Status of the Photosensor Testing Facility (PTF) at TRIUMF Tom Feusels for Team PTF University of British Columbia 6th Open Meeting for the Hyper-Kamiokande Project Jan 30, 2015 T. Feusels (UBC) PTF Status 30/01/2015 1 / 30 Table of


slide-1
SLIDE 1

Status of the Photosensor Testing Facility (PTF) at TRIUMF

Tom Feusels for Team PTF

University of British Columbia

6th Open Meeting for the Hyper-Kamiokande Project Jan 30, 2015

  • T. Feusels (UBC)

PTF Status 30/01/2015 1 / 30

slide-2
SLIDE 2

Table of Contents

1

PTF

2

8” HPD Performance at PTF

3

PTF Measurements

  • T. Feusels (UBC)

PTF Status 30/01/2015 2 / 30

slide-3
SLIDE 3

Table of Contents

1

PTF

2

8” HPD Performance at PTF

3

PTF Measurements

  • T. Feusels (UBC)

PTF Status 30/01/2015 3 / 30

slide-4
SLIDE 4

Objectives

Y.Nishimura

Characterize and study optical properties of PMTs under consideration for HK and near detector. Measure PMT acceptance. Map reflectivity of PMT: important for reconstruction algorithms. Measure wavelength and magnetic field dependence. ...

  • T. Feusels (UBC)

PTF Status 30/01/2015 4 / 30

slide-5
SLIDE 5

The PTF setup

x y z rotation θ tilt φ receiver PMT gantry 0 gantry 1

photo- cathode water- level

5 stepping motors for each of two manipulator arms (gantries) ⇒ 5D (x,y,z, rotation, tilt) Optical box contains USB powered board with 3-axis magnetometer, accelerometer and gyroscope (Phidget). Active cancellation with Helmholtz coil, passive cancellation with two layers of g-iron. Light shielding with dark curtains.

  • T. Feusels (UBC)

PTF Status 30/01/2015 5 / 30

slide-6
SLIDE 6

The PTF setup

Phidget Receiver PMT Monitor PMT Light source

Waterproof optical box with laser, monitor and receiver PMT attached to the head of the gantry arm. Multiple light sources: 405 and 467nm pulsed laser, Xe lamp with several filters. Collimator, polarizer and beam splitter.

  • T. Feusels (UBC)

PTF Status 30/01/2015 6 / 30

slide-7
SLIDE 7

Mechanical system: alignment

Improvement of mechanical stability and reliability. Alignment of boundaries of X, Y and Z motion for both gantries with respect to each other up to 2 mm and maximize phase space. Alignment of rotation axis: needs to be improved. February: Full survey and realignment using laser trackers. Alignment measurements and measurement accuracy studies under development.

  • T. Feusels (UBC)

PTF Status 30/01/2015 7 / 30

slide-8
SLIDE 8

DAQ software

DAQ and control of PTF through MIDAS. Improvements of the DAQ software: increase of phase space for scans and reliability of long scans. User friendly interface for taking measurements. New measurement sequences being tested: alignment, reproducibility, normal incidence scans,... Automatic conversion of MIDAS files to ROOT files with a flat tree structure. Automatic run logging.

  • T. Feusels (UBC)

PTF Status 30/01/2015 8 / 30

slide-9
SLIDE 9

Water system

UV sterilizer

2stage 1mm, 50nmultrafilter venturi vacuum systemfor degassing Water pumping and purification system ready. Tests in the PTF for early March when tank is in place. Particle counters for water quality monitoring installation at end of March.

  • T. Feusels (UBC)

PTF Status 30/01/2015 9 / 30

slide-10
SLIDE 10

Table of Contents

1

PTF

2

8” HPD Performance at PTF

3

PTF Measurements

  • T. Feusels (UBC)

PTF Status 30/01/2015 10 / 30

slide-11
SLIDE 11

HPD setup in PTF: noise control

July 2014: 8” HPD prototype arrives at TRIUMF. Proper shielding and grounding of power supply cables crucial for noise reduction. Motors switched off during measurement to limit noise from motor controller. Increased AD bias voltage to almost maximum rating (340V) by increasing LV control voltage (2.33 to 2.8V).

