EYP 2006 EYP 2006
Differential Scanning Differential Scanning Calorimetry - - PowerPoint PPT Presentation
Differential Scanning Differential Scanning Calorimetry - - PowerPoint PPT Presentation
Differential Scanning Differential Scanning Calorimetry Calorimetry Cooking with Chemicals Cooking with Chemicals Clare Rawlinson Clare Rawlinson School of Pharmacy School of Pharmacy University of Bradford University of
EYP 2006 EYP 2006
Outline Outline
- Brief history of thermal analysis
Brief history of thermal analysis
- Theory of thermal analysis techniques
Theory of thermal analysis techniques
– – Thermal Gravimetric Analysis (TGA) Thermal Gravimetric Analysis (TGA) – – Differential Scanning Calorimetry (DSC) Differential Scanning Calorimetry (DSC)
- Generating valid data
Generating valid data
– – Calibration Calibration – – Sample preparation Sample preparation
- Interpreting data and Applications
Interpreting data and Applications
– – Real events Real events – – Artefacts Artefacts
- Recent advances
Recent advances
EYP 2006 EYP 2006
Calorimetry Calorimetry
- Calorimetry
Calorimetry
– – The study of heat transfer during The study of heat transfer during physical and chemical processes physical and chemical processes
- Calorimeter
Calorimeter
– – A device for measuring the heat A device for measuring the heat transferred transferred Lavoisier Lavoisier and and Laplace Laplace (1782 (1782-
- 1784):
1784):
- il was burned in a lamp (
- il was burned in a lamp (Fig 9
Fig 9) held in ) held in a bucket (Fig. 8) held in a wire mesh a bucket (Fig. 8) held in a wire mesh cage ( cage (f f) )
- surrounded by ice in spaces
surrounded by ice in spaces b b and and a a of
- f
the double walled container a foot in the double walled container a foot in diameter diameter
- lid (
lid (F F) was topped with ice, as was a ) was topped with ice, as was a mesh lid (not shown) beneath it that mesh lid (not shown) beneath it that covered the inner volume covered the inner volume b b
EYP 2006 EYP 2006
Oil lamps to Guinea Pigs Oil lamps to Guinea Pigs… …
- Measured heat production of
Measured heat production of the metabolic processes in the metabolic processes in the ice bath calorimeter the ice bath calorimeter
- Outer
jacket prevented Outer jacket prevented conduction of heat from the conduction of heat from the external environment which external environment which would have also melted the would have also melted the ice ice
- From latent heat of fusion for
From latent heat of fusion for ice (334 J/gram ice at 0 ice (334 J/gram ice at 0 º ºC) C) Lavoisier Lavoisier converted the rate converted the rate
- f water formation to heat
- f water formation to heat
production production
- In 10 hours 370 grams of ice
In 10 hours 370 grams of ice melted melted Guinea pig produced 12,358 J per hour of heat Guinea pig produced 12,358 J per hour of heat (12.4 kJ/hr) (12.4 kJ/hr)
EYP 2006 EYP 2006
Basic Principles of Thermal Analysis
Modern instrumentation used for thermal analysis usually consists of four parts:
- sample/sample holder
- sensors to detect/measure a property of the sample and the
temperature
- an enclosure within which the experimental parameters may be
controlled
- a computer to control data collection and processing
EYP 2006 EYP 2006
TGA and DSC TGA and DSC
Thermogravimetric Analysis (TGA)
– – mass change of a substance measured as function of mass change of a substance measured as function of temperature whilst the substance is subjected to a controlled temperature whilst the substance is subjected to a controlled temperature programme temperature programme1
1
– – mass is lost if the substance contains a volatile fraction mass is lost if the substance contains a volatile fraction
Differential Scanning Calorimetry (DSC)
– – provides information about thermal changes that do not involve a provides information about thermal changes that do not involve a change in sample mass change in sample mass1
1
– – more commonly used technique than TGA more commonly used technique than TGA – Two basic types of DSC instruments: heat-flux and power compensation
1Haines, P. J. (2002) The Royal Society of Chemistry, Cambridge.
EYP 2006 EYP 2006
Sample holder :
sample and reference are connected by a low-resistance heat flow path Aluminium, stainless, platinum sample pans
Sensors: Sensors:
- temperature sensors
temperature sensors
- usually thermocouples
usually thermocouples
Furnace:
- ne block for both sample
and reference cells
Temperature controller:
- temperature difference between the sample and reference is
measured
Heat Flux DSC
sample pan inert gas vacuum heating coil reference pan thermocouples
EYP 2006 EYP 2006
Sample holder Sample holder :
:
- Aluminium
Aluminium, platinum, stainless steel pans , platinum, stainless steel pans
Sensors: Sensors:
- Pt resistance
Pt resistance thermocouples. thermocouples.
- Separate sensors
Separate sensors and heaters for the and heaters for the sample and reference sample and reference
Furnace: Furnace:
- separate blocks for sample and reference cells
separate blocks for sample and reference cells
Temperature controller: Temperature controller:
- differential thermal power is supplied to the heaters to mainta
differential thermal power is supplied to the heaters to maintain the in the temperature of the sample and reference at the program value temperature of the sample and reference at the program value
sample pan ΔT = 0 inert gas vacuum inert gas vacuum
individual heaters
reference pan thermocouple
Power Compensated DSC Power Compensated DSC
EYP 2006 EYP 2006
Outline Outline
- Brief history of thermal analysis
Brief history of thermal analysis
- Theory of thermal analysis techniques
Theory of thermal analysis techniques
– – Thermal Gravimetric Analysis (TGA) Thermal Gravimetric Analysis (TGA) – – Differential Scanning Calorimetry (DSC) Differential Scanning Calorimetry (DSC)
- Generating valid data
Generating valid data
– – Calibration Calibration – – Sample preparation Sample preparation
- Interpreting data and Applications
Interpreting data and Applications
– – Real events Real events – – Artefacts Artefacts
- Recent advances
Recent advances
EYP 2006 EYP 2006
DSC Calibration
Baseline Calibration
evaluation of the thermal resistance of the sample and reference sensors measurements
- ver the temperature
range
- f
interest
EYP 2006 EYP 2006
DSC Calibration
Temperature
- match melting onset temperatures to the known melting points of
standards analyzed by DSC
- should be calibrated as close to desired temperature range as possible
Heat flow
- use calibration standards of known heat capacity, slow accurate heating
rates (0.5–2.0 °C/min), and similar sample and reference pan weights
calibrants
- high purity
- accurately known enthalpies
- thermally stable
- light stable
- not hygroscopic
- do not react (pan, atmosphere)
metals
- Indium 156.6 °C; 28.45 J/g
- Zinc 419.47°C, 108.17 J/g
inorganics
- KNO3 128.7 °C
- KClO4 299.4 °C
- rganics
- polystyrene 105 °C
- benzoic acid 122.3 °C; 147.3 J/g
EYP 2006 EYP 2006
Sample Preparation
accurately-weighed samples (~3-20 mg, usually 3-5 mg for simple powders) small sample pans (0.1 mL) of inert or treated metals (Al, Pt, stainless) several pan configurations, e.g., open , pinhole, or hermetically-sealed pans same material and configuration should be used for the sample and the reference material should completely cover the bottom of the pan to ensure good thermal contact avoid overfilling the pan to minimize thermal lag from the bulk of the material to the sensor
Al Pt alumina Ni Cu quartz
* small sample masses and low heating rates increase resolution, but at the expense
- f
sensitivity
EYP 2006 EYP 2006
Purge Gases Purge Gases
- Sample may react with air
Sample may react with air -
- oxidising or burning
- xidising or burning
- Control moisture content of atmosphere
Control moisture content of atmosphere
- Use inert gas e.g. nitrogen or argon
Use inert gas e.g. nitrogen or argon
- Flowing purge gas
Flowing purge gas
- In some cases deliberately choose reactive gas, e.g.
In some cases deliberately choose reactive gas, e.g.
