Gas Flow Checking the Lotix Joy Osborne TOC Technical Product Manager Agenda Finding a Leak/Restriction Signs of a leak/Restriction Causes of Gas Flow Loss Causes of a Leak or Restriction Fittings - Overtighten/Loose Connection Furnace -


slide-1
SLIDE 1

Gas Flow Checking the Lotix

Joy Osborne TOC Technical Product Manager

slide-2
SLIDE 2

Agenda

Causes of Gas Flow Loss Signs of a leak/Restriction Finding a Leak/Restriction

slide-3
SLIDE 3

Causes of a Leak or Restriction

Fittings - Overtighten/Loose Connection Furnace - Heating Causes Expansion and Cooling Causing Contracting Solids - Particulates/ Salt/ Quartz Wool/ Catalyst in Pathway Random – Movement/ Vibrations/ Shipping

slide-4
SLIDE 4

When to Leak Check After Install Before and After Maintenance Poor Analytical Results Weekly

slide-5
SLIDE 5

Signs of a Leak

slide-6
SLIDE 6

Analytical Signs of a Leak

Low ABS Counts for Standards

<200 abs counts per 1ppmC Example: <200,000 abs counts for 1000ppmC

High %RSD

>5%

Low M Value in y=mx+b Equation for your Calibration Curve

0.5ml sample loop >200 0.3ml sample loop > 120

slide-7
SLIDE 7

Signs in a Report

Good Good Good

slide-8
SLIDE 8

Signs in a Report

slide-9
SLIDE 9

Flow Check

180- 220ml/min

Checkpoint 1: Valve 11 Com Port In

Should be within 10%

  • f

Checkpoint 1

Checkpoint 10: NDIR Bev A Line Out

slide-10
SLIDE 10

Step by Step Gas Flow Checking

slide-11
SLIDE 11

Tools Needed

Flow Meter Plugs Connectors Peek Nut Tighten Tool Tube Cutters

slide-12
SLIDE 12

Manual Exert

slide-13
SLIDE 13

Divide and Conquer

slide-14
SLIDE 14

Divide and Conquer – Gas Pathway in Ready Mode

Gas Source Pressure Regulator 3 Different Ts Flow Restrictor Valve 11 Sides of Perm Dryer Furnace Condensate Loop Mist Trap Valve 10 Perm Dryer Halogen Scrubber Sample Filter NDIR Detector Vent

slide-15
SLIDE 15

Before Checkpoint 1 ■ Common of valve 11 must be >180ml, min ■ Ideally >200ml/min ■ <200ml/min prior to the valve

  • f 11

Gas Source Pressure Regulator 3 Different Ts Flow Restrictor

slide-16
SLIDE 16

Divide and Conquer – Gas Pathway in Ready Mode

Gas Source Pressure Regulator 3 Different Ts Flow Restrictor Valve 11 Sides of Perm Dryer Furnace Condensate Loop Mist Trap Valve 10 Perm Dryer Halogen Scrubber Sample Filter NDIR Detector Vent

slide-17
SLIDE 17

Check Furnace Out

Valve 11 Sides of Perm Dryer Furnace

slide-18
SLIDE 18

Divide and Conquer – Gas Pathway in Ready Mode

Gas Source Pressure Regulator 3 Different Ts Flow Restrictor Valve 11 Sides of Perm Dryer Furnace Condensate Loop Mist Trap Valve 10 Perm Dryer Halogen Scrubber Sample Filter NDIR Detector Vent

slide-19
SLIDE 19

Side Arm of Mist Trap

Furnace Condensate Loop Mist Trap

slide-20
SLIDE 20

Divide and Conquer – Gas Pathway in Ready Mode

Gas Source Pressure Regulator 3 Different Ts Flow Restrictor Valve 11 Sides of Perm Dryer Furnace Condensate Loop Mist Trap Valve 10 Perm Dryer Halogen Scrubber Sample Filter NDIR Detector Vent

slide-21
SLIDE 21

Halogen Scrubber – Cu Side In

Mist Trap Valve 10 Perm Dryer

slide-22
SLIDE 22

Divide and Conquer – Gas Pathway in Ready Mode

Gas Source Pressure Regulator 3 Different Ts Flow Restrictor Valve 11 Sides of Perm Dryer Furnace Condensate Loop Mist Trap Valve 10 Perm Dryer Halogen Scrubber Sample Filter NDIR Detector Vent

slide-23
SLIDE 23

Divide and Conquer – Gas Pathway in Ready Mode

Gas Source Pressure Regulator 3 Different Ts Flow Restrictor Valve 11 Sides of Perm Dryer Furnace Condensate Loop Mist Trap Valve 10 Perm Dryer Halogen Scrubber Sample Filter NDIR Detector Vent

slide-24
SLIDE 24

Leak Free! ■ Thank you for listening!