  • T. Feusels (UBC)

PTF Status 30/01/2015 11 / 30

slide-12
SLIDE 12

HPD signal

Motors switched off during measurement to limit noise from motor controller. Properly shielding and common grounding of power supply cables crucial for noise reduction. Increased AD bias voltage to almost maximum rating (340V) by increasing LV control voltage (2.33 to 2.8V).

  • T. Feusels (UBC)

PTF Status 30/01/2015 12 / 30

slide-13
SLIDE 13

8” HPD pedestal and single PE distribution

100 150 200 250 300 350 400 1000 2000 3000 4000 5000

ADC value

1 PE resolution: 11.24% Peak to valley ratio: 5.25 Signal to noise ratio: 12.27

100 150 200 250 300 350 400 1 10

2

10

3

10

4

10

1 PE resolution: 11.24% Peak to Valley ratio: 5.25 Signal to noise ratio: 12.27 ADC value

Excellent 1pe resolution. Very good peak to valley ratio and signal to noise1 ratio.

1Signal to noise ratio = ratio of SPE peak to pedestal sigma

  • T. Feusels (UBC)

PTF Status 30/01/2015 13 / 30

slide-14
SLIDE 14

8” HPD pedestal and single PE distribution

LV (V) 2.3 2.4 2.5 2.6 2.7 2.8 2 4 6 8 10 12 14 16 18 20 1PE re solution(%) P/V ra tio S/N ra tio

AD Bias Dependence ADC value 200 400 600 800 1000 1200 50 100 150 200 250 300 350 400

Optimal 8” HPD performance at PTF at LV control voltage of about 2.75-2.8V. Individual PE peaks become very clear up to 5 PE in charge distribution.

  • T. Feusels (UBC)

PTF Status 30/01/2015 14 / 30

slide-15
SLIDE 15

HPD timing resolution

Fire laser at HPD. Use 500 MHz digitizer (Caen V1730) for acquiring waveform of monitor PMT and HPD. Width of ∆t = fitted monitor PMT LE time - fitted HPD PMT LE time is resolution.

  • T. Feusels (UBC)

PTF Status 30/01/2015 15 / 30

slide-16
SLIDE 16

HPD performance at PTF

8” HPD @Kamioka 8” HPD @PTF (LV: 2.8V) 1PE resolution σ/µ 12% 11.2% Peak to valley ratio 5.2 5.25 Signal to Noise 12 12.3 Timing resolution σ (ns) 1.1 1.05 Same performance of 8” HPD in PTF as in Kamioka after improved noise control. Next: Dark noise, pulse shape and stability (gain, DN). PTF in good shape for 20” PMT measurements.

  • T. Feusels (UBC)

PTF Status 30/01/2015 16 / 30

slide-17
SLIDE 17

Table of Contents

1

PTF

2

8” HPD Performance at PTF

3

PTF Measurements

  • T. Feusels (UBC)

PTF Status 30/01/2015 17 / 30

slide-18
SLIDE 18

8” HPD: first intensity map

x y z = fixed tilt φ = -90° gantry 0

photo- cathode

X position (m) 0.05 0.1 0.15 0.2 0.25 0.3 Y position (m) 0.15 0.2 0.25 0.3 0.35 0.4 0.45 0.5 1 1.5 2 2.5 3 HPD Mean charge response (PE)

Scanning the PMT with the laser pointing vertically downwards in 2cm steps. Using fixed pedestal subtraction and 1PE gain here to calculate average PE per point. Real asymmetry? → Rotate PMT. Actual acceptance should be measured with normal incidence: need PMT curvature and new measurement sequence.

  • T. Feusels (UBC)

PTF Status 30/01/2015 18 / 30

slide-19
SLIDE 19

HPD vertical 5mm scan with two HPD orientations

0.5 1 1.5 2 2.5 3 3.5 X position (m) 0.05 0.1 0.15 0.2 0.25 0.3 Y position (m) 0.15 0.2 0.25 0.3 0.35 0.4 0.45

HPD Mean charge response (p.e.) HPD Mean charge response (p.e.)

