Development of a QC method for assembly of heat conductive material - PowerPoint PPT Presentation

Development of a QC method for assembly of heat conductive material on new ATLAS pixel sensor module Yuji Onishi Yamanaka Lab Yuji Onishi 01 Year-End Presentation 2019/12/23(Mon) Yuji Onishi 2 5~10 times of LHC (250 fb -1 per year)


  1. Development of a QC method for assembly of heat conductive material on new ATLAS pixel sensor module Yuji Onishi Yamanaka Lab Yuji Onishi 01 Year-End Presentation 2019/12/23(Mon)

  2. Yuji Onishi 2 5~10 times of LHC 
 (250 fb -1 per year) 3000 fb -1 12 year after upgraded introduction • LHC will be upgraded to HL-LHC 
 • L arge H adron C ollider • proton - proton collision in √s = 13 TeV. • Luminosity ~ 10 34 cm -2 s -1 • Integrated luminosity ~ 200 fb -1 (until 2018, run2) 
 • H igh L uminosity LHC • Luminosity 
 • Integrated luminosity 


  3. Yuji Onishi 3 introduction Pixel detector For HL-LHC, we will upgrade the pixel detector. To upgrade, we have to do • ATLAS inner pixel detector will be upgraded for HL-LHC • mass production • QA/QC ( Q uality A ssurance/ Q uality C ontrol )

  4. Yuji Onishi sensor silicon we must detect those “thermal bad” we can not cool the sensor enough. Cooling pipe Thermal glue Cooling Cell FPC ASIC ASIC introduction 4 module. • in production of new pixel sensor module • Cooling Cell (carbon) will be loaded • Because front-end ASIC generate heat (10~20W). • Connected to Cooling pipe • with bad glueing, 


  5. Yuji Onishi 5 
 can know the behaviour of thermal bad. 
 
 introduction • how to detect “thermal bad” • candidate methods • leakage current • surface temperature gradient 
 • by comparing health module and intentional bad module, we • Now I am evaluating the sensitivity of these methods

  6. Yuji Onishi 6 temperature with health FE-I4 module 
 •use module which is intentionally failed to glue, 
 •bad contact between Cell and cooling pipe introduction • procedure • prepare temperature monitor system 
 • Now going on measuring the leakage current and surface • same measurement with intentional bad thermal conduction • exam) 


  7. Yuji Onishi Temperature monitor system calibration looks good. 
 into the climate chamber. and digital thermometer In calibration, I put 8 NTCs R is NTC resistance T is temperature, A,B,C is coefficient, 
 Steinhart-Hart equation temperature coefficient NTC means thermistor which has negative 
 • 7 (NTC6&7 is not important) • Temperature monitor • measure the NTC resistance • calibrate by using 


  8. Yuji Onishi 8 measuring setup ( 2 NTCs already implemented on module flex) digital thermometer & humidity sensor • setup to measure the leakage current & surface temperature • attach NTCs on surface of sensor module with thermal paste. • read the resistance of module NTCs with multimeter. 
 • with cables connected (HV, LV, LAN)

  9. Yuji Onishi changing climate chamber temperature 
 NTC4 NTC3 NTC2 NTC1 NTC0 w/ and w/o ASIC working. Sensor module measuring setup 9 • setup to measure the leakage current & surface temperature • purpose : • to know behaviour of below values at several temperature • leakage current • surface temperature 
 • To do so, • I measure these values 


  10. Yuji Onishi NTC5 is on readout PCB board it means all NTC measure the temperature of same temperature NTC0~4 is almost same surface temperature It looks good to measure NTC 6,7 is on air. • • NTC 0~4 is on module surface. • check NTC temperature without ASIC operation measuring setup 10 silicon plate • setup to measure the leakage current & surface temperature • To confirm temperature readout, 


  11. Yuji Onishi 11 measurement • Now going on measuring leakage current & surface temperature

  12. Yuji Onishi 12 bad” Temperature monitor system & measurement setup 
 Summary • I am working on development the method how to detect “thermal • I am trying two candidate way • leakage current & surface temperature gradient 
 • I prepared 
 • I am going on measuring leakage current & surface temperature

  13. Yuji Onishi 13 Prospects temperature with health module 
 •use module which is intentionally failed to glue, 
 •make bad thermal contact between Cell and cooling pipe • I will continue to measure the leakage current and surface • same measurement with intentional bad thermal conduction • exam) 


  14. Back up

  15. Yuji Onishi 15 (for humidity) NTC reference resister ~10kΩ 4.6kΩ monitor system ADC 5V GND • circuit

Download Presentation
Download Policy: The content available on the website is offered to you 'AS IS' for your personal information and use only. It cannot be commercialized, licensed, or distributed on other websites without prior consent from the author. To download a presentation, simply click this link. If you encounter any difficulties during the download process, it's possible that the publisher has removed the file from their server.