development of semiochemical slow release formulations
play

Development of semiochemical slow-release formulations as biological - PowerPoint PPT Presentation

Development of semiochemical slow-release formulations as biological control devices against aphids Dr. Stphanie Heuskin Post-doc 1. Laboratory of Analytical Chemistry, Gembloux Agro-Bio Tech-University of Lige, Belgium 2. Evolutionary


  1. Development of semiochemical slow-release formulations as biological control devices against aphids Dr. Stéphanie Heuskin Post-doc 1. Laboratory of Analytical Chemistry, Gembloux Agro-Bio Tech-University of Liège, Belgium 2. Evolutionary Ecology and Genetics Group, Earth and Life Institute, Catholic University of Louvain, Belgium

  2. SOLAPHID project (WALEO 2) Funding from the Belgium Walloon Region (2006-2011) “Biotechnologies related to the industrial production of insects used in biological control” 5 teams : chemistry – formulation – entomology – chemical ecology - industrial production

  3. Summary General introduction Objective The choice of semiochemicals and their origin How to analyse and quantify semiochemicals? How to purify semiochemicals? How to formulate semiochemicals? Is the formulation efficient? Conclusions and perspectives

  4. General introduction

  5. The aphid problem Damages to crops: virus and disease transmitter Economical and agricultural problem Pesticide control is limited - resistance of pest insects Biological control - non species-specific - unsafe for environment and human health

  6. Biological control as pest management strategy “The use of natural enemies to reduce the damage caused by a pest population” Attraction of aphid natural enemies

  7. Aphid tritrophic system 1 st level 3 rd level 2 nd level Host plant Aphids Predators Parasitoids  Chemical communication : semiochemicals

  8. Semiochemicals Plant – insect – insect chemical communication signals Pheromones Allelochemicals - alarm - allomones: + emitting species - sex - kairomones: +receptor species - aggregation - synomones: + emitting, + receptor - trail - host marking - … A same molecule can act as a pheromone and as an allelochemical substance

  9. Objective

  10. Global objective To develop natural semiochemical slow-release formulations as biological control devices attractive towards aphid natural enemies Which semiochemicals ? Which formulation ? Analysis and quantification ? Natural origin ? Efficiency ? Purification ? Release ? Attractiveness ?

  11. The choice of semiochemicals and their natural origin

  12. E- β -farnesene Sesquiterpene (C 15 H 24 ) - Aphid alarm pheromone 1 - Kairomone: attraction of aphid predators ( Episyrphus balteatus De Geer) 2-4 and aphid parasitoids ( Aphidius ervi Haliday) 5-6 1 Bowers et al., 1972 4 Verheggen et al., 2009 2 Francis et al., 2005 5 Du et al., 1998 3 Verheggen et al., 2008 6 Powell et al., 2003

  13. E- β -caryophyllene Sesquiterpene (C 15 H 24 ) - Reducer of aphid reproduction 1 - Attractive towards aphid parasitoids ( A. ervi Haliday) 2 1 Tomova et al., 2005 2 Sasso et al., 2009

  14. Natural matrix for sesquiterpenes  Essential oils - Matricaria chamomilla L. (Asteraceae): E- β - farnesene - Nepeta cataria L. (Lamiaceae): E- β - caryophyllene

  15. Essential oil characterisation : Gas chromatography Chromatography : a technique for separating the components of a mixture (liquid or gas) on the basis of differences in their affinity for a stationary (solid or liquid) and a mobile phase (liquid or gas) Gas chromatography - mixture : gas (headspace or vaporisation of a liquid) - stationary phase : liquid or polymer in capillary column - mobile phase : gas (inert carrier)

  16. Essential oil characterisation : Gas chromatography 4 Chromatogram 1 Most common : Mass 2 spectrometer, FID 3 He, H 2 or Program of T° : optimisation of the separation N 2 of the components of the mixture

  17. Gas chromatography : Ultra Fast GC >< Classic GC column Ultra Fast GC Classic GC

  18. Gas chromatography : Ultra Fast GC >< Classic GC - Ramp of T ° : 100 – 1200 ° C/min - Ramp of T ° : 10-30 ° C/min - Column : 2 – 5 m, 0.1 mm ID - Column : 10 – 30 m, 0.25-0.32 mm ID  Time for 1 analysis < 5 min  Time for 1 analysis > 35 min Ultra Fast GC Classic GC

  19. Essential oil characterisation Matricaria chamomilla L. (originated from Nepal) GC-MS (identification) N ° Major compounds Retention % index E- β -farnesene 1 1456 42.6 2 Germacrene D 1478 2.9 3 bicyclogermacrene 1494 1.9 (E,E)- α -farnesene 4 1506 8.3 α -bisabolol oxide B 5 1649 4.4 Fast GC-FID α -bisabolone oxide A 6 1673 4.5 (quantification ) 7 Chamazulene 1715 1.1 α -bisabolol oxide A 8 1735 21.1 9 Cis-ene-yne-dicycloether 1802 5.9 Heuskin S.et al ., 2009, J. Chrom. A, 1216, 2768-2775.

