Measurement Uncertainty - Error & Uncertainty Measurement - - PowerPoint PPT Presentation

measurement uncertainty error
SMART_READER_LITE
LIVE PREVIEW

Measurement Uncertainty - Error & Uncertainty Measurement - - PowerPoint PPT Presentation

Instrumentation (and Error & Process Control) Uncertainty Fall 1393 Bonab University Error Measurement Uncertainty - Error & Uncertainty Measurement errors are impossible to avoid We can minimize their magnitude by Good


slide-1
SLIDE 1

Instrumentation (and

Process Control)

Fall 1393 Bonab University

Error & Uncertainty

slide-2
SLIDE 2

Measurement Uncertainty - Error

  • Measurement errors are impossible to avoid
  • We can minimize their magnitude by
  • Good measurement system design
  • Appropriate analysis and processing of measurement data
  • All error sources: How to eliminate or reduce their magnitude
  • Errors:
  • Arise during the measurement process *
  • Arise due to later corruption of the measurement signal (by induced noise during transfer of the signal)
  • In any measurement system it’s important to:
  • Reduce errors to the minimum possible level
  • quantify the maximum remaining error that may exist in output reading
  • What if system final output is calculated by combining together two or more measurements?
  • How each separate measurement error be combined  best estimate of the final output error
  • Error main categories:
  • Systematic
  • Random

2

Error & Uncertainty

slide-3
SLIDE 3

Measurement Uncertainty - Error

  • Systematic:
  • Describe errors in the output readings that are consistently on one side of the correct

reading, that is, either all errors are positive or are all negative

  • System disturbance during measurement
  • The effect of environmental changes
  • Bent needles, use of uncalibrated instruments, drift, poor cabling, …
  • The remaining is quantified by the quoted accuracy
  • Random:
  • (precision errors) are perturbations of the measurement in either side of the true value

caused by random and unpredictable effects, such that positive errors and negative errors

  • ccur in approximately equal numbers
  • mainly small, but large perturbations occur from time to time
  • Human observation of analog device + interpolation
  • Electrical noise
  • Can be largely removed by: many measurements  averaging or other statistical techniques
  • The best way is to express them in probabilistic terms (say, 95% CI)

3

Error & Uncertainty

slide-4
SLIDE 4

Sources of Systematic Error

  • Disturbance of the measured system by the act of measurement
  • Mercury-in-glass thermometer
  • Orifice plate
  • General rule: the process of measurement always disturbs

system being measured

  • Accurate understanding of the mechanisms of

system disturbance is needed to minimize it

  • Case: Electric circuits:
  • The´venin’s theorem *
  • Rm acts as a shunt
  • Rm increase  the ratio = 1
  • Practical issues (increasing moving-coil instrument’s Rm)
  • Solve: changing the spring constant
  • Ruggedness changes, and needs better friction
  • So, usually improving one aspect  introduce another problem
  • Using active devices improves this limit
  • Case: measuring instrument in a bridge circuit

4

Error & Uncertainty

slide-5
SLIDE 5

Sources of Systematic Error

  • Example:

R1 = 400 O; R2 = 600 O; R3 = 1000 O; R4 = 500 O; R5 = 1000 O The voltage across AB is measured by a voltmeter whose internal resistance is 9500 O. What is the measurement error caused by the resistance of the measuring instrument?

  • Solution:

RAB=500  measurement error = EO – Em = EO(1-9500/10000) = 0.05 EO  5%

5

Error & Uncertainty

slide-6
SLIDE 6

Sources of Systematic Error

  • Environmental disturbances (modifying inputs)
  • static and dynamic characteristics specified for measuring instruments are only valid for

particular environmental conditions

  • These specified conditions must be reproduced during calibration
  • Its magnitude quantified by:
  • sensitivity drift
  • zero drift (both included in the specifications)
  • Env. Disturbance is difficult to determine
  • Example: A small closed box (0.1 kg)  scale says 1kg

(a) a 0.9 kg rat in the box (real input) (b) an empty box with a 0.9 kg bias on the scale due to a temperature change (environmental input) (c) a 0.4 kg mouse in the box together with a 0.5 kg bias (real þ environmental inputs)

  • Thus, the magnitude of any environmental input must be measured before the value of the

measured quantity (the real input) can be determined from the output reading of an instrument

  • Designers’ choice:
  • Reduce the susceptibility of measuring instruments to environmental inputs
  • Quantify the effects of environmental inputs and correct for them in the instrument output reading

