Stereotactic Breast Brachytherapy Applied Peripherally Using a Compression Applicator and HDR 192 Ir Mark J Rivard, PhD, 1 Raymond J Bricault Jr., MSc, 2 Christopher S Melhus, MSc, 1 Jessica R Hiatt, MSc, 3 Piran Sioshansi, PhD, 2 and David E


slide-1
SLIDE 1

s

Mark J Rivard, PhD,1 Raymond J Bricault Jr., MSc,2 Christopher S Melhus, MSc,1 Jessica R Hiatt, MSc,3 Piran Sioshansi, PhD,2 and David E Wazer, MD,1,3

1 Tufts-New England Medical Center, Boston, MA 2 Advanced Radiation Therapy, LLC, Billerica, MA

3 Rhode Island Hospital, Providence, RI

Stereotactic Breast Brachytherapy Applied Peripherally Using a Compression Applicator and HDR 192Ir

slide-2
SLIDE 2

Full Disclosure

Rivard & Wazer are medical advisors to ART Bricault & Sioshansi are officers of ART AccuBoost is FDA approved only for breast brachytherapy

slide-3
SLIDE 3

Treatment Process Overview

slide-4
SLIDE 4

AccuBoost System: Equipment Integration

mammography unit W-collimated applicators computed radiography

system

slide-5
SLIDE 5

Stereotactic Brachytherapy Image Guidance

before alignment after alignment

slide-6
SLIDE 6

Applicator: Mammography Paddle Interface

compression paddle applicator alignment

slide-7
SLIDE 7

AccuBoost Applicator Schematic

slide-8
SLIDE 8

AccuBoost Irradiation Beam

slide-9
SLIDE 9

Ionization Chamber & Film Measurements

slide-10
SLIDE 10

Central Plane Dose: 5 cm Applicator

slide-11
SLIDE 11

Central Plane Dose: 6 cm Applicator

slide-12
SLIDE 12

Central Plane Dose: 7 cm Applicator

slide-13
SLIDE 13

Depth Dose Profile: 6 cm Applicator

slide-14
SLIDE 14

Monte Carlo Simulation Geometry

cross-sectional view planar view through HDR dwells

slide-15
SLIDE 15

Calculated AccuBoost Dose Profiles

slide-16
SLIDE 16

Calculated Skin / Center Dose Ratio

slide-17
SLIDE 17

Dosimetric Comparison

slide-18
SLIDE 18

Summary

  • AccuBoost system integrates

– mammography – computed radiography – externally-collimated HDR 192Ir source

for stereotactic breast brachytherapy

  • measured & simulated dosimetry: good agreement
  • forthcoming multi-institutional clinical use