0.5 1 1.5 2 2.5 3 3.5 X position (m) 0.05 0.1 0.15 0.2 0.25 0.3 Y position (m) 0.15 0.2 0.25 0.3 0.35 0.4 0.45

HPD Mean charge response (p.e.) HPD Mean charge response (p.e.)

Fine grained scan shows interesting non uniform pattern with large local variation. Circle with increased intensity possibly from internal reflection on metal inside the PMT (also observed by Hamamatsu). Pattern rotates as HPD rotates. Next: low intensity scan to map gain dependency.

  • T. Feusels (UBC)

PTF Status 30/01/2015 19 / 30

slide-20
SLIDE 20

HPD vertical 5mm scan with two wavelengths

0.5 1 1.5 2 2.5 3 3.5 X position (m) 0.05 0.1 0.15 0.2 0.25 0.3 Y position (m) 0.15 0.2 0.25 0.3 0.35 0.4 0.45

HPD Mean charge response (p.e.) HPD Mean charge response (p.e.)

0.5 1 1.5 2 2.5 3 3.5 4 4.5 X position (m) 0.05 0.1 0.15 0.2 0.25 0.3 Y position (m) 0.15 0.2 0.25 0.3 0.35 0.4 0.45

HPD Mean charge response (p.e.) HPD Mean charge response (p.e.)

405nm 467nm Different laser intensity for both! Similar pattern seen in both wavelengths.

  • T. Feusels (UBC)

PTF Status 30/01/2015 20 / 30

slide-21
SLIDE 21

HPD vertical 5mm scan with 5◦ tilt

x y z = fixed tilt φ = -85° gantry 0

photo- cathode

0.5 1 1.5 2 2.5 3 3.5 X position (m) 0.05 0.1 0.15 0.2 0.25 0.3 Y position (m) 0.15 0.2 0.25 0.3 0.35 0.4 0.45

HPD Mean charge response (p.e.) HPD Mean charge response (p.e.)

0.5 1 1.5 2 2.5 3 3.5 X position (m) 0.05 0.1 0.15 0.2 0.25 0.3 Y position (m) 0.15 0.2 0.25 0.3 0.35 0.4 0.45

HPD Mean charge response (p.e.) HPD Mean charge response (p.e.)

0.5 1 1.5 2 2.5 3 3.5 X position (m) 0.05 0.1 0.15 0.2 0.25 0.3 Y position (m) 0.15 0.2 0.25 0.3 0.35 0.4 0.45

HPD Mean charge response (p.e.) HPD Mean charge response (p.e.)

Similar pattern seen, although different HPD cross section scanned.

  • T. Feusels (UBC)

PTF Status 30/01/2015 21 / 30

slide-22
SLIDE 22

HPD vertical 5mm scan with 5◦ tilt

0.5 1 1.5 2 2.5 3 3.5 4 X position (m) 0.05 0.1 0.15 0.2 0.25 0.3 Y position (m) 0.15 0.2 0.25 0.3 0.35 0.4 0.45

HPD Mean charge response (p.e.) HPD Mean charge response (p.e.)

0.5 1 1.5 2 2.5 3 3.5 X position (m) 0.05 0.1 0.15 0.2 0.25 0.3 Y position (m) 0.15 0.2 0.25 0.3 0.35 0.4 0.45

HPD Mean charge response (p.e.) HPD Mean charge response (p.e.)

0.5 1 1.5 2 2.5 3 3.5 X position (m) 0.05 0.1 0.15 0.2 0.25 0.3 Y position (m) 0.15 0.2 0.25 0.3 0.35 0.4 0.45

HPD Mean charge response (p.e.) HPD Mean charge response (p.e.)

0.5 1 1.5 2 2.5 3 3.5 X position (m) 0.05 0.1 0.15 0.2 0.25 0.3 Y position (m) 0.15 0.2 0.25 0.3 0.35 0.4 0.45

HPD Mean charge response (p.e.) HPD Mean charge response (p.e.)

Similar pattern seen, although different HPD cross section scanned.