– – hydrogen to reduce an oxide to metal hydrogen to reduce an oxide to metal – – carbon dioxide which affects decomposition of metal carbonate carbon dioxide which affects decomposition of metal carbonate
- Removes waste products from sublimation or
Removes waste products from sublimation or decomposition decomposition
EYP 2006 EYP 2006
Outline Outline
- Brief history of thermal analysis
Brief history of thermal analysis
- Theory of thermal analysis techniques
Theory of thermal analysis techniques
– – Thermal Gravimetric Analysis (TGA) Thermal Gravimetric Analysis (TGA) – – Differential Scanning Calorimetry (DSC) Differential Scanning Calorimetry (DSC)
- Generating valid data
Generating valid data
– – Calibration Calibration – – Sample preparation Sample preparation
- Interpreting data and Applications
Interpreting data and Applications
– – Real events Real events – – Artefacts Artefacts
- Recent advances
Recent advances
EYP 2006 EYP 2006
Typical Features of a DSC Trace (Polymorphic System)
sulphapyridine
endothermic events melting sublimation solid-solid transitions desolvation chemical reactions exothermic events crystallization solid-solid transitions decomposition chemical reactions baseline shifts glass transition Exo
EYP 2006 EYP 2006
Melting Processes by DSC
Pure substances
- linear melting curve
- melting
point defined by
- nset
temperature
eutectic melt
Melting with decomposition
- exothermic
- endothermic
Impure substances
- Broad,
asymmetric melting peak
- melting
characterized at peak maxima
- eutectic impurities
may produce a second peak
EYP 2006 EYP 2006
Definition of Transition Temperature
157.81°C 156.50°C 28.87J/g
- 2.5
- 2.0
- 1.5
- 1.0
- 0.5
0.0 0.5
Heat Flow (W/g)
140 145 150 155 160 165 170 175
Temperature (°C)
Exo Up Universal V3.3B TA Instruments
extrapolated
- nset temperature
peak melting temperature
Exo
EYP 2006 EYP 2006
Enthalpy of Fusion
157.81°C 156.50°C 28.87J/g
- 2.5
- 2.0
- 1.5
- 1.0
- 0.5
0.0 0.5
Heat Flow (W/g)
140 145 150 155 160 165 170 175
Temperature (°C)
Exo Up Universal V3.3B TA Instruments
Exo
EYP 2006 EYP 2006
Enthalpy of Fusion by DSC
More difficult where multiple thermal events leading to stable melt
e.g. solid-solid transitions (A to B) before melt, or where melt / recrystallisation before melt Estimate from sum all areas
For a single (well-defined) melting endotherm area under peak minimal decomposition/sublimation readily measured for high melting polymorph can be measured for low melting polymorph Endo
EYP 2006 EYP 2006
Purity by DSC
1-3 mg samples in hermetically- sealed pans are recommended Peak width a valuable measure
- f purity:
impurities lower the melting point Less pure (non-perfect) crystals melt first followed by purer larger crystals polymorphism interferes with purity determination, especially when a transition occurs in the middle of the melting peak Accurate measurement of ΔHf needs pure samples of polymorphs
benzoic acid
Plato, C.; Glasgow, Jr., A.R. Anal. Chem., 1969, 41(2), 330-336.
