method optimization series purge parameters selection amy
play

Method Optimization Series: Purge Parameters Selection Amy Nutter - PowerPoint PPT Presentation

Method Optimization Series: Purge Parameters Selection Amy Nutter August, 2018 1 Overview The fundamentals of purge parameters Purge gas Purge volume Purge flow rate and time In-vial vs. traditional Purge and trap 101: What


  1. Method Optimization Series: Purge Parameters Selection Amy Nutter August, 2018 1

  2. Overview ■ The fundamentals of purge parameters ■ Purge gas ■ Purge volume ■ Purge flow rate and time ■ In-vial vs. traditional

  3. Purge and trap 101: What is purge and Trap? ■ Purge and Trap is a technique used to improve the sensitivity of volatile samples beyond the capabilities of static headspace. 3

  4. Purge and trap 101: How does purge and Trap work? ■ Purge Equilibrium Theory ■ Compounds migrate out of a concentrated solution and into a dilute solution to reach equilibrium. ■ Purge and Trap ■ The headspace above the sample is constantly replaced with fresh gas, preventing equilibrium from being reached. 4

  5. Static Headspace VS Purge and Trap Purge and Trap Static Headspace

  6. Basic Purge and Trap Set Up – Purge Purge Gas 6-port valve Heated Transfer Line Gas Chromatograph Trap Carrier/Desorb To Vent Gas

  7. Method Parameters Of Purge and Trap ■ Trap selection ■ Sample volume ■ Purge Method (in-vial vs traditional) ■ Purge gas type and volume (time and flow rate) ■ Dry Purge (time and flow rate) ■ Desorb (time and GC flow rate) 7

  8. Method Parameters Of Purge and Trap ■ Trap selection ■ Sample volume ■ Purge Method (in-vial vs traditional) ■ Purge gas type and volume (time and flow rate) ■ Dry Purge (time and flow rate) ■ Desorb (time and GC flow rate) 8

  9. What is the Purge of Purge and Trap? ■ Process through which the VOCs are stripped out of the sample for analysis ■ Purging allows us to collect only the volatile portion of the sample, solid, liquid portions are retained in the glassware and not transferred to GC inlet ■ Facilitates sample concentration without the use of expensive and/or toxic solvents 9

  10. Fundamentals of Purge Gas ■ Must be inert ■ Cannot react with VOCs or any other component of the sample, known or unknown ■ Cannot be combustible and/or unstable ■ H 2 or O 2 ■ Must be free of VOCs ■ Cannot contribute to the total VOC sample collected, creating false positives and/or “dirty” blanks 10

  11. Purge Gas Selection ■ Helium ■ Traditional choice for superior inertness, stability and compatibility with carrier gas for GCMS ■ Is somewhat expensive and occasionally difficult to source ■ Nitrogen ■ Good alternative to Helium thanks to similar inertness and low cost/ability to use generators ■ Is a larger molecule which can have minor affect on purge efficiency and MS vacuum during desorb 11

  12. He to N 2 Purge Gas Comparison – EPA 524.3/4 ■ P&T parameters Helium/Nitrogen 12

  13. He to N 2 Purge Gas Comparison – EPA 524.3/4 ■ GC Parameters ■ 524.3 13

  14. He to N 2 Purge Gas Comparison – EPA 524.3/4 ■ GC Parameters ■ 524.4 14

  15. Results - Gases 15

  16. Results - Midrange 524.3 524.4 16

  17. Results – Late Eluters 17

  18. Purge Gas Conclusions ■ Method requirements were easily met with both purge gases ■ Excellent sensitivity was seen across all ranges of the chromatogram ■ This is due to sensitivity of modern MSDs as well as the use of Helium as the carrier gas 18

  19. Considerations of Purge Volume ■ Flow rate ■ Faster flow means more purge gas in the same length of time ■ Flow too aggressively, can cause pressure issues and/or push compounds too deeply into sorbent bed ■ Purge time ■ Longer times mean more purge gas ■ Purge too long, waste of purge gas and/or push compounds too deeply into sorbent bed 19

  20. Considerations of Purge Volume 20

  21. Purge Volume Comparison – EPA 524 – Parameters 21

  22. Purge Volume Comparison – EPA 524 – Data 22

  23. Purge Volume Comparison – EPA 524 – Data 23

  24. Purge Volume Comparison – EPA 524 – Data 24

  25. Purge Volume Comparison – EPA 524 – Data 25

  26. Purge Volume Comparison – EPA 524 – Data 26

  27. Purge Volume Comparison – EPA 524 – Data 27

  28. Purge Volume Comparison – EPA 524 – Conclusions ■ Purge time can be greatly reduced, resulting in more samples processed during a time period ■ Gases can be affected at higher purge flow rates ■ Heavier compounds can decrease with the shorter purge times ■ Staying within recommended 524.2 parameters – data should look great 28

  29. In-Vial vs. Traditional Purge – 8260 – Parameters 29

  30. In-Vial vs. Traditional Purge – 8260 – Parameters 30

  31. In-Vial vs. Traditional Purge – 8260 – Data 31

  32. In-Vial vs. Traditional Purge – 8260 – Data 32

  33. In-Vial vs. Traditional Purge – 8260 – Data 33

  34. In-Vial vs. Traditional Purge – 8260 – Data 34

  35. In-Vial vs. Traditional Purge – 8260 – Data 35

  36. In-Vial vs. Traditional Purge – 8260 – Data 36

  37. In Vial vs. Traditional Purge – 8260 – Conclusions ■ Method requirements were easily met with both In- Vial and Traditional purge ■ MDLs for In-Vial purge were slightly higher ■ Increased % Accuracy/Recovery for Traditional purge 37

  38. Teledyne Tekmar’s New Online Store! ■ https://store.teledynetekmar.com ■ One convenient stop for all of your Tekmar product needs ■ See pictures and get part numbers 38

  39. Thank You! For more information: Amy.Nutter@Teledyne.com Amy Nutter Website: www.teledynetekmar.com Phone: 1-800-874-2004 TekmarSales@Teledyne.com Sales: Tekmar_IntlSales@Teledyne.com Tekmar_CSC@Teledyne.com Customer Support: Technical Support: TekmarSupport@Teledyne.com Check out our website for all new applications! You can also find us on Twitter, Facebook, and LinkedIn

Download Presentation
Download Policy: The content available on the website is offered to you 'AS IS' for your personal information and use only. It cannot be commercialized, licensed, or distributed on other websites without prior consent from the author. To download a presentation, simply click this link. If you encounter any difficulties during the download process, it's possible that the publisher has removed the file from their server.

Recommend


More recommend