DETERMINING CALIBRATION INTERVALS BY AS-FOUND CALIBRATION A study - - PowerPoint PPT Presentation

determining calibration intervals by as found calibration
SMART_READER_LITE
LIVE PREVIEW

DETERMINING CALIBRATION INTERVALS BY AS-FOUND CALIBRATION A study - - PowerPoint PPT Presentation

DETERMINING CALIBRATION INTERVALS BY AS-FOUND CALIBRATION A study of high pressure turbinemeters by the Danish TSO Energinet Rune Hviid / Max Hansen 1 AGENDA 3 Cases for determining calibration interval Introduction to the Danish Gas


slide-1
SLIDE 1

DETERMINING CALIBRATION INTERVALS BY AS-FOUND CALIBRATION

A study of high pressure turbinemeters by the Danish TSO “Energinet”

1

Rune Hviid / Max Hansen

slide-2
SLIDE 2

AGENDA

Introduction to the Danish Gas network Calibration interval after: Recalibration, new meter of refurbishment of meter Case 1: Site conditions to determining calibration intervals. Case 2: Follow guidelines – Common for Danish gas companies Case 3: Study stability of meter - by recalibration to determine calibration intervals

2

3 Cases for determining calibration interval

  • 0,6
  • 0,4
  • 0,2
0,2 0,4 0,6 1 2 3 4 5 6 7 8 9

Absolute change in relative WME error

Service time (Years)

Evolution of metering error over time

83009883 83025024 83025992 80038506 83005897 83010724 83007604 83026450 83038508 80038504 80038505 80038510 80038511 80038512 83006011 83007531 83025990 83026451 83026452 83026453 83026455 Middelændring med fortegn Konf+ Konf- 74769
slide-3
SLIDE 3

THE DANISH GAS STEEL PIPELINE NETWORK

3

1 TS TSO – 80 80 Barg rg pipeli eline 3 DS DSO – 40 40 Barg rg pipeli eline

(DSO 4 barg plastic pipe network not shown)

42 Custody dy t trans nsfer p r points with pressure regulation

and metering (turbine)

slide-4
SLIDE 4

THE DANISH GAS STORAGE

4

Lille lle Torup – Ultrasonic meters Salt c cavern g rn gas storage Sten enli lill lle - Ultrasonic meters Au Auari rifer stora rage

slide-5
SLIDE 5

CASE 1: SITE CONDITIONS

Meter contamination

  • Coal dust - (from surface plants active coal filter)
  • Oil residue - (from underground previous storage use in Lille Torup gas storage)
  • Deposits on Ultrasonic sensors

Deterioration by operation conditions - Rapid pressure change

  • Compressor start
  • One-way valves
  • Damage to turbine blades

Calibration interval: Gas storage uses ultrasonic meters with 2 year due to heavy contamination

5

Heavy contamination – short calibration intervals

slide-6
SLIDE 6

CASE 2: FOLLOW GUIDELINES

Text from guidelines: “flowmeters are to be recalibrated with 5 year intervals”. “However, if calibration history study or test is available, year 10 interval is possible for new meters”.

6

Danish gas companies have common guidelines for calibration of flow meters.

Guidelines

Englis lish t transla latio ion § 18

  • 18. Network companies that uses measurements made in accordance with

section 1, § 1, section 3, is required, upon request, to inform the buyers of the measurement uncertainty of the amount of gas (energy) that is settled. The network companies concerned are also obliged to provide, at the request of the customers, the required documentation of the information provided by calibration certificates, cf. 3, as well as any additional calculations and statistical information to document the total measurement uncertainty.

  • Sec. 2. There must be a sound measurement technique for the information
  • provided. The measuring equipment in question must be calibrated at such a

frequency that there is a sufficient likelihood of compliance with determined or agreed measurement uncertainties during the period between the calibrations. Regardless of the calibration intervals determined, renewal calibration must be performed if the measuring equipment has been damaged or subjected to interference that affects its target accuracy.

  • Sec. 3. Calibration shall be performed as accredited calibration or foreign

calibration recognized in this country and shall be documented in calibration certificates with traceability to recognized international primary

LEGI GISLATUR URE

BEK number 1037 - 17/10/2006

slide-7
SLIDE 7

CASE 3: STUDY STABILITY OF METERS

Flowmeter guidelines allow extended calibration intervals if studies or tests of flow meters are made which prove what the metering error is below the allowable limit in time period of the extended calibration interval. 22 turbine meters were studies by Energinet by recalibration to show this in 2005, Last similar study was in 1984. Study 22 meters calibrated at 4500 euro, total 100.000 EURO Calibration interval extended from 5 to 8 years saving 30.000 EURO per year for all 84 meters (custody), in calibration alone, add on-top saving on fewer new meter purchaces, Study payback time 2 years.

7

Scope: Reevaluation of Calibration Interval (set in 1984 to 8 years)

slide-8
SLIDE 8

CASE 3: STUDY STABILITY OF METERS

8

Meter error in comissioning calibration and recalibration compaired

  • 1,5
  • 1
  • 0,5

0,5 1 20 40 60 80 100

Absolute change in relative error

Calibration flow in % of Qmax

Recalibration of 22 turbinemeters

83010724 83005897 80038506 83026450 83025992 83007604 80038508 83025990 83006011 83025024 83007531 80038504 80038505 83026451 83026452 83026455 83026453 80038511 80038510 80038512 83009883 Middelændring med fortegn 95% Konfidens 95% Konfidens 74769

slide-9
SLIDE 9

CASE 3: STUDY STABILITY OF METERS

9

Meter error in comissioning calibration and recalibration compairet

  • 0,6
  • 0,4
  • 0,2

0,2 0,4 0,6 1 2 3 4 5 6 7 8 9

Absolute change in relative WME error

Service time (Years)

Evolution of metering error over time

83009883 83025024 83025992 80038506 83005897 83010724 83007604 83026450 83038508 80038504 80038505 80038510 80038511 80038512 83006011 83007531 83025990 83026451 83026452 83026453 83026455 Middelændring med fortegn Konf+ Konf- 74769

slide-10
SLIDE 10

STUDY CONCLUSIONS

10

2005 study conclusions compaired with 1984 study

Calibration interval of 8 years is well choosen – change in spread of WME equal to 4-5 year of service life Change in WME over 8 years were found to be ~ 0,72% Allowable change is set to 0,9 % in guidelines Previous studies have shown a change of 0.03-0.08% per year [1,2] (0.64 per 8 years)

1. DS/EN 1776:1999: Gasforsyningssystemer – Målestationer for naturgas –Funktionskrav. 2. Time dependent performance of turbine gas meters (FLOMEKA maj 2003)

slide-11
SLIDE 11

RECALIBRATION, NEW METER OR REFURBISHMENT

Considerations: Price of new meter Price of refurbishing of meter (Refurbishment is a meter rebuild with some/all internal parts changed) Price of (re)calibration Meter size Recalibration if: Turbine Requirement 1: Turbine was dismounted with no failures since last calibration (8 years) Turbine Requirement 2: Examination report of turbine revealed no points of concern. Set turbine recalibration interval: 5 year. New meter if: Price of refurbishing ≤ 70% of new meter. (Turbines above G1000 fulfill this) Set calibration interval: 8 years Refurbished if: Old meter is in storage ( and savings over new metes as described above) Set calibration interval: 8 years Energinet: 65% new meters , 30% refurbished meter, 5% recalibrated meters

11

Guidelines Energinet uses on meter change

slide-12
SLIDE 12

DETERMINING CALIBRATION INTERVALS

Case 1: Site conditions to determining calibration intervals. Case 2: Follow guidelines – Common for Danish gas companies Case 3: Study stability of meter - by recalibration to determine calibration intervals

12

The 3 ways

slide-13
SLIDE 13

13

slide-14
SLIDE 14

DECENTRAL POWERPLANTS

14

Fuel Coa

  • al

Natura ral g gas Water/nat

  • at. g

gas Coa

  • al/Biomas

mass Other er mult lti Wast ste Biomas mass Biog iogas Windfarms Ele lectric icity p y prod

  • ducing u

unit its Centra ral powerplant Decentra ral powerplant >25 M 5 MW Decentra ral powerplant ≤25 MW