Dosimetry of an x-ray tube and Irradiation results of a DEPFET - - PowerPoint PPT Presentation
Dosimetry of an x-ray tube and Irradiation results of a DEPFET - - PowerPoint PPT Presentation
Dosimetry of an x-ray tube and Irradiation results of a DEPFET matrix Carried out at the University of Karlsruhe IMPRS Workshop Motivation: The KEKB accelerator and High luminosity collider Asymetric energy for studying
Motivation: The KEKB accelerator and…
- High luminosity collider
- Asymetric energy for
studying
- Measure CKM matrix
elements
- CP violation
Measuring device:
7/21/2009
IMPRS Workshop
2
B B S Y e e
) 4 (
…the Belle II detector
- The two innermost layers
- f the vertex detector will
be consisted of DEPFET matrices
- Ionizing Radiation from:
- Beam-Gas Interaction
- Synchrotron radiation
will damage those matrices
- Ca. 1 Mrad/a (=10 kGy/a)
- Simulating of these various
sources with x-ray radiation
7/21/2009
IMPRS Workshop
3
What are DEPFETs?
7/21/2009
IMPRS Workshop
4
- Depend on sideward
depletion: Internal Gate
- Generated charges will be
stored in the Internal Gate
- Modulation of Drain-
Source current
- Clear operation necessary
- Beneath Gate
metallization: Layer of silicon dioxide Sensitive to ionizing radiation
Damage mechanism
- Generataion of
electron/hole pairs in the
- xide
- Mobility of holes is weak in
comparison to electrons
- Holes will be trapped near
the interface and stay there for O(h) till O(a)
- Also: Increase of interface
state density
7/21/2009
IMPRS Workshop
5
Dosimetry of the x-ray facility in Karlsruhe
- To quantify the damage
done by ionizing radiation Applied Dose is necessary
- Calibration was given,
Recheck it Problem Challenge started …
7/21/2009
IMPRS Workshop
6
Recipe for succsesfull Calibration
- X-ray source
- Detector
a) Fast readout, highly sensitive, good spectral response, low noise, superb amplifier b) If not (a), then find a compromise of (a)
- Energy spectrum would be good
a) Measure one b) Simulate one
Thanks to my colleague: Oksana Brovchenko
7/21/2009
IMPRS Workshop
7
- 1. Determine dose rate in silicon diode
- Dose rate measurement via
depleted diode
- Measuring of reverse-current
- X-ray photons generate
electron/hole-pairs in Si- Bulk
– Every charge carrier pair represents an energy of 3.6 eV – With the x-ray generated current one gets the deposited power, with the mass of the diode dose rate in Si
7/21/2009
IMPRS Workshop
8
- 2. Making use of the spectrum
- Spectrum of tungsten anode (including a 0.4 mm
Be filtering, black)
- Generate via absorption function of Zr a new
transmitted spectrum (red)
7/21/2009
IMPRS Workshop
9
- 3. Dose in SiO2
- Dose measurement
(diode) power of spectrum (Si, blue) is known
- Original spectrum
(black) Filter (Zr, red) Absorption in SiO2 (green)
- Final dose is
determined
7/21/2009
IMPRS Workshop
10
New dose rates
- Every dose rate matches to a specific set of
parameters, let’s assume
- U=60 kV (max. tube voltage)
- I=33 mA (max. tube current)
- Distance is 155 mm ( )
- Dose rate in silicon (300 µm)
- Dose rate in silicon dioxide (180 nm)
7/21/2009
IMPRS Workshop
11
New dose rates
7/21/2009 11 2
1 r D
s Gy D
µm mm mA kv Si
305 , |
300 , 155 , 33 , 60
s Gy D
nm mm mA kV SiO
239 , |
180 , 155 , 33 , 60
2
Back to the DEPFET: Setup and DAQ
- Study with a 6x16 Minimatrix
- Important contacts on PCB
easy accessibility
- Drain contact needed probe
needle
- Several irradiation and
measurement steps
- Readout duration of input
characteristic of all 96 pixels
- ~ 6…7h
– min. 4 days of room temperature annealing
- DAQ via LabVIEW: Sweep of
Gate voltage, Drain current is measured
7/21/2009
IMPRS Workshop
12
Results of input characteristics √IDS(UGS)
7/21/2009
IMPRS Workshop
13
Change of threshold voltage
7/21/2009
IMPRS Workshop
14
Change of gain gm
- Input characteristic curve
fitted with
– I=aU²+bU+c – Gain gm=mU+b via dI/dU m=2a no numeric deviation – Gain evaluated at Drain current = 50 µA
- Maybe effect is part of
setup and readout process needs to be rechecked to find out if effect still
- ccurs
7/21/2009
IMPRS Workshop
15
Outlook & Conclusions
- Dosimetry
– A good agreement between simulated spectrum and dose measurement with silicon diode has been found – Further investigations are under way:
- Spectrum has been measured (matches fine to the simulated one, except low energy photons
and L-Lines of tungsten)
- Dosimetry with RadFETs have been done
- DEPFET
– Shift of threshold voltage matches to previous single pixel measurements (strongly dependent on gate voltage) – Spreading of threshold voltage after irradiation increased
- threshold voltage shift ∆ = ±0,12 V
unirradiated
- threshold voltage shift ∆ = ± 0,20 V
at 142 krad (1.42 kGy)
- threshold voltage shift ∆ = ±0,22 V
at 2.37 Mrad (23.65 kGy)
– Impact on Belle II
- Steering chips (Switcher) won‘t have problems with the spreading
- Change of gm interferes on gq (via gate oxide capacitance). Variations in Drain-Source current
may affect the current readout chip.
7/21/2009
IMPRS Workshop
16
Backup Slides
7/21/2009
IMPRS Workshop
17
Change of threshold voltage vs. Dose
7/21/2009
IMPRS Workshop
18