  • T. Feusels (UBC)

PTF Status 30/01/2015 22 / 30

slide-23
SLIDE 23

HPD receiver PMT

X position (m) 0.05 0.1 0.15 0.2 0.25 0.3 Y position (m) 0.15 0.2 0.25 0.3 0.35 0.4 0.45

0.5 1 1.5 2 2.5 3 3.5

X position (m) 0.05 0.1 0.15 0.2 0.25 0.3 Y position (m) 0.15 0.2 0.25 0.3 0.35 0.4 0.45

0.1 0.2 0.3 0.4 Fitting circle using bins with number of PE between 0.45 and 1.2. Inner circle seems offset from center of circle. Use center of PMT sphere in gantry coordinate system for acceptance, measure height manually. Probably only one bin where reflected photon is seen, matches offset of receiver PMT wrt laser.

  • T. Feusels (UBC)

PTF Status 30/01/2015 23 / 30

slide-24
SLIDE 24

HPD receiver PMT

X position (m) 0.14 0.145 0.15 0.155 0.16 0.165 Y position (m) 0.295 0.3 0.305 0.31 0.315 0.32

0.002 0.004 0.006 0.008 0.01 0.012 0.014 0.016 0.018

X position (m) 0.05 0.1 0.15 0.2 0.25 0.3 Y position (m) 0.15 0.2 0.25 0.3 0.35 0.4 0.45

0.1 0.2 0.3 0.4 Detailed mm stepsize scan shows profile of reflected light!

  • T. Feusels (UBC)

PTF Status 30/01/2015 24 / 30

slide-25
SLIDE 25

Summary

Improvements in both software and hardware of PTF. Setup at PTF shows same excellent HPD performance as setup in Kamioka. Water system is ready to be tested in-situ. February: finalize 8” HPD performance characterization, normal incidence and angular acceptance studies. Improvement of alignment and measurement of reproducibility and accuracy. Measurement of PMT curvature and study of effect on normal incidence measurement. 3 20” PMTs will arrive at TRIUMF mid/end of Feb: preparation for installation in tank

  • ngoing.

March: First measurements in water.

  • T. Feusels (UBC)

PTF Status 30/01/2015 25 / 30

slide-26
SLIDE 26

THANKS

UBC, TRIUMF and visitors: Sophie Berkman, Ben Krupicz, Patrick de Perio, Wayne Faszer, Tom Feusels, Mark Hartz, Akira Konaka, Harish Kugel, Thomas Lindner, James Linqiao Liu, Philip Lu, Andy Miller, David Morris, Corina Nantais, Carl Reithmeier, Fabrice Reti` ere, Mark Scott, Nils Smit-Anseeuw, Yusuke Suda, Hiro Tanaka, Shimpei Tobayama, Peter Vincent, Michael Wilking, Stan Yen, Aaron Zimmer

  • T. Feusels (UBC)

PTF Status 30/01/2015 26 / 30

slide-27
SLIDE 27

Backup

  • T. Feusels (UBC)

PTF Status 30/01/2015 27 / 30

slide-28
SLIDE 28

Magnetic field compensation

In addition to the Earth magnetic field, presence of additional magnetic fields (eg. from TRIUMF Cyclotron). Active cancellation with Helmholtz coil, passive cancellation with two layer of g-iron. Field scans with phidgets mounted on gantry arms. Calibration of offsets in magnetometer on phidget using accurate Gaussmeter (up to 0.1mG).

  • T. Feusels (UBC)

PTF Status 30/01/2015 28 / 30

slide-29
SLIDE 29

Optical system

Phidget Receiver PMT Monitor PMT Light source

Light sources: 405nm pulsed laser and Xe lamp with filters. Connected through fiber to optical boxes. Collimator, polarizer, beam splitter. Monitor PMT and receiver PMT.

  • T. Feusels (UBC)

PTF Status 30/01/2015 29 / 30

slide-30
SLIDE 30

HPD signal before noise reduction

150 200 250 300 350 400 450 ADC value 50 100 150 200 250 300 350 400

Pedestal 1 PE

100 200 300 400 500 600 ADC value 20 40 60 80 100 120 140 160 180 200 220

Pedestal 1 PE 2 PE 4 PE 3 PE

  • T. Feusels (UBC)

PTF Status 30/01/2015 30 / 30