Exo
97% 99% 99.9%
EYP 2006 EYP 2006
Glass Transitions
transition from a disordered solid to a liquid appears as a step (endothermic direction) in the DSC curve gradual enthalpy change may occur, producing an endothermic peak superimposed on the glass transition characterized by change in heat capacity (no heat absorbed or evolved) Exo
EYP 2006 EYP 2006
Effect of Heating Rate
many transitions (evaporation, crystallization, decomposition, etc.) are kinetic so shift to higher temp. when heated at a higher rate increasing the scanning rate increases sensitivity, while decreasing the scanning rate increases resolution to obtain thermal event temperatures close to the true thermodynamic value, slow scanning rates (e.g., 1–5 K/min) should be used Rapid scanning can obscure thermal events Advantageous in fast scan DSC, e.g. 500K/min
EYP 2006 EYP 2006
Recognizing Artefacts
Sample movement in pan cool air entry into cell sample pan distortion Pan moves in furnace mechanical shock / knock bench electrical effects, power spikes, etc. atmosphere changes burst of pan lid Closing /
- pening pan
hole, e.g. sublimation sensor contamination
EYP 2006 EYP 2006
Ensuring correct interpretation of DSC Ensuring correct interpretation of DSC
- You can
You can’ ’t t
- Can minimise misinterpretation
Can minimise misinterpretation
- Essential to have valid data to interpret
Essential to have valid data to interpret
– – Calibration, reproducible data, appropriate sampling etc Calibration, reproducible data, appropriate sampling etc
- Kinetics / thermodynamics at elevated temps
Kinetics / thermodynamics at elevated temps
– – High temp can speed kinetics High temp can speed kinetics – – event would happen at room event would happen at room temperature but slowly temperature but slowly – – Effect activated by increased temp (overcome activation energy) Effect activated by increased temp (overcome activation energy)
- event would not happen at room temperature
event would not happen at room temperature
- DSC shows excipients interact at 120
DSC shows excipients interact at 120º ºC C
– – Does not necessarily show interaction at room temp Does not necessarily show interaction at room temp
EYP 2006 EYP 2006
Polymorph Screening and Indentification
thermal stability
– melting – crystallization – solid-state transformations – desolvation – glass transition – sublimation – decomposition
heat flow
– heat of fusion – heat of transition – heat capacity
mixture analysis
– physical purity (crystal forms, crystallinity) – chemical purity
- phase diagrams / interactions
–––––– Form I –––––– Form II –––––– Variable Hydrate –––––– Dihydrate –––––– Acetic acid solvate Form III Form I Form II –––––– Form I –––––– Form II –––––– Variable Hydrate –––––– Dihydrate –––––– Acetic acid solvate Form III Form I Form II
- 2.5
- 2.0
- 1.5
- 1.0
- 0.5
0.0 0.5 1.0
Heat Flow (W/g)
50 100 150 200 250 300 350
Temperature (°C)
––––––– Form I ––––––– Form II ––––––– Variable Hydrate ––––––– Dihydrate ––––––– Acetic acid solvate Exo Up Form III Form I Form II
- 2.5
- 2.0
- 1.5
- 1.0
- 0.5
0.0 0.5 1.0
Heat Flow (W/g)
50 100 150 200 250 300 350
Temperature (°C)
––––––– Form I ––––––– Form II ––––––– Variable Hydrate ––––––– Dihydrate ––––––– Acetic acid solvate Exo Up Form III Form I Form II
Exo
EYP 2006 EYP 2006
Effect of Phase Impurities
Lot A: pure low melting polymorph – melting observed Lot B: seeds of high melting polymorph induce solid-state transition below the melting temperature of the low melting polymorph
2046742 FILE# 022511DSC.1 2046742 FILE# 022458 DSC.1 Form II ?
- 5
- 4
- 3
- 2
- 1
Heat Flow (W/g)
80 130 180 230 280
Temperature (°C)
Exo Up Universal V3.3B TA Instruments
Lot A - pure Lot B - seeds
lots A & B of polymorph (identical by XRD) are different by DSC:
EYP 2006 EYP 2006
Outline Outline
- Brief history of thermal analysis
Brief history of thermal analysis
- Theory of thermal analysis techniques
Theory of thermal analysis techniques
– – Thermal Gravimetric Analysis (TGA) Thermal Gravimetric Analysis (TGA) – – Differential Scanning Calorimetry (DSC) Differential Scanning Calorimetry (DSC)
- Generating valid data
Generating valid data
– – Calibration Calibration – – Sample preparation Sample preparation
- Interpreting data and Applications
Interpreting data and Applications
– – Real events Real events – – Artefacts Artefacts
- Recent advances
Recent advances
EYP 2006 EYP 2006
Microcalorimetry Microcalorimetry
- High sensitivity DSC
High sensitivity DSC
- Solutions
Solutions
- Scan range typically
Scan range typically 0-
- 120
120 ° °C C
- Scanning rate of 0
Scanning rate of 0-
- 120
120 ° °C/hr C/hr
- Reverse scan rate 0
Reverse scan rate 0-
- 45
45 ° °C/hr C/hr (depending on efficiency (depending on efficiency
- f cooling tank)
- f cooling tank)
- Useful for looking at low
Useful for looking at low energy modifications energy modifications
- e.g. protein relaxation and
e.g. protein relaxation and refolding, polymer refolding, polymer characterisation characterisation
trehlose
EYP 2006 EYP 2006
Modulated DSC Heating Profile
Modulated DSC (MDSC)
introduced in 1993; “heat flux” design sinusoidal (or square-wave or sawtooth) modulation is superimposed on the underlying heating ramp total heat flow signal contains all
- f the thermal transitions of
standard DSC Fourier Transformation analysis is used to separate the total heat flow into its two components: reversing and non-reversing heat flow increased sensitivity, resolution and the ability to separate multiple thermal events
EYP 2006 EYP 2006
Heat capacity (reversing heat flow) glass transition melting
MDSC for Polymorph Characterization
Reversing (heat flow component)- 0.50
- 0.45
- 0.40
- 0.35
- 0.30
- 0.25
- 0.20
- 0.15
- 0.10
- 0.05
0.00
Rev Heat Flow (W/g)
110 120 130 140 150 160 170 180
Temperature (°C)
DSC010622b.1 483518 HCL (POLYMORPH 1) DSC010622d.1 483518 HCL Exo Up Universal V3.3B TA Non-reversing (heat flow component)- 0.8
- 0.6
- 0.4
- 0.2
0.0 0.2
Nonrev Heat Flow (W/g)
110 120 130 140 150 160 170 180
Temperature (°C)
DSC010622b.1 483518 HCL (POLYMORPH 1) DSC010622d.1 483518 HCL Exo Up Universal V3.3B TALot A Lot B Lot A Lot B reversing heat flow non-reversing heat flow
Kinetic (non-reversing heat flow) crystallization decomposition evaporation
EYP 2006 EYP 2006
‘ ‘Hyper Hyper’ ’ DSC DSC
- Fast scanning DSC
Fast scanning DSC
- Only possible with power compensated
Only possible with power compensated
- Normal equipment
Normal equipment ≈ ≈ 100 100 º ºC/min C/min
- Specialised up to 500
Specialised up to 500 º ºC/min C/min
- Increased sensitivity, loss of resolution
Increased sensitivity, loss of resolution
- e.g. amorphous content in mainly crystalline sample
e.g. amorphous content in mainly crystalline sample
– change of specific heat at Tg Tg is linear relationship to the amorphous content – – Conventional DSC 10% amorphous limit of detection Conventional DSC 10% amorphous limit of detection – – Hyper DSC Hyper DSC <1% amorphous easily detected <1% amorphous easily detected Lappalainen Lappalainen, M., I. , M., I. Pitkanen Pitkanen, et al. (2006). , et al. (2006). International Journal of International Journal of Pharmaceutics Pharmaceutics 307 307(2): 150 (2): 150-
- 155.
155.
EYP 2006 EYP 2006
Best Practices for Thermal Analysis
proper instrument calibration use purge gas (N2 or He) to remove corrosive off-gases small sample size good thermal contact between the sample and the temperature-sensing device proper sample encapsulation start temperature well below expected transition temperature slow scanning speeds
(Unless aiming to obscure thermal transitions, e.g fast scan DSC)
avoid decomposition in the DSC
(Run TGA first – its easier to clean up!)
EYP 2006 EYP 2006
Caution Caution… …
- It is a bulk tool
It is a bulk tool
– – Analysing the gross average of events in a sample Analysing the gross average of events in a sample – – Conversely, small powder sample in DSC may not Conversely, small powder sample in DSC may not represent packing of powder bulk in represent packing of powder bulk in decomposition studies decomposition studies
- Instrument error in DSC typically
Instrument error in DSC typically ± ± 0.5 0.5 -
- 1
1º ºC C
- In Scanning modes, thermal events may be
In Scanning modes, thermal events may be “ “smeared smeared” ” by a thermal lag by a thermal lag – – Sample temperature not keeping up with Sample temperature not keeping up with instrument instrument – – Bigger effect at higher heating rates Bigger effect at higher heating rates – – Typically 1 Typically 1º ºC at 10 C at 10º ºC/min C/min
EYP 2006 EYP 2006
And more caution! And more caution!
- Thermal analysis tells you what is happening at
Thermal analysis tells you what is happening at the temperature it happens at! the temperature it happens at! – – Care when extrapolating to room temperature Care when extrapolating to room temperature stability / interaction stability / interaction
- Don
Don’ ’t over t over-
- interpret data
interpret data
- Care when using thermal analysis in isolation
Care when using thermal analysis in isolation
- Artefacts / heating rate effects etc
Artefacts / heating rate effects etc
- Couple with other analytical tools
Couple with other analytical tools – – Heated X Heated X-
- ray, heated vibrational
ray, heated vibrational spectroscopy, hot stage microscope spectroscopy, hot stage microscope
EYP 2006 EYP 2006
Acknowledgements Acknowledgements
Professor Adrian Williams, University of Reading Professor Adrian Williams, University of Reading Dr Ian Grimsey, University of Bradford Dr Ian Grimsey, University of Bradford Dr Peter Timmins, Bristol Myers Squibb Dr Peter Timmins, Bristol Myers Squibb Dr Wendy Dr Wendy Hulse Hulse, University of Bradford , University of Bradford Luciana Luciana DeMatos DeMatos, University of Sheffield , University of Sheffield
EYP 2006 EYP 2006
Questions Questions
EYP 2006 EYP 2006
Reversing and Non-Reversing Contributions to Total DSC Heat Flow
dQ/dt = Cp . dT/dt + f(t,T)
total heat flow resulting from average heating rate reversing signal heat flow resulting from sinusoidal temperature modulation (heat capacity component) non-reversing signal (kinetic component)
e.g. see Pharmaceutical Research: 17 (6): 696-700, June 2000 Craig, DQM et al.
EYP 2006 EYP 2006
Some Common Thermal Analysis Techniques
Differential Thermal Analysis (DTA)
- the temperature difference between a sample and an inert reference material, ΔT = TS -
TR, is measured as both are subjected to identical heat treatments
Differential Scanning Calorimetry (DSC)
- the sample and reference are maintained at the same temperature, even during a
thermal event (in the sample)
- the energy required to maintain zero temperature differential between the sample and
the reference, dΔq/dt, is measured
Isothermal titration calorimetry (ITC)
- The temperature of a “reaction” is kept constant whilst the energy change is measured
Thermogravimetric Analysis (TGA)
- the change in mass of a sample on heating is measured
There are various techniques in which a physical property is measured as a function of temperature, while the sample is subjected to a predefined heating or cooling program.
EYP 2006 EYP 2006
Thermogravimetric Analysis (TGA)
- thermobalance
to monitor sample weight as a function
- f temperature
- weight
calibration using known weights
- temperature
calibration based
- n
ferromagnetic transition
- f
Curie point standards (e.g., Ni)
- larger sample masses, lower
temperature gradients, and higher purge rates minimize undesirable buoyancy effects
12.15% 19.32% 29.99% 20 40 60 80 100 120Weight (%)
20 40 60 80 100 120 140 160Time (min)
Universal V3.7A TA InstrumentsEYP 2006 EYP 2006
Differential Thermal Analysis
Sample holder: Sample and reference cells Sensors: Thermocouples, one for the sample and one for the reference Furnace: Block containing sample and reference cells Temperature controller: Controls temperature program Advantages:
- instruments can be used at very high
temperatures
- instruments are highly sensitive
- flexibility in sample volume/form
- characteristic transition or reaction
temperatures can be determined
Disadvantages:
uncertainty of heats of
fusion and transition temperatures
EYP 2006 EYP 2006
- development of “hyphenated” techniques for simultaneous analysis
TG-DTA TG-DSC TG-FTIR TG-MS
15.55% (0.9513mg) 24.80°C 100.0% 179.95°C 84.45%
- 1.8
- 0.8
0.2 1.2 2.2 3.2 4.2
Temperature Difference (µV/mg)
- 40
40 80 120
Weight (%)
20 70 120 170 220 270
Temperature (°C)
Exo Up Universal V3.3B TA InstrumentsTG-DTA trace of sodium tartrate
“Hyphenated” Techniques
- thermal techniques alone are insufficient to prove the existence of polymorphs
and solvates
- other techniques should be used, e.g., microscopy, diffraction, and spectroscopy