11/3/2014 1
High Performance Liquid Chromatography
Print version
Updated: 3 November 2014
David Reckhow CEE 772 # 18 1 David Reckhow CEE 772 # 18 2
High Performance Liquid Chromatography David Reckhow CEE 772 # 18 - - PDF document
11/3/2014 Updated: 3 November 2014 Print version High Performance Liquid Chromatography David Reckhow CEE 772 # 18 1 HPLC System David Reckhow CEE 772 # 18 2 1 11/3/2014 Instrument Basics INJECTION PUMP POINT DETECTOR COLUMN
Updated: 3 November 2014
David Reckhow CEE 772 # 18 1 David Reckhow CEE 772 # 18 2
MOBILE PHASE PUMP INJECTION POINT RECORDER DETECTOR COLLECTOR COLUMN
David Reckhow CEE 772 # 18 3
Adsorption
Normal Phase – polar bed, non polar mobile
phase (n-hexane, tetrahydrofuran)
Reverse Phase – non-polar bed w/ polar mobile
phase (methanol, water, acetonitrile mixture)
* most common
Ion Exchange
Stationary bed ionically charged surface,
Use with ionic or ionizable samples Stronger charge = longer elution time Mobile Phase – aqueous buffer
Size Exclusion
Column material precise pore sizes Large molecules first, then small
David Reckhow CEE 772 # 18 4
David Reckhow CEE 772 # 18 5
David Reckhow CEE 772 # 18 6
Pumps solvent through stationary phase
Smaller packing requires higher pressure by
pump
Larger packing and lower pump pressure is
usable for most procedures, except SEC
Stable flow rate - (not affected by pump)
0.01-10 mL/min Normal flow rate stability < 1 % Max psi 5000
Pump should be inert to solvents, buffer
Stainless steel; titanium; resistant minerals
(sapphires and ruby); PTFE (Teflon)
David Reckhow CEE 772 # 18 7
I. Constant Pressure
a) Pressurized coil b) Pressure intensifier
II. Constant Flow Pump
a) Piston *** most widely used b) Syringe
Modern pumps are highly efficient and
David Reckhow CEE 772 # 18 8
In-line metal coil system
Reduces pulse to +/- 3 % Low cost, possible contamination Limited range +/- 50-100 psi
T-type
flow does not pass through coil < 0.1 % pulse reduction Same limitations as above
Bellows, Spring Loaded
best but most expensive
David Reckhow CEE 772 # 18 9
David Reckhow CEE 772 # 18 10
UV Detector
Substances that absorb light from 180 to 350 nm 254 nm common General detector, most organic compounds Good for non UV absorbing solvents
Fluorescence
very sensitive to a few analytes which do fluoresce
(phenanthrene)
Derivative methods to attach ‘fluorophores’ to analytes Excitation at 280-305 nm and emission at 340-500 nm
Refractive Index Electrochemical Conductivity
David Reckhow CEE 772 # 18 11
David Reckhow CEE 772 # 18 12
Filtered using 0.2 μm filter Extends pump life Protects column from clogs
Displacement w/ less soluble gas Vacuum application Heat solvent
David Reckhow CEE 772 # 18 13 David Reckhow CEE 772 # 18 14
Isocratic elution
Gradient elution:
David Reckhow CEE 772 # 18 15
Isocr
tic c Separa ration tion (B : : Ac Acet etoni
rile) le) Gradient ent Separation tion
Condit Conditions: ions:
Column : : 0.46 0.46 * * 25cm 25cm Hypers Hypersil ODS ODS
rate : : 1.0 0 mL/min mL/min
nt : Aqu Aqueous B us Buffer ( (pH 3.5) a and Ac Acet etonitril rile (1) (1) benzyl alcohol, (2) benzyl alcohol, (2) Phenol Phenol (3) (3) 3’, 3’, 4’- 4’- dimetho hoxy-tolu
(4) (4) benzoin benzoin (5) (5) ethyl l benzoa benzoate, (6) (6) toluen luene (7) (7) 2, 2, 6 6 -dime
luen ene, e, (8) (8) o-
metho thoxybip iphenyl l
David Reckhow CEE 772 # 18 16
Stainless steel Common sizes:
10,12.5, 15, 25 cm long 4.6 mm i.d.
