beam diagnostics for ion sources
play

Beam Diagnostics for Ion Sources CERN Accelerator School 2012 Uli - PowerPoint PPT Presentation

Beam Diagnostics for Ion Sources CERN Accelerator School 2012 Uli Raich CERN BE/BI CAS Slovakia 25.5 - 8.6. 2012 Uli Raich CERN BE/BI 1 The LHC and its injectors High particle density, small emittance -> high luminosity CAS Slovakia


  1. Beam Diagnostics for Ion Sources CERN Accelerator School 2012 Uli Raich CERN BE/BI CAS Slovakia 25.5 - 8.6. 2012 Uli Raich CERN BE/BI 1

  2. The LHC and its injectors High particle density, small emittance -> high luminosity CAS Slovakia 25.5 - 8.6. 2012 Uli Raich CERN BE/BI 2

  3. CERN accelerator chain for Hadrons • Source: up to 100 KeV • RFQ: up to some MeV • Linac: 50 Mev – few GeV Synchrotons: up to some TeV • CAS Slovakia 25.5 - 8.6. 2012 Uli Raich CERN BE/BI 3

  4. Source and RFQ Source and LEBT determine beam properties later in the accelerator chain Need to measure beam parameters before entering the RFQ CAS Slovakia 25.5 - 8.6. 2012 Uli Raich CERN BE/BI 4

  5. LEBT • Transport beam from the source to the RFQ CAS Slovakia 25.5 - 8.6. 2012 Uli Raich CERN BE/BI 5

  6. Parameters to be measured • Beam Intensity – Faraday Cup (destructive) – Transformer (non destructive) • Transverse Profile – Wire Harps and Wire Scanners – Residual Gas Monitors • Transverse Phase space – Slit/Grid device – Allison Scanner – Pepperpot • Energy and Energy Spread – Spectrometer CAS Slovakia 25.5 - 8.6. 2012 Uli Raich CERN BE/BI 6

  7. LEBT Commissioning Stages Faraday Emittance 1 Source meter Cup Emittance Faraday Source Sol. 1 2 meter Cup Slit Source Sol. 3 1 RFQ Input SEM grid Faraday Emittance Faraday Cup + Source Sol. Sol. 4 SEM meter Cup 1 2 Grid Beam Current CAS Slovakia 25.5 - 8.6. 2012 Uli Raich CERN BE/BI 7 Transformer (BCT)

  8. Parameters to be measured • Beam Intensity – Faraday Cup (destructive) – Transformer (non destructive) • Transverse Profile – Wire harps – Residual Gas Monitors • Transverse Phase space – Slit/Grid device – Allison Scanner – Pepperpot • Energy and Energy Spread – Spectrometer CAS Slovakia 25.5 - 8.6. 2012 Uli Raich CERN BE/BI 8

  9. Faraday Cup • Source intensity measured by a retractable Faraday Cup • Secondary electron emission is suppressed by polarization voltage which also eliminates parasitic electrons created in the source • Pneumatic in/out mechanism on PLC is used to enter and retract the cup into/from the beam • Oscilloscope is used for signal observation • A ~ 1 MHz sampling ADC may be used to acquire the Faraday Cup signal CAS Slovakia 25.5 - 8.6. 2012 Uli Raich CERN BE/BI 9

  10. Faraday Cup pieces active electrode guard ring Faraday Cup body CAS Slovakia 25.5 - 8.6. 2012 Uli Raich CERN BE/BI 10

  11. Electro-static Field in Faraday Cup In order to keep secondary electrons within the cup a repelling voltage is applied to the polarization electrode Since the electrons have energies of less than 20 eV some 100V repelling voltage is sufficient CAS Slovakia 25.5 - 8.6. 2012 Uli Raich CERN BE/BI 11

  12. Energy of secondary emission electrons 90keV 50keV 30keV I total vs. eV 1.6 • With increasing repelling voltage the electrons do not escape the 1.4 Faraday Cup any more and the 1.2 current measured stays stable. • At 40V and above no decrease in 1 the Cup current is observed any I(µA) 0.8 more 0.6 0.4 0.2 0 0.1 1 10 100 1000 CAS Slovakia 25.5 - 8.6. 2012 Uli Raich CERN BE/BI 12 V

  13. Faraday Cup with water cooling For higher intensities water cooling may be needed CAS Slovakia 25.5 - 8.6. 2012 Uli Raich CERN BE/BI 13

  14. Parameters to be measured • Beam Intensity – Faraday Cup (destructive) – Transformer (non destructive) • Transverse Profile – Wire harps – Residual Gas Monitors • Transverse Phase space – Slit/Grid device – Allison Scanner – Pepperpot • Energy and Energy Spread – Spectrometer CAS Slovakia 25.5 - 8.6. 2012 Uli Raich CERN BE/BI 14

  15. Current Transformers Magnetic field Fields are very low r i Capture magnetic field lines with cores of high r o relative permeability (CoFe based amorphous alloy Vitrovac: μ r = 10 5 ) Beam current µ µ r β = qeN qeN c 2 0 r 0 L lN ln = = I beam 2 π r t l i Uli Raich CERN BE/BI CAS Slovakia 25.5 - 8.6. 2012 15

  16. The passive AC transformer The active AC transformer The ideal transformer R F t − R L s dI τ = = U ( t ) I ( t ) e droop beam R L R L U L beam N dt A C S C S R Inductance L of the winding τ = L C Transformer output signal rise s s Beam signal L L τ = ≈ L τ = droop R R + droop + R R f L R L L A Uli Raich CERN BE/BI CAS Slovakia 25.5 - 8.6. 2012 16

  17. Principle of a fast current transformer • 500MHz Bandwidth Low droop (< 0.2%/ µ s) Image • Current Ceramic Gap BEAM 80nm Ti Coating ⇒ 20 Ω to improve Calibration winding impedance CAS Slovakia 25.5 - 8.6. 2012 Uli Raich CERN BE/BI 17

  18. Current Transformers Good magnetic shielding avoids interference from nearby pulsing magnets Shielding simulation and test measurements have been done CAS Slovakia 25.5 - 8.6. 2012 Uli Raich CERN BE/BI 18

  19. Typical Transformer Signal Calibration signal before after beam pulse Digitization of 400 µs pulse at 10 MHz Measures • total intensity • intensity per Booster ring Background suppression by software CAS Slovakia 25.5 - 8.6. 2012 Uli Raich CERN BE/BI 19

  20. The DC current transformer AC current transformer can be extended to very long droop times but not to DC Measuring DC currents is needed on DC ion sources Must provide a modulation frequency Takes advantage of non/linear magnetisation curve B H CAS Slovakia 25.5 - 8.6. 2012 Uli Raich CERN BE/BI 20

  21. Principle of DCCT V a -V b Synchronous V a V b detector modulator Power supply beam Compensation current I feedback =-I beam R V=RI beam CAS Slovakia 25.5 - 8.6. 2012 Uli Raich CERN BE/BI 21

  22. Modulation of a DCCT without beam dB U = NA B dt B=f(t) = ∫ Udt + B B H 0 NA Modulation current has only odd harmonic frequencies since the signal is symmetric 1 2 3 4 5 CAS Slovakia 25.5 - 8.6. 2012 Uli Raich CERN BE/BI 22

  23. Modulation of a DCCT with beam B=f(t) B H Sum signal becomes non-zero Even harmonics appear 1 1 2 3 4 5 CAS Slovakia 25.5 - 8.6. 2012 Uli Raich CERN BE/BI 23

  24. Modulation current difference signal with beam • Difference signal has 2 ω m • ω m typically 200 Hz – 10 kHz • Use low pass filter with ω c << ω m • Provide a 3rd core, normal AC transformer to extend to higher frequencies CAS Slovakia 25.5 - 8.6. 2012 Uli Raich CERN BE/BI 24

  25. Photo of DCCT internals CAS Slovakia 25.5 - 8.6. 2012 Uli Raich CERN BE/BI 25

  26. Parameters to be measured • Beam Intensity – Faraday Cup (destructive) – Transformer (non destructive) • Transverse Profile – Wire harps and scanners – Residual Gas Monitors • Transverse Phase space – Slit/Grid device – Allison Scanner – Pepperpot • Energy and Energy Spread – Spectrometer CAS Slovakia 25.5 - 8.6. 2012 Uli Raich CERN BE/BI 26

  27. SEMGrids for Profile Meas. • SEMGrid resolution: up to 0.5mm, up to 36 wires • New analogue electronics for 36 under design • Needs time resolved measurements (200 kHz) • New VME readout card has been developed (36 channels), series of 50 cards have been produced • In/out mechanism by motor with PLC control CAS Slovakia 25.5 - 8.6. 2012 Uli Raich CERN BE/BI 27

  28. Wire Scanners Slowly drives the wire through the beam Measures wire position and collected current on the wire Reconstructs the beam profile CAS Slovakia 25.5 - 8.6. 2012 Uli Raich CERN BE/BI 28

  29. Parameters to be measured • Beam Intensity – Faraday Cup (destructive) – Transformer (non destructive) • Transverse Profile – Wire harps and scanners – Residual Gas Monitors • Transverse Phase space – Slit/Grid device – Allison Scanner – Pepperpot • Energy and Energy Spread – Spectrometer CAS Slovakia 25.5 - 8.6. 2012 Uli Raich CERN BE/BI 29

  30. Ionisation Profile Monitor • An Ionization Profile Monitor (IPM) measures beam profile by collecting rest gas molecules/electrons ionized by the beam. • The ions/electrons are guided by electric field to MCP • Gas injection may be needed to increase yield • Micro-channel plates age, and need to be replaced. P. Forck GSI

  31. Luminescence Monitor • Gas fluorescence monitor measures light emitted by atoms/molecules excited by the beam. • Cross sections much lower than for ionization • Light emittance isotropically. • What is the rest gas pressure? F. Becker et al, GSI

  32. Parameters to be measured • Beam Intensity – Faraday Cup (destructive) – Transformer (non destructive) • Transverse Profile – Wire harps and scanners – Residual Gas Monitors • Transverse Phase space – Slit/Grid device – Allison Scanner – Pepperpot • Energy and Energy Spread – Spectrometer CAS Slovakia 25.5 - 8.6. 2012 Uli Raich CERN BE/BI 32

  33. Emittance measurements • If for each beam particle we x’ plot its position and its transverse angle we get a Beam size particle distribution who’s x boundary is an usually ellipse. • The projection onto the x axis is the beam size CAS Slovakia 25.5 - 8.6. 2012 Uli Raich CERN BE/BI 33

Download Presentation
Download Policy: The content available on the website is offered to you 'AS IS' for your personal information and use only. It cannot be commercialized, licensed, or distributed on other websites without prior consent from the author. To download a presentation, simply click this link. If you encounter any difficulties during the download process, it's possible that the publisher has removed the file from their server.

Recommend


More recommend