dosimetry of an x ray tube
play

Dosimetry of an x-ray tube and Irradiation results of a DEPFET - PowerPoint PPT Presentation

Dosimetry of an x-ray tube and Irradiation results of a DEPFET matrix Carried out at the University of Karlsruhe IMPRS Workshop Motivation: The KEKB accelerator and High luminosity collider Asymetric energy for studying


  1. Dosimetry of an x-ray tube and Irradiation results of a DEPFET matrix Carried out at the University of Karlsruhe IMPRS Workshop

  2. Motivation: The KEKB accelerator and… • High luminosity collider • Asymetric energy for studying       ( 4 ) e e Y S B B • Measure CKM matrix elements • CP violation Measuring device: IMPRS Workshop 7/21/2009 2

  3. … the Belle II detector • The two innermost layers of the vertex detector will be consisted of DEPFET matrices • Ionizing Radiation from: • Beam-Gas Interaction • Synchrotron radiation will damage those matrices Ca. 1 Mrad/a (=10 kGy/a) • • Simulating of these various sources with x-ray radiation IMPRS Workshop 7/21/2009 3

  4. What are DEPFETs? • Depend on sideward depletion: Internal Gate • Generated charges will be stored in the Internal Gate • Modulation of Drain- Source current • Clear operation necessary • Beneath Gate metallization: Layer of silicon dioxide Sensitive to ionizing radiation IMPRS Workshop 7/21/2009 4

  5. Damage mechanism • Generataion of electron/hole pairs in the oxide • Mobility of holes is weak in comparison to electrons • Holes will be trapped near the interface and stay there for O(h) till O(a) • Also: Increase of interface state density IMPRS Workshop 7/21/2009 5

  6. Dosimetry of the x-ray facility in Karlsruhe • To quantify the damage done by ionizing radiation Applied Dose is necessary • Calibration was given,  Recheck it  Problem Challenge started … IMPRS Workshop 7/21/2009 6

  7. Recipe for succsesfull Calibration • X-ray source • Detector a) Fast readout, highly sensitive, good spectral response, low noise, superb amplifier b) If not (a), then find a compromise of (a) • Energy spectrum would be good a) Measure one b) Simulate one Thanks to my colleague: Oksana Brovchenko IMPRS Workshop 7/21/2009 7

  8. 1. Determine dose rate in silicon diode • Dose rate measurement via depleted diode • Measuring of reverse-current • X-ray photons generate electron/hole-pairs in Si- Bulk – Every charge carrier pair represents an energy of 3.6 eV – With the x-ray generated current one gets the deposited power, with the mass of the diode  dose rate in Si IMPRS Workshop 7/21/2009 8

  9. 2. Making use of the spectrum • Spectrum of tungsten anode (including a 0.4 mm Be filtering, black) • Generate via absorption function of Zr a new transmitted spectrum (red) IMPRS Workshop 7/21/2009 9

  10. 3. Dose in SiO 2 • Dose measurement (diode)  power of spectrum (Si, blue) is known • Original spectrum (black)  Filter (Zr, red)  Absorption in SiO 2 (green) • Final dose is determined IMPRS Workshop 7/21/2009 10

  11. New dose rates New dose rates • Every dose rate matches to a specific set of parameters, let’s assume • U=60 kV (max. tube voltage) • I=33 mA (max. tube current) 1  • Distance is 155 mm ( ) D  2 r • Dose rate in silicon (300 µm) Gy   | 0 , 305 D 60 , 33 , 155 , 300 Si kv mA mm µm s • Dose rate in silicon dioxide (180 nm) Gy   | 0 , 239 D SiO 60 kV , 33 mA , 155 mm , 180 nm 2 s IMPRS Workshop 7/21/2009 7/21/2009 11 11

  12. Back to the DEPFET: Setup and DAQ Study with a 6x16 Minimatrix • • Important contacts on PCB  easy accessibility Drain contact needed probe • needle Several irradiation and • measurement steps Readout duration of input • characteristic of all 96 pixels ~ 6…7h • – min. 4 days of room temperature annealing DAQ via LabVIEW: Sweep of • Gate voltage, Drain current is measured IMPRS Workshop 7/21/2009 12

  13. Results of input characteristics √I DS (U GS ) IMPRS Workshop 7/21/2009 13

  14. Change of threshold voltage IMPRS Workshop 7/21/2009 14

  15. Change of gain g m • Input characteristic curve fitted with – I=aU²+bU+c – Gain g m =mU+b via dI/dU  m=2a  no numeric deviation – Gain evaluated at Drain current = 50 µA • Maybe effect is part of setup and readout process  needs to be rechecked to find out if effect still occurs IMPRS Workshop 7/21/2009 15

  16. Outlook & Conclusions • Dosimetry – A good agreement between simulated spectrum and dose measurement with silicon diode has been found – Further investigations are under way: Spectrum has been measured (matches fine to the simulated one, except low energy photons • and L-Lines of tungsten) • Dosimetry with RadFETs have been done • DEPFET – Shift of threshold voltage matches to previous single pixel measurements (strongly dependent on gate voltage) – Spreading of threshold voltage after irradiation increased threshold voltage shift ∆ = ±0,12 V unirradiated • threshold voltage shift ∆ = ± 0,20 V at 142 krad (1.42 kGy) • threshold voltage shift ∆ = ±0,22 V at 2.37 Mrad (23.65 kGy) • – Impact on Belle II • Steering chips (Switcher) won‘t have problems with the spreading • Change of g m interferes on g q (via gate oxide capacitance). Variations in Drain-Source current may affect the current readout chip. IMPRS Workshop 7/21/2009 16

  17. Backup Slides IMPRS Workshop 7/21/2009 17

  18. Change of threshold voltage vs. Dose IMPRS Workshop 7/21/2009 18

Download Presentation
Download Policy: The content available on the website is offered to you 'AS IS' for your personal information and use only. It cannot be commercialized, licensed, or distributed on other websites without prior consent from the author. To download a presentation, simply click this link. If you encounter any difficulties during the download process, it's possible that the publisher has removed the file from their server.

Recommend


More recommend