Analysis of Compressed Medical Air by Prashant S. Nagathan- - - PowerPoint PPT Presentation

analysis of compressed medical air
SMART_READER_LITE
LIVE PREVIEW

Analysis of Compressed Medical Air by Prashant S. Nagathan- - - PowerPoint PPT Presentation

Analysis of Compressed Medical Air by Prashant S. Nagathan- Graduate student of University at Buffalo Dr. Robert E. Baier-Professor, University at Buffalo Introduction Medical air produced on site Medical air is used for


slide-1
SLIDE 1

Analysis of Compressed Medical Air

by Prashant S. Nagathan- Graduate student of University at Buffalo

  • Dr. Robert E. Baier-Professor, University at Buffalo
slide-2
SLIDE 2

Introduction

  • Medical air produced on site
  • Medical air is used for patients-neonates,

patients suffering from respiratory disease

  • Medical air is governed by NFPA under USP
  • No specification on microbial contamination

limits

slide-3
SLIDE 3

Materials and Methods

  • Medical air samples were taken at 5 hospitals

and at dental school

  • SMA sampler and Modified Andersen sampler

used for analysis of microbial contamination in medical air

  • Both samplers were operated for 3 minutes and

1 minute @ 1CFM

  • In hospital-1 the medical air was also sampled

for 35 minutes using both samplers @ 1CFM

slide-4
SLIDE 4

Modified Andersen sampler & SMA sampler operating in hospital-1

slide-5
SLIDE 5

Modified Andersen sampler

Stainless steel hose Stainless steel inlet Andersen Sampler Air flow meter Pressure regulator Pressure gauge

slide-6
SLIDE 6

Materials and Methods (contd.)

  • Four media-BA, R2A, TSA, RBA were

used to detect human associated, environmental associated, thermophilic bacteria and fungi, respectively

  • BA was kept at 35oC for 4 days
  • R2A was kept at 23+/-3oC for 4 days
  • RBA was kept at 23+/-3oC for 7 days
  • TSA was kept at 56oC for 4 days
slide-7
SLIDE 7

Check for Microbial growth on the four media

SMA sampler Modified Andersen sampler

slide-8
SLIDE 8

Air impactor

slide-9
SLIDE 9

Materials and Methods (contd.)

  • Germanium prism placed in Air impactor
  • Medical air was bombarded on germanium

prism kept in the Air impactor for 1 hour

  • Germanium prism held in front of

connector by attachments

  • Medical air deposits on prism were

analyzed for organic matter and particulate debris by Infrared spectroscopy, SEM and EDX

slide-10
SLIDE 10

Prism kept in the prism holder held in front of connector

Germanium prism Prism holder Connector

slide-11
SLIDE 11

Typical IR spectra-medical air

Hospital-1 Hospital-2 Hospital-3 Hospital-4

slide-12
SLIDE 12

Results

  • Twenty four (24) media of each type were

used for medical air sampling at each hospital

  • No microorganisms were detected on the

media from 5 hospitals

  • From the IR spectroscopy analysis no

chemical bands were detected except for hydrocarbons in hospital-5

slide-13
SLIDE 13

IR spectrum-medical air

Hospital-5

slide-14
SLIDE 14

Results (contd.)

  • SEM showed particles were in size range

0.1 micrometer to 100 micrometer

  • EDX analysis showed particles contained

these elements: Cu, Sn, Ca, Cl, K, Fe, Li, Zn, Al, Mg, Br, Si

slide-15
SLIDE 15

SEM and EDX of particle agglomerate on prism

Sn, Cu, Zn, Li elements detected from the agglomerate of particles by EDX shown in the figure

slide-16
SLIDE 16

SEM and EDX of the particle from the medical air bombarded

  • n prism

KBr detected by the EDX shown in the figure.

slide-17
SLIDE 17

SEM and EDX of the particle from the medical air bombarded

  • n prism

KCl salt particle detected by EDX

slide-18
SLIDE 18

SEM and EDX of the particle from the medical air bombarded

  • n prism

Li, Zn, Cu, Ca, Sn, Cl elements are detected by EDX

slide-19
SLIDE 19

SEM and EDX of the particle from the medical air bombarded on prism

Size of the particle is approximately 100 micrometers

slide-20
SLIDE 20

Analysis of Compressed air from the dental school

  • Centralized compressor supplies the

compressed air for the labs and dental clinic

  • Compressor is oil-free and refrigerant

based drying

  • SMA, Modified Andersen sampler used for

analysis of air for detection of microorganisms

  • Both samplers were operated for 35

minutes @ 1CFM

slide-21
SLIDE 21

Analysis of Compressed air from the dental school (contd.)

Modified Andersen sampler

  • perating in the lab

SMA sampler operating in the lab

slide-22
SLIDE 22

Analysis of Compressed air from the dental school (contd.)

Compressed air samples from air tap using SMA, Modified Andersen sampler, SMA on top of Andersen sampler, 1CFM, 35 min

slide-23
SLIDE 23

Analysis of Compressed air from the dental school (contd.)

Control

Compressed air samples from dental chair using Modified Andersen sampler, 1CFM, 35 min

slide-24
SLIDE 24

Analysis of Compressed air from the dental school (contd.)

Air tap

Prism holder Germanium

prism

slide-25
SLIDE 25

Analysis of Compressed air from the dental school (contd.)

  • Germanium prism placed in the prism

holder was kept in front of the air tap

  • The compressed air was bombarded on

the prism for 1 hour

  • Compressed air deposits on prism were

analysed for organic matter and particulate debris by Infrared spectroscopy, SEM and EDX

slide-26
SLIDE 26

Analysis of Compressed air from the dental school (contd.)

IR spectrum from the compressed air deposits-lab 308 IR spectrum from the compressed air deposits-lab B30

slide-27
SLIDE 27

Analysis of Compressed air from the dental school (contd.)

Elements-Al, Sn, S detected by EDX

slide-28
SLIDE 28

Analysis of Compressed air from the dental school (contd.)

Elements- Cu, Zn, Cl, S detected by EDX

slide-29
SLIDE 29

Discussion

  • Dr. Bjerring and Dr. Oberg had detected

bacterial contamination in medical air in

  • ne hospital of Denmark
  • No detectable microorganisms in medical

air, confirming effectiveness of use of oil- free compressors and refrigerant based drying

  • Collected previously uncharacterized

metallic and flux particles from medical air systems at all five test sites

slide-30
SLIDE 30

Discussion (contd.)

  • Particles detected could derive from fluxes

during the welding operations of piping joints

  • The efficiencies of the two sampler could

not be compared because no microorganisms detected in medical air in any case

slide-31
SLIDE 31

Future work

  • Cross check Modified Andersen and SMA

sampler in controlled air volumes seeded with known microorganisms

  • Detection of Legionella bacteria in medical

facilities

  • Concentration of particles in the medical

air

  • Detection of SO2, NO and NO2 during

peak hours of traffic

slide-32
SLIDE 32

Acknowledgement

I sincerely thank

  • Praxair Inc.
  • Dr. Robert E. Baier
  • Plant operation managers of the Hospitals