Interference and Point-of-Care Testing Devices Nam K. Tran, PhD, - - PowerPoint PPT Presentation

interference and point of care testