED BROADBAND CAVITY ENHANCED ABSORPTION SPECTROSCOPY
Cavity Ring Down Spectroscopy
Paul S. Johnston Kevin K. Lehmann Department of Chemistry University of Virginia
Broad Bandwidth
- Light sources: broad bandwidth dye lasers, Free
electron lasers, fs-lasers, LEDs, arc-source
- Engeln & Meier, Fourier transform CRDS,
1996
Cavity Ring Down Spectroscopy
1996
- Thorpe & Ye, 2007
– Mode lock sources with cavities a multiple of the laser repetition rate allows much improved transmission – Cavity dispersion a challenge
Source for Broad Bandwidth Coherent Radiation: Supercontinuum Photonic Crystal Fibers
- Material: Pure Silica
- Core diameter: 4.8 + 0.2 µm
- Cladding diameter: 125 + 3 µm
- Zero dispersion wavelength: 1040 + 10 nm
- Nonlinear Coefficient at 1060 nm: 11 (W·Km)-1
Cavity Ring Down Spectroscopy
www.crystal-fibre.com
Supercontinuum parameters
- Input
– Average power: 1.0 W – Rep rate: 30 KHz
Cavity Ring Down Spectroscopy
– Rep rate: 30 KHz – Pulse energy: 34 µJ, 10 ns – Peak power: 3400 W
- Output
– Average output power: 0.29 W
! "##$$
Supercontinuum Output
Cavity Ring Down Spectroscopy
White light sources
Cavity Ring Down Spectroscopy
%&&'
Prism Ring-down Resonator
Output
θb
Input
θb
P- polarization
Cavity Ring Down Spectroscopy
b
6 meter radius
- f curvature
P- polarization
- G. Engel et al., in Laser Spectroscopy XIV International Conference,
- Eds. R. Blatt et al. pgs. 314-315 (World Scientific, 1999).
- Wide spectral coverage - Simultaneous
detection of multiple species
- Compact ring geometry (optical isolation)
Advantages of Prism Cavity
Cavity Ring Down Spectroscopy
- Compact ring geometry (optical isolation)
- No dielectric coatings (harsh environments)
- Coupling can be optimized for broadband
Broadband system using white light from photonic crystal fiber
Mode Matching Mirrors 20x objective
Fiber Output
Collimating mirror PC Fiber
Cavity Ring Down Spectroscopy
()"* %#*'"**+,- (%."* " .**/'"& 01%2/ Time
λ
CCD array Sample Cell Gas inlet Mirrors Nd:Vanadate laser
Loss Due to Dispersion changing Brewster’s Angle in Fused Silica
ppm R 1 ) 1 . , 46 . 1 ( = °
2 4 ) 1 (
6 2 4
) , ( δθ δθ
n n
n R
−
=
- Cavity Ring Down Spectroscopy
ppm R ppm R 7 . 98 ) . 1 , 46 . 1 ( 1 ) 1 . , 46 . 1 ( = ° = °
**Only Brewster’s angle loss**
Modeling Cavity Loss
Cavity Ring Down Spectroscopy
- Model:
loss angle s Brewster' scattering Loss + =
High Equivalent Reflectivity
Ring-down Test for Fused Silica Prism
Fused Silica Prisms (built in 43cm cavity) ring down tau/ppm loss vs. wavelength
40 50 60 70 icrosecond) 90 110 130 150
loss
Tau Trend measure Tau ppm loss Trend Measure PPM Loss
Near-IR Prism Cavity Loss Measurements (Tiger Optics) Cavity Ring Down Spectroscopy
10 20 30 40 1300 1350 1400 1450 1500 1550 1600 1650 1700
Wavelength(nm)
Ringdown(tau-mic 10 30 50 70 90
ppm lo
Tau measurenment at 1310, 1368, 1377,1392,1522,1531, 1578,1635,1671nm. Every Diode laser Temp scan from (40
- r ) 35~0 Celsisus
degree.
Cavity enhanced spectroscopy
- Measure time integrated intensity
- Cavity Ring Down Spectroscopy
- Advantages
– Relatively high sensitivity – Simpler set up
- Sensitivity limitations
– Residual mode structure – Laser noise
- Berden, G.; Peeters, R.; Meijer, G. Int. Rev. Phys. Chem. 2000, 19, 565.
Atmospheric Oxygen
Cavity Ring Down Spectroscopy
Current Status
- Collaborating with Tiger Optics to
commercialize a detector of multiple chemical species.
- Will try near-IR spectroscopy with InGaAs
array detector
- Will use FT-IR for dispersion
Cavity Ring Down Spectroscopy
- Will use FT-IR for dispersion
- Have begun building mode-locked (80 MHz)
super-continuum source that we expect > 10 W average power.
– Frequency comb of source can be matched to frequency comb of cavity transmission to greatly improve transmission – Potentially can use “Vernier” principle to improve upon resolution of spectrograph
Applications of Broad band CRDS
- Combustion and Plasma diagnostics
– parallel detection improves S/N ratio if we have unstable sample – “single shot” determination of temperature Cavity Ring Down Spectroscopy – “single shot” determination of temperature
- Breath Analysis Applications
– both species and isotopic composition studies
- Could be combined with optical comb technology for
high accuracy metrology.
Other Cavity Spectroscopy related projects
- Methane Isotope Ratio Instrument
– 13C/12C and D/H ratios – CH4 in oceans, air , and emitted from permafrost. – Possible mission to Mars
- Detect NO and other molecules in human breath
Cavity Ring Down Spectroscopy
- Detect NO and other molecules in human breath
- Fabricate and test Prisms from CaF2 (UV) and
BaF2 (IR)
- Collaboration with group looking for WIMPS
with liquid Ar detector
– Need to detect H2O, O2, and N2 impurities < 1 ppb – Plan to try to detect N2 via 3Σ state produced in discharge.
Contact Information
- Phone 243-2130
- Office: Rm. 124 in Chemistry Building
- Email: Lehmann@virginia.edu
Cavity Ring Down Spectroscopy
- Email: Lehmann@virginia.edu