  20. Essential oil characterisation Nepeta cataria L. (originated from Canada) GC-MS 3 2 Retention N ° Major compounds % 1 index 4 1 (Z,E)-nepetalactone 1353 8.4 % 2 (E,Z)-nepetalactone 1377 22.5 % E- β -caryophyllene 3 1415 58.9 % Fast GC-FID 3 α -humulene 4 1465 3.9 % 2 1 4

  21. How to analyse and quantify semiochemicals ? Heuskin S.et al., 2009, J. Chrom. A, 1216, 2768-2775 Heuskin S.et al., 2010, J. Pharm. Biomed. Anal., 53, 962-972

  22. Quantification of semiochemicals: various steps 1. Quantification with internal standard 2. Optimisation of analytical method: resolution of compounds 3. Validation of analytical method: - calibration curve - evaluation of validation criteria according to norms

  23. Quantification with internal standard Why an internal standard? - to avoid the problem of variation of injected volume in GC with autosampler How to add an internal standard? - in reference solutions to construct calibration curve : the same concentration of IS in all the levels of concentration of analytes - in routine samples at a known concentration

  24. Quantification with internal standard Which internal standard? - Compound of the same family than the analytes - Retention time of IS close to the rentention time of analytes - Response factor close to 1: F = (Area A .Conc IS / Area IS .Conc A ) - Not naturally present in the routine sample  Here : IS = longifolene

  25. Optimisation of the analytical method Ultra Fast GC analysis IS  Good resolution of peaks in less than 5 min. R s = 2(t R E- β -caryophyllene – t R longifolene )/(W longifolene - W E- β -caryophyllene ) R s = 1,65 > 1,5  OK

  26. Analytical validation Objective of an analytical method for quantification : To be able to quantify the more precisely the routine samples x i ↔ µ T Results True value

  27. Analytical validation Objective of a validation : To give to the laboratory the garantees that the results are within acceptance limits │ x i - µ T │ < λ Bias λ = acceptance limits

  28. 5 concentrations Calibration curve * 3 replicates Calibration curve of E- β -farnesene Peak area of EBF / Peak area of IS 2,000 1,500 1,000 0,500 0,000 0,000 0,500 1,000 1,500 2,000 Concentration EBF / Concentration IS Blank y = 0,9558x + 0,0053 R² = 0,9999

  29. Analytical validations 1. « Classic » validation ISO 5725, GLP standard operating procedures : criteria validated 1 by 1 2. «Accuracy profile » validation Guidelines of the SFSTP* : Total error concept : combination of systematic and random errors Accuracy = Trueness + Precision *SFSTP= Société Française des Sciences et Techniques Pharmaceutiques

  30. 1. « Classic » validation Linearity > 0.996 < 2.75 90 <x< 110

  31. 1. « Classic » validation Precision of the method RSD % < 8% RSD % < 6% RSD % < 16% RSD % < 12%

  32. 2. Accuracy profile validation Trueness - Bias – Systematic error Precision – Repeatability + Intermediate precision – Random error Accuracy - Trueness + Precision – Total error

  33. λ = acceptance limits LLOQ HLOQ Accuracy profile

  34. Linearity of the method

  35. How to purify semiochemicals from essential oils ?

  36. Purification of components : chromatographic techniques Solid-Liquid chromatography Essential oil Solvent of elution Silica gel Semiochemical of interest 4. Collection of the 1. Essential oil in 2. Beginning of the semiochemical of the head of the elution with solvent 3. Elution process interest column

  37. Purification of components : chromatographic techniques Solid-Liquid chromatography - Mixture : liquid – essential oil - Stationary phase : solid – silicagel - Mobile phase : liquid – solvent of elution Goal : To obtain highly purified semiochemicals without solvent  Evaporation of solvent of elution

  38. Choice of the solvent of elution By thin layer chromatography  Choice of solvent based on : - Best separation of compounds on silica - Importance of solvent boiling point N-pentane (36°C) Standard of reference Essential oil

  39. Essential oil fractionation By liquid column chromatography Preliminary tests Small scale liquid column chromatography 1 ml essential oil deposited on 11 g dried silicagel Elution with n-pentane Collection of fractions (1.5 ml) Dilution Fast GC analysis

  40. Matricaria chamomilla fractionation 100 80 Percentage (%) 60 40 20 0 0 10 20 30 40 50 60 70 80 90 100 Elution volume (ml) E-β-farnesene α-farnesene Germacrene D Monoterpenes Chamazulene % E,E- α - % Elution volume (ml) % EBF % Germacrene D % chamazulene farnesene monoterpenes 0 - 10,5 (F0) 0 0 0 0 0 10,5 - 16,5 (F1) 0 0 0 100 0 16,5 - 22,5 (F2) 0 - 82 7,8 - 26 3 - 5 47 - 2 0 22,5 – 51 (F3) 86,3 - 76 4 - 1,4 5,7 - 22 0 0 51 – 72 (F4) 72 - 56 1,4 - 1,6 22 - 33 0 0 72 – 90 (F5) 55 - 33 1,6 33 - 41 0 0,5 - 16

Download Presentation
Download Policy: The content available on the website is offered to you 'AS IS' for your personal information and use only. It cannot be commercialized, licensed, or distributed on other websites without prior consent from the author. To download a presentation, simply click this link. If you encounter any difficulties during the download process, it's possible that the publisher has removed the file from their server.

Recommend


More recommend