Perspectives on Efficacy Standards Development of a Standard Test - - PowerPoint PPT Presentation

perspectives on efficacy standards
SMART_READER_LITE
LIVE PREVIEW

Perspectives on Efficacy Standards Development of a Standard Test - - PowerPoint PPT Presentation

Perspectives on Efficacy Standards Development of a Standard Test Method for Validating the Antimicrobial Efficacy of Whole Room UV Devices Matthew Hardwick, PhD ResInnova Laboratories 1 INTRO Ultraviolet germicidal irradiation has a


slide-1
SLIDE 1

1

Perspectives on Efficacy Standards

Development of a Standard Test Method for Validating the Antimicrobial Efficacy of Whole Room UV Devices

Matthew Hardwick, PhD ResInnova Laboratories

slide-2
SLIDE 2

2

INTRO

  • Ultraviolet germicidal irradiation has a

long history

  • 1st documented in mid– late 19th

century*

  • Water treatment (early 1900s)
  • Air decontamination (1930s)
  • Surface decontamination (?????)

*Schmarda LK. Der Einfluss des Lichtes auf die Infusionsthierchen. Med Jahrbüucher des k. k. Österreichischen Staates. 1845;54:257–70. Downes A, Blunt TP. The influence of light upon the development of bacteria. Nature. 1877;16:218.

slide-3
SLIDE 3

3

WHOLE ROOM DISINFECTION

  • Introduced to healthcare around 2007
  • Industry has grown dramatically recently
  • Do these devices work?
  • Numerous research articles
  • Clinical trials
  • Consistency is an issue (Boyce/Donskey 2019)
slide-4
SLIDE 4

4

FEDERAL REGULATION

  • EPA kicked the can down the road in 1976
  • Regulation without Registration
  • FIFRA section 2(q)(1) and section 7
  • Therefore, no guidance or standards regarding efficacy testing
  • Lead to different approaches in testing and confusion for the

consumer

slide-5
SLIDE 5

5

TESTING STANDARDS

slide-6
SLIDE 6

6

TESTING STANDARDS

International Standards Organization (ISO)

“International Standards make things work. They give world-class specifications for products, services and systems, to ensure quality, safety and efficiency. They are instrumental in facilitating international trade.”

slide-7
SLIDE 7

7

TESTING STANDARDS

ASTM International

“Over 12,000 ASTM standards operate globally. Defined and set by us, they improve the lives of millions every day. Combined with our innovative business services, they enhance performance and help everyone have confidence in the things they buy and use.”

slide-8
SLIDE 8

8

TESTING STANDARDS

  • American National Standards Institute (ANSI)
  • NSF International
  • ANSI/NSF 55
slide-9
SLIDE 9

9

TESTING STANDARDS

Goals

  • Quality
  • Consistency
  • Consumer Confidence
slide-10
SLIDE 10

10

TESTING STANDARDS

ASTM E3135-18 Standard Practice for Determining Antimicrobial Efficacy

  • f

Ultraviolet Germicidal Irradiation Against Microorganisms on Carriers with Simulated Soil

slide-11
SLIDE 11

11

IUVA TESTING STANDARD

Goals

  • Provide clarity on the antimicrobial efficacy of the UV

device in a controlled setting

  • Utility to manufacturers and benefit the industry
  • Improve public health
slide-12
SLIDE 12

12

IUVA TESTING STANDARD

Scope

The antimicrobial efficacy of a UV generating device will be evaluated by inoculating a specific strain of a known micro-

  • rganism on a test carrier and exposing it to the UV device using a

standard test protocol. After exposure, the test carrier will be assayed and quantified for surviving microbes to determine the log reduction achieved by the device.

slide-13
SLIDE 13

13

IUVA TESTING STANDARD

UV Devices

  • All whole room UV devices emitting radiation in the UV

spectrum (200-400 nm)

  • Includes portable and self-propelled devices
  • Includes devices that deploy as multiple units
slide-14
SLIDE 14

14

IUVA TESTING STANDARD

Testing Facility

  • Still in development
  • Includes a full patient room with a side bathroom
  • Performed in an accredited laboratory/organization
slide-15
SLIDE 15

15

IUVA TESTING STANDARD

Testing Organisms

  • Still in development
  • Clostridioides difficile (C. difficile) spores
  • Considered the most difficult to kill of the relevant HAI-

pathogens

  • Spores prepared according to EPA MB-28-04
slide-16
SLIDE 16

16

IUVA TESTING STANDARD

Testing Organisms (cont)

  • Spores to be plated on stainless steel disks according to ASTM

E2197

  • Plated with 5% serum load as the organic carrier
  • ASTM E3135-18
slide-17
SLIDE 17

17

IUVA TESTING STANDARD

Testing System

  • Inoculated disks will be positioned at distinct spots to reflect

relevant/important surfaces

  • Bedrails and center of the bed
  • Shaded areas (underside of the overbed table)
  • Near and far positions on the floor
  • Key areas of the bathroom
slide-18
SLIDE 18

18

IUVA TESTING STANDARD

Testing System (cont)

  • Positioning and exposure time of the UV device according to

manufacturer’s recommendation

  • Device positioning and UV exposure should reflect real-world

use of the device

  • In other words, the data should reflect how the device will

be used in a healthcare setting

slide-19
SLIDE 19

19

IUVA TESTING STANDARD

Other Important Considerations

  • Description of the Testing Chamber/Room
  • Environmental Conditions
  • Pass/Fail Criteria
  • Report Format
slide-20
SLIDE 20

20

THANK YOU!

  • Troy Cowan
  • Ashish Mathur
  • John Boyce
  • IUVA Healthcare Working Group
slide-21
SLIDE 21

www.resinnovalabs.com

8807 Colesville Rd Suite 300 Silver Spring, MD 20910