Gas Flow Checking the Lotix Joy Osborne TOC Technical Product Manager Agenda Finding a Leak/Restriction Signs of a leak/Restriction Causes of Gas Flow Loss Causes of a Leak or Restriction Fittings - Overtighten/Loose Connection Furnace -
SLIDE 2
SLIDE 3
Causes of a Leak or Restriction
Fittings - Overtighten/Loose Connection Furnace - Heating Causes Expansion and Cooling Causing Contracting Solids - Particulates/ Salt/ Quartz Wool/ Catalyst in Pathway Random – Movement/ Vibrations/ Shipping
SLIDE 4
When to Leak Check After Install Before and After Maintenance Poor Analytical Results Weekly
SLIDE 5
Signs of a Leak
SLIDE 6
Analytical Signs of a Leak
Low ABS Counts for Standards
<200 abs counts per 1ppmC Example: <200,000 abs counts for 1000ppmC
High %RSD
>5%
Low M Value in y=mx+b Equation for your Calibration Curve
0.5ml sample loop >200 0.3ml sample loop > 120
SLIDE 7
Signs in a Report
Good Good Good
SLIDE 8
Signs in a Report
SLIDE 9
Flow Check
180- 220ml/min
Checkpoint 1: Valve 11 Com Port In
Should be within 10%
- f
Checkpoint 1
Checkpoint 10: NDIR Bev A Line Out
SLIDE 10
Step by Step Gas Flow Checking
SLIDE 11
Tools Needed
Flow Meter Plugs Connectors Peek Nut Tighten Tool Tube Cutters
SLIDE 12
Manual Exert
SLIDE 13
Divide and Conquer
SLIDE 14
Divide and Conquer – Gas Pathway in Ready Mode
Gas Source Pressure Regulator 3 Different Ts Flow Restrictor Valve 11 Sides of Perm Dryer Furnace Condensate Loop Mist Trap Valve 10 Perm Dryer Halogen Scrubber Sample Filter NDIR Detector Vent
SLIDE 15
Before Checkpoint 1 ■ Common of valve 11 must be >180ml, min ■ Ideally >200ml/min ■ <200ml/min prior to the valve
- f 11
Gas Source Pressure Regulator 3 Different Ts Flow Restrictor
SLIDE 16
Divide and Conquer – Gas Pathway in Ready Mode
Gas Source Pressure Regulator 3 Different Ts Flow Restrictor Valve 11 Sides of Perm Dryer Furnace Condensate Loop Mist Trap Valve 10 Perm Dryer Halogen Scrubber Sample Filter NDIR Detector Vent
SLIDE 17
Check Furnace Out
Valve 11 Sides of Perm Dryer Furnace
SLIDE 18
Divide and Conquer – Gas Pathway in Ready Mode
Gas Source Pressure Regulator 3 Different Ts Flow Restrictor Valve 11 Sides of Perm Dryer Furnace Condensate Loop Mist Trap Valve 10 Perm Dryer Halogen Scrubber Sample Filter NDIR Detector Vent
SLIDE 19
Side Arm of Mist Trap
Furnace Condensate Loop Mist Trap
SLIDE 20
Divide and Conquer – Gas Pathway in Ready Mode
Gas Source Pressure Regulator 3 Different Ts Flow Restrictor Valve 11 Sides of Perm Dryer Furnace Condensate Loop Mist Trap Valve 10 Perm Dryer Halogen Scrubber Sample Filter NDIR Detector Vent
SLIDE 21
Halogen Scrubber – Cu Side In
Mist Trap Valve 10 Perm Dryer
SLIDE 22
Divide and Conquer – Gas Pathway in Ready Mode
Gas Source Pressure Regulator 3 Different Ts Flow Restrictor Valve 11 Sides of Perm Dryer Furnace Condensate Loop Mist Trap Valve 10 Perm Dryer Halogen Scrubber Sample Filter NDIR Detector Vent
SLIDE 23
Divide and Conquer – Gas Pathway in Ready Mode
Gas Source Pressure Regulator 3 Different Ts Flow Restrictor Valve 11 Sides of Perm Dryer Furnace Condensate Loop Mist Trap Valve 10 Perm Dryer Halogen Scrubber Sample Filter NDIR Detector Vent