spicp ms tier 1 method
play

spICP-MS tier 1 method People behind the experiments Anna - PowerPoint PPT Presentation

spICP-MS tier 1 method People behind the experiments Anna Katarzyna Undas (RIKILT) NanoDefine Outreach Event Brussels, September 2017 NanoDefine is funded by the European Community's Seventh Framework Programme under Grant


  1. spICP-MS – tier 1 method “People behind the experiments” Anna Katarzyna Undas (RIKILT) NanoDefine Outreach Event Brussels, September 2017 NanoDefine is funded by the European Community's Seventh Framework Programme under Grant Agreement-604347

  2. RIKILT (Wageningen University & Research) is your innovative partner for safe and authentic food. We are specialized in (forensic) measurements, perform top level research and by arrangement 24/7 available in case of emergencies Wageningen Expertise areas: • Food fraud and Composition • Effects of substances on humans and animals • Measuring and detecting substances • Safe food production • Training & Consultancy

  3. RIKILT (Wageningen University & Research) RIKILT coordinated EU-projects NanoLyse and NanoDefine and was a research partner in many more EU projects focusing on nanotechnology Wageningen

  4. NanoDefine project – people behind experiments

  5. spICP-MS – EU nanomaterial definition Nanomaterial EU definition: “A natural, incidental or manufactured material containing particles, in an unbound state or as an aggregate or as an agglomerate and where, for 50 % or more of the particles in the number size distribution, one or more external dimensions is in the size range 1 nm - 100 nm .” • Determination of size • Determination of size distribution • Number based concentration

  6. spICP-MS method

  7. spICP-MS – method validation Validation based on: • IUPAC “Harmonized Guidelines for Single - Laboratory Validation of Analytical Methods” • Eurachem Guide “Fitness for Purpose of Analytical Methods” • Commision Decision 2002/657/EC 5 days, n=3 (LOD,LOQ, n=5) TiO 2 in sunscreen Al 2 O 3 in toothpaste 20 – 500 nm 48 – 500 nm Working range 2.4 A , 8.0 B , 19.9 C 5.3 A , 5.8 B , 2.3 C Repeatability (RSD r ) [%] Intermediate precision (RSD RL ) [%] 4.0 A , 16.5 B , 21.3 C 15 A , 24 B , 29 C LOD LODs 20 nm LODs 48 nm LODc 12 mg/kg prod. (10 peaks) LODc 61 mg/kg prod. (10 peaks) LOQ LOQs 22 nm LOQs 70 nm LOQc 30 mg/kg prod. (30 peaks) LOQc 180 mg/kg prod. (30 peaks) Measurement uncertainty [%] 5 A , 25 B , 33 C 17 A , 25 B , 33 C A particle diameter, B particle number conc., C particle mass conc.

  8. Analysis of commercially available toothpaste Product Median size [nm] a 154 b 126 c 143 d 140 e 123 f 110 g 143 h 106

  9. Interesting posters

  10. www.NanoDefine.eu This project has received funding from the European Union’s Seventh Programme for research, technological development and demonstration under grant agreement No 604347

Download Presentation
Download Policy: The content available on the website is offered to you 'AS IS' for your personal information and use only. It cannot be commercialized, licensed, or distributed on other websites without prior consent from the author. To download a presentation, simply click this link. If you encounter any difficulties during the download process, it's possible that the publisher has removed the file from their server.