Method Optimization Series: Purge Parameters Selection Amy Nutter - - PowerPoint PPT Presentation

method optimization series purge parameters selection amy
SMART_READER_LITE
LIVE PREVIEW

Method Optimization Series: Purge Parameters Selection Amy Nutter - - PowerPoint PPT Presentation

Method Optimization Series: Purge Parameters Selection Amy Nutter August, 2018 1 Overview The fundamentals of purge parameters Purge gas Purge volume Purge flow rate and time In-vial vs. traditional Purge and trap 101: What


slide-1
SLIDE 1

Method Optimization Series: Purge Parameters Selection Amy Nutter August, 2018

1

slide-2
SLIDE 2

Overview ■ The fundamentals of purge parameters ■ Purge gas ■ Purge volume

■ Purge flow rate and time

■ In-vial vs. traditional

slide-3
SLIDE 3

Purge and trap 101: What is purge and Trap? ■ Purge and Trap is a technique used to improve the sensitivity of volatile samples beyond the capabilities

  • f static headspace.

3

slide-4
SLIDE 4

Purge and trap 101: How does purge and Trap work? ■ Purge Equilibrium Theory

■ Compounds migrate out of a concentrated solution and into a dilute solution to reach equilibrium.

■ Purge and Trap

■ The headspace above the sample is constantly replaced with fresh gas, preventing equilibrium from being reached.

4

slide-5
SLIDE 5

Static Headspace VS Purge and Trap

Static Headspace Purge and Trap

slide-6
SLIDE 6

Basic Purge and Trap Set Up – Purge

Trap Gas Chromatograph Heated Transfer Line 6-port valve To Vent Carrier/Desorb Gas Purge Gas

slide-7
SLIDE 7

Method Parameters Of Purge and Trap ■ Trap selection ■ Sample volume ■ Purge Method (in-vial vs traditional) ■ Purge gas type and volume (time and flow rate) ■ Dry Purge (time and flow rate) ■ Desorb (time and GC flow rate)

7

slide-8
SLIDE 8

Method Parameters Of Purge and Trap ■ Trap selection ■ Sample volume ■ Purge Method (in-vial vs traditional) ■ Purge gas type and volume (time and flow rate) ■ Dry Purge (time and flow rate) ■ Desorb (time and GC flow rate)

8

slide-9
SLIDE 9

What is the Purge of Purge and Trap? ■ Process through which the VOCs are stripped out of the sample for analysis ■ Purging allows us to collect only the volatile portion

  • f the sample, solid, liquid portions are retained in the

glassware and not transferred to GC inlet ■ Facilitates sample concentration without the use of expensive and/or toxic solvents

9

slide-10
SLIDE 10

Fundamentals of Purge Gas

■ Must be inert

■ Cannot react with VOCs or any other component of the sample, known or unknown

■ Cannot be combustible and/or unstable

■ H2 or O2

■ Must be free of VOCs

■ Cannot contribute to the total VOC sample collected, creating false positives and/or “dirty” blanks

10

slide-11
SLIDE 11

Purge Gas Selection

■ Helium

■ Traditional choice for superior inertness, stability and compatibility with carrier gas for GCMS ■ Is somewhat expensive and occasionally difficult to source

■ Nitrogen

■ Good alternative to Helium thanks to similar inertness and low cost/ability to use generators ■ Is a larger molecule which can have minor affect on purge efficiency and MS vacuum during desorb

11

slide-12
SLIDE 12

He to N2 Purge Gas Comparison – EPA 524.3/4 ■ P&T parameters

12

Helium/Nitrogen

slide-13
SLIDE 13

He to N2 Purge Gas Comparison – EPA 524.3/4 ■ GC Parameters

■ 524.3

13

slide-14
SLIDE 14

He to N2 Purge Gas Comparison – EPA 524.3/4 ■ GC Parameters

■ 524.4

14

slide-15
SLIDE 15

Results - Gases

15

slide-16
SLIDE 16

Results - Midrange

16

524.4 524.3

slide-17
SLIDE 17

Results – Late Eluters

17

slide-18
SLIDE 18

Purge Gas Conclusions ■ Method requirements were easily met with both purge gases ■ Excellent sensitivity was seen across all ranges of the chromatogram ■ This is due to sensitivity of modern MSDs as well as the use of Helium as the carrier gas

18

slide-19
SLIDE 19

Considerations of Purge Volume

■ Flow rate

■ Faster flow means more purge gas in the same length of time ■ Flow too aggressively, can cause pressure issues and/or push compounds too deeply into sorbent bed

■ Purge time

■ Longer times mean more purge gas ■ Purge too long, waste of purge gas and/or push compounds too deeply into sorbent bed

19

slide-20
SLIDE 20

Considerations of Purge Volume

20

slide-21
SLIDE 21

Purge Volume Comparison – EPA 524 – Parameters

21

slide-22
SLIDE 22

Purge Volume Comparison – EPA 524 – Data

22

slide-23
SLIDE 23

Purge Volume Comparison – EPA 524 – Data

23

slide-24
SLIDE 24

Purge Volume Comparison – EPA 524 – Data

24

slide-25
SLIDE 25

Purge Volume Comparison – EPA 524 – Data

25

slide-26
SLIDE 26

Purge Volume Comparison – EPA 524 – Data

26

slide-27
SLIDE 27

Purge Volume Comparison – EPA 524 – Data

27

slide-28
SLIDE 28

Purge Volume Comparison – EPA 524 – Conclusions

■ Purge time can be greatly reduced, resulting in more samples processed during a time period ■ Gases can be affected at higher purge flow rates ■ Heavier compounds can decrease with the shorter purge times ■ Staying within recommended 524.2 parameters – data should look great

28

slide-29
SLIDE 29

In-Vial vs. Traditional Purge – 8260 – Parameters

29

slide-30
SLIDE 30

In-Vial vs. Traditional Purge – 8260 – Parameters

30

slide-31
SLIDE 31

In-Vial vs. Traditional Purge – 8260 – Data

31

slide-32
SLIDE 32

In-Vial vs. Traditional Purge – 8260 – Data

32

slide-33
SLIDE 33

In-Vial vs. Traditional Purge – 8260 – Data

33

slide-34
SLIDE 34

In-Vial vs. Traditional Purge – 8260 – Data

34

slide-35
SLIDE 35

In-Vial vs. Traditional Purge – 8260 – Data

35

slide-36
SLIDE 36

In-Vial vs. Traditional Purge – 8260 – Data

36

slide-37
SLIDE 37

In Vial vs. Traditional Purge – 8260 – Conclusions

■ Method requirements were easily met with both In- Vial and Traditional purge ■ MDLs for In-Vial purge were slightly higher ■ Increased % Accuracy/Recovery for Traditional purge

37

slide-38
SLIDE 38

Teledyne Tekmar’s New Online Store! ■ https://store.teledynetekmar.com ■ One convenient stop for all of your Tekmar product needs ■ See pictures and get part numbers

38

slide-39
SLIDE 39

Thank You!

For more information:

Amy Nutter Amy.Nutter@Teledyne.com

Website:

www.teledynetekmar.com

Phone: 1-800-874-2004 Sales:

TekmarSales@Teledyne.com Tekmar_IntlSales@Teledyne.com

Customer Support:

Tekmar_CSC@Teledyne.com

Technical Support: TekmarSupport@Teledyne.com Check out our website for all new applications! You can also find us on Twitter, Facebook, and LinkedIn