6

Error & Uncertainty

slide-7
SLIDE 7

Sources of Systematic Error

  • Changes in characteristics due to wear in instrument components (with time)
  • Systematic errors can frequently develop over a period of time because of wear in

instrument components

  • Recalibration often provides a full solution
  • Resistance of connecting leads
  • Example: a resistance thermometer
  • Often thermometer is separated by 100 meters (20-gauge copper wire is 7 Ω)
  • Also: a temperature coefficient of 1 mΩ/oC
  • Care:
  • Cross section (resistance)
  • Route (not to pick up noise)

7

Error & Uncertainty

slide-8
SLIDE 8

Reduction of Systematic Errors

  • Prerequisite : a complete analysis of the measurement system that identifies all

sources of error

  • Simple faults: bent meter needles, poor cabling practices…
  • other error sources require more detailed analysis and treatment
  • Careful Instrument Design
  • Reducing the sensitivity (strain gauge to temperature)  cost
  • Calibration
  • All instruments suffer from drift in their characteristics  it depends on:
  • environmental conditions
  • Frequency of use
  • More frequent calibration = lower drift-related error
  • Method of Opposing Inputs
  • compensates : effect of an environmental input by introducing an equal and opposite

environmental input that cancels it out

8

Error & Uncertainty

slide-9
SLIDE 9

Reduction of Systematic Errors

  • Method of Opposing Inputs
  • Example:
  • If the coil resistance Rcoil is sensitive to temperature,

environmental input ( temperature change) will alter the value of the coil current for a given applied voltage  alter the pointer output reading

  • Compensation: introducing a compensating resistance Rcomp

where Rcomp has a temperature coefficient equal in magnitude but opposite in sign to that of the coil

  • High-Gain Feedback
  • Unknown voltage Ei is applied to
  • A motor of torque constant Km
  • Resistance spring constant Ks
  • Effect of environment on motor/spring = Dm/DS

9

Error & Uncertainty

slide-10
SLIDE 10

Reduction of Systematic Errors

  • High-Gain Feedback
  • No environment input: displacement Xo = KmKsEi

, but changes with environment

  • But if we close the loop:
  • Adding amplifier: Ka
  • Feedback device: Kf

 high Ka 

  • Only Kf !  we have to be concerned only with Df
  • Signal Filtering
  • corruption of reading by periodic noise
  • often at a frequency of 50 Hz caused by pickup through the close proximity to apparatus or

current-carrying cables

  • High frequency noise (mechanical oscillation/vibration)
  • Appropriate filter (LP

, BP , BS) reduces noise amplitude

10

Error & Uncertainty

slide-11
SLIDE 11

Reduction of Systematic Errors

  • Signal Filtering
  • Example: passive RC LP filter
  • Manual Correction of Output Reading
  • Errors due to
  • system disturbance during the act of measurement
  • Environmental changes
  • a good measurement technician reduce errors
  • by calculating the effect of such systematic errors
  • making appropriate correction to readings
  • Not easy (needs all disturbances quantified)
  • Intelligent Instruments
  • Contain extra sensors that measure the value of environmental inputs
  • Automatically compensate the value of the output reading
  • ability to deal very effectively with systematic errors (explained later)

11

Error & Uncertainty

slide-12
SLIDE 12

Quantification of Systematic Errors

12

Error & Uncertainty

  • Do all practical steps have been taken to eliminate or reduce the magnitude of

systematic errors?  quantify the maximum likely systematic error

  • Quantification of Individual Systematic Error Components
  • first complication: exact value for a component = ?  use best estimate:
  • Environmental condition errors
  • Environment effect?
  • Assume midpoint environmental conditions
  • specify maximum measurement error as ±x% of the output reading
  • If fluctuations occur over a short period of time (random draughts of hot or cold air) this is a

rather a random error

  • Calibration errors
  • The maximum error just before the instrument is due for recalibration becomes the

basis for estimating the maximum likely error

slide-13
SLIDE 13

Quantification of Systematic Errors

13

Error & Uncertainty

  • Calibration errors
  • Example (a pressure transducer ):
  • Recalibration frequency: once the measurement error has grown to +1% of the full-scale
  • Range: 0 to 10 bar
  • How can its inaccuracy be expressed in the form of a ±x% error in the output reading?
  • Solution:
  • Just before recalibration: error grown to +0.1 bar (1% of 10 bar)
  • Half this maximum error, 0.05 bar, should be subtracted from all measurements
  • Error:
  • just after calibration: -0.05 bar ( -0.5% of FSR)
  • just before the next recalibration: +0.05 bar (+0.5% of FSR)
  • Inaccuracy due to calibration error: ±0.05% of FSR
  • System disturbance (as well as loading) errors
  • Maximum likely error = 2x (worst-case system loading)  Likely error: ±x  ±y% FSD
slide-14
SLIDE 14

Quantification of Systematic Errors

14

Error & Uncertainty

  • Calculation of Overall Systematic Error
  • Total systemic error: often composed of several separate components
  • measurement system loading
  • Environmental factors
  • Calibration errors
  • worst-case prediction of maximum error: simply add up each separate systematic error
  • Example: 3 components of systematic error with a magnitude of ±1% each, a worst-case prediction

error: sum of the separate errors = ±3%

  • However, it is very unlikely that all components be at their max/min simultaneously
  • Usual course of action: combine separate sources root-sum-squares method
  • n components:
  • Warning: manufacturers data sheets  measurement uncertainty/inaccuracy = best

estimate (manufacturer gives) about performance

  • When it’s new, used under specified conditions, and recalibrated at the recommended

frequency

slide-15
SLIDE 15

Quantification of Systematic Errors

15

Error & Uncertainty

  • This can only be a starting point in estimating the measurement accuracy

(achievable in use)

  • Many sources (systematic error) may apply in a particular situation (not

included in the accuracy calculation in the manufacturer’s data sheet)

  • Example:
  • 3 separate sources of systematic error are identified in a measurement system
  • After reducing the magnitude of these errors as much as possible, the magnitudes of the three errors

are estimated:

  • System loading:

+1.2% (Xm-Xt)

  • Environmental changes: 0.8%
  • Calibration error:

0.5%

  • Calculate the maximum possible total systematic error and the likely system error by the root-sum-

square method.

  • Solution:
  • The maximum possible system error = ±(1.2 + 0.8 + 0.5)% = ±2.5%
  • Applying the root-sum-square: likely error = ± √1.22 + 0.82 + 0.52 = ±1.53%
slide-16
SLIDE 16

Sources and Treatment of Random Errors

  • Caused by unpredictable variations (precision errors)
  • Human observation
  • electric noise
  • random environmental changes (draught), etc.
  • Small perturbations either side of the correct value (positive & negative errors
  • ccur in approximately equal numbers)  largely eliminated by averaging
  • (but ≠ 0, reason: finite number of measurements)
  • The degree of confidence (how close mean value is to the correct value)?
  • can be indicated by standard deviation or variance
  • parameters describing distribution about the mean value

16

Error & Uncertainty

slide-17
SLIDE 17

Statistical Analysis of Measurements Subject to Random Errors

  • Mean and Median Values
  • average value of a set of measurements of a constant quantity:
  • Median (was easier for a computer to find)  even number  midway
  • Mean (slightly closer to the correct value)
  • Example:
  • length of a steel bar (mm) is measured by a number of different observers

17

Error & Uncertainty Set Mean Median

A

398 420 394 416 404 408 400 420 396 413 430 409 408

B

409 406 402 407 405 404 407 404 407 407 408 406 407

C

409 406 402 407 405 404 407 404 407 407 408 406.5 406 406 410 406 405 408 406 409 406 405 409 406 407

  • Which one more

confidence?

  • Low-Spread: say range

430-394=36 vs 409-402=7

  • Median closer

to mean

slide-18
SLIDE 18

Statistical Analysis of Measurements Subject to Random Errors

  • Standard Deviation and Variance
  • Spread = range between the largest and the smallest value  not a very good way of

examining distribution

  • Much better: variance or standard deviation  start with deviation (error)
  • di = xi – xmean
  • Variance:
  • Standard deviation: 
  • Definitions: infinite number of data  not in practice
  • Finite measurements: xmean ≠ true mean (µ)
  • A better prediction: Bessel correction
  • Finite number:
  • Example:
  • Previous sets of measurement

18

Error & Uncertainty Set Mean Median spread

V σ

A

409 408 36 137 11.7

B

406 407 7 4.2 2.05

C

406.5 406 8 3.53 1.88

_ _

slide-19
SLIDE 19

Statistical Analysis of Measurements Subject to Random Errors

  • Graphical Data Analysis Techniques—Frequency Distributions
  • Simplest: Histogram
  • Bands/bins of equal width across the range of measurement
  • # measurements within each band
  • Finding the # of bands/bins (Sturgis Rule):
  • Example: 23 measurements in set-C
  • Bins:  5
  • Span: 402-410mm
  • Width?  2mm works
  • Care in choice of boundaries:
  • No measurements on the boundary
  • Say, put the middle bin on the Mean (406.5)
  • Large enough # of measurement & truly

Random error  symmetry

  • Usually, error is of most concern  deviation

19

Error & Uncertainty

# Measurements

mm

slide-20
SLIDE 20

Gaussian (Normal) Distribution

  • Measurement sets that only contain random errors  a distribution with a

particular shape that is called Gaussian

  • Frequency of small deviations from the mean >>

the frequency of large deviations

  • Measurements in a data set subject to random

errors lie inside deviation boundaries of ±σ  68%

  • Lie inside deviation boundaries of ±2σ

 95.4%

  • Lie inside deviation boundaries of ±3σ

 99.7%

20

±1.96σ  95% (Very common)

Error & Uncertainty

slide-21
SLIDE 21

Standard Error of the Mean

  • We examined: how measurements with random errors are distributed about the

mean

  • However, we know: error exists (mean value of a finite set - true value)
  • The standard deviation of mean values of a series of finite

sets of measurements relative to the true mean = standard error of the mean  α

  • Question:
  • if we use the mean value of a finite set of measurements to

predict the true value  what is the likely error?

  • S.d. of error = α  68% of deviations around true value within ±α

21

Error & Uncertainty

Mean of 10

slide-22
SLIDE 22

Standard Error of the Mean

  • Question:
  • if we use the mean value of a finite set of measurements to

predict the true value  what is the likely error?

  • S.d. of error = α  68% of deviations around true value within ±α
  • Means: with 68% certainty that the magnitude of the error

does not exceed |α|

  • For data set C , n = 23, σ = 1.88  α = 0.39
  • The length (average): 406.5 ± 0.4 (68% confidence limit)
  • ±2α  length : 406.5 ± 0.8 (95.4% confidence limit)
  • ±3α  length : 406.5 ± 1.2 (99.7% confidence limit)

22

Error & Uncertainty

Mean of 10

Not so good

slide-23
SLIDE 23

Estimation of Random Error in a Single Measurement (n=1)

  • Usually, not practical (repeated measurements  find average)
  • Or measured quantity is not constant
  • What: likely magnitude of error?
  • Often: calculate the error within 95%confidence limits  ± 1.96σ
  • However, it was only maximum likely deviation from calculated mean
  • Not the true value
  • Add: standard error of the mean to the likely maximum deviation value (95%)
  • Example:
  • A standard mass is measured 30 times (same instrument), σ=0.46  α=0.08
  • Now, measure an unknown mass 105.6 kg, how should the mass value be expressed?
  • Solution:
  • ± 1.96(σ+α) = ±1.06  mass: 105.6±1.06 kg

23

Error & Uncertainty

slide-24
SLIDE 24

Rogue Data Points (Data Outliers)

  • Very large error measurements sometimes occur (random & unpredictable)
  • Error magnitude: much larger than the expected random variations
  • Sources:
  • Sudden transient voltage surges
  • Incorrect data recording
  • Accepted practice:
  • Discard these data points
  • Threshold: ±3σ
  • Practical problem:
  • When a new dataset is measured (S.D. is not known)
  • It’s possible to have outlier in measurements
  • Simple solution:
  • Any new set of measurement  Histogram
  • Examine to spot outliers
  • Exclude if any  calculate ±3σ to test future measurements

24

* * *

Boxplot

Error & Uncertainty

slide-25
SLIDE 25

Aggregation of Measurement System Errors

  • Usually, 2 or more sources of measurement error
  • Total likely error in output reading?
  • Forms of aggregation:
  • A measurement have both: systematic (±x) & random errors (±y)
  • Often likely maximum error:
  • A measurement system have several measurement each with separate errors
  • Say, different instruments/transducers  add, subtract, multiply, divide
  • Example: as S = y + z
  • Problem: error term is not expressed as percentage of calculated value for S

25

Error & Uncertainty

slide-26
SLIDE 26

Aggregation of Measurement System Errors

  • Problem: error term is not expressed as percentage of calculated value for S
  • Statistical analysis:
  • Example: A circuit requirement for a resistance of 550 (2 resistors of nominal values 220

and 330 in series)

  • If each resistor has a tolerance of ±2%, the error in the sum?
  • It can be shown that the error (e) for subtraction is the same

26

Error & Uncertainty