Length for optimum separation dictated
Filled with stationary phase material
David Reckhow CEE 772 # 18 17
Silica gel :
polymer composed of tetrahedral silicon atoms
Spherical (superior, more expensive)
Particle size and shape, surface area, and pore
Also, pH of gel surface, # active silanol groups,
David Reckhow CEE 772 # 18 18
David Reckhow CEE 772 # 18 19 David Reckhow CEE 772 # 18 20
David Reckhow CEE 772 # 18 21
Elution
Eluent strength
David Reckhow CEE 772 # 18 22
methanol-water or acetonitrile-water
Avoid to measure a sample that pH value is greater than 7.5 in a reversed –phase column, because of hydrolysis of the siloxane.
David Reckhow CEE 772 # 18 23
separation the components. In a reverse-phase column, the reverse is true
column, the reverse is true
David Reckhow CEE 772 # 18 24
particles compounds and ions compounds: precipitation upon contact with the stationary
compounds: co-elute and cause extraneous peaks and interfere with detection and/or quantification. Prolongs the life of the analytical column
David Reckhow CEE 772 # 18 25
each solute band spreads as it moves through the column the later eluting bands will spread more peak shape follow a Gaussian distribution
Baseline
David Reckhow CEE 772 # 18 26
The breadth of a Gaussian curve is directly related to the variance б2 or standard deviation б
David Reckhow CEE 772 # 18 27
Particle size of packings Column diameters Extra-column volume
Effect of mobile-phase flow rate
David Reckhow CEE 772 # 18 28
David Reckhow CEE 772 # 18 29
David Reckhow CEE 772 # 18 30
Extra-Column-Volume = sample volume + connecting tubing volume + fitting volume + detector cell volume
David Reckhow CEE 772 # 18 31
A minimum in H (or a maximum in efficiency) at low flow Particle size
Plate height
David Reckhow CEE 772 # 18 32
David Reckhow CEE 772 # 18 33
A properly packed HPLC column will give
Changes in either the physical or chemical
f
W0.05
David Reckhow CEE 772 # 18 34
Causes Cure Sample solvent stronger than the mobile phase Dissolve sample in mobile phase or at least reduce the strength of the sample solvent as much as possible Sample mass
Reduce the amount (mass) of sample injected. Silanol interaction with amines (affects late eluting peaks most)
25mM to 50mM recommended
10 mM TEA is usually sufficient.
See Figure 6 for a ranking of C18 phases by silanol activity Adsorption of acids
25 mM to 50 mM is usually sufficient
1% acetic acid or 0.1% TFA is usually sufficient Column void (affects early eluting peaks most)
David Reckhow CEE 772 # 18 35
Samples in solution Solutions must be filtered, centrifuged Some samples may need to be extracted
pH is important for ionized species
David Reckhow CEE 772 # 18 36
Evaluate the maximum wood sorption
Competitive sorption with metal and
Sorption and desorption under different pH
David Reckhow CEE 772 # 18 37
Sources:
nature: forest fire human behaviors: fuel burning, excess pesticide…
Characteristic
low solubility --- accumulation toxic, carcinogenic, mutagenic
Research interest Analysis
GC/FID: sensitivity but background interferences HPLC : necessary sensitivity and higher specificity
David Reckhow CEE 772 # 18 38
UV detector: MeOH : Water=90:10
y = 370607x - 3783.2 R2 = 0.9991 y = 253704x + 9291.7 R2 = 0.993 y = 139763x + 6567.8 R2 = 0.9831 50000 100000 150000 200000 250000 300000 350000 400000 0.2 0.4 0.6 0.8 1 1.2 In acetonitrile In Methanol In D.I. Water
David Reckhow CEE 772 # 18 39
UV and Fluo detector : in solution with 0.01 M CaCl2
and 200 mg/L NaN3,
MeOH : Water=90:10
C18
y = 2258129.5260 x - 21533.3953 R2 = 0.9990 y = 262017x - 819.82 R2 = 0.9999 200000 400000 600000 800000 1000000 1200000 1400000 1600000 1800000 0.2 0.4 0.6 0.8 1
FLUO UV
David Reckhow CEE 772 # 18 40
y = 2271926.286242 x - 21210.991782 R
2 = 0.998869
20000 40000 60000 80000 100000 120000 140000 160000 180000 200000 0.02 0.04 0.06 0.08 y = 1857707.241214x R2 = 0.999440 20000 40000 60000 80000 100000 120000 140000 160000 180000 0.02 0.04 0.06 0.08 0.1 y = 1866320.463236x - 518.501729 R2 = 0.999478
20000 40000 60000 80000 100000 120000 140000 160000 180000 0.02 0.04 0.06 0.08 0.1
Fluorescence detector:
David Reckhow CEE 772 # 18 41
To next lecture
David Reckhow CEE 772 # 18 42