SLIDE 6 Definition of OCT criteria
6 Xience metallic stent Absorb bioresorbable scaffold
Intraluminal defect: 4 points
Neointimal thickness (NIT) was measured from the endoluminal side of the strut hyperintense signal to the lumen contour following the center of gravity of the vessel NIT was drawn from the endoluminal border of the black box to the lumen
- contour. Because the mean
± SD thickness of the endoluminal frame post- implantation (without tissue coverage) is 34 ± 6 μm. uncovered struts at follow- up were defined when the NIT was ≤30 μm. Malapposed struts were defined whenever the distance (endoluminal strut leading edge – lumen) was larger than the metallic strut thickness 100 μm. Malapposed struts were defined when the abluminal border of the hyperintense frame was separated from the vessel wall.
Xience metallic stent Absorb bioresorbable scaffold
any irregular mass “floating” into the lumen has been classified as intraluminal mass
Strut thickness 90 μm and the axial resolution of the OCT is around 10 μm. Strut thickness 150 μm
Gomez-Lara J, Serruys PW. JACC Cardiovasc Interv. 2011 Dec;4(12):1271-80; Garcia-Garcia HM, Serruys PW, JACC Cardiovasc Imaging. 2014 Nov;7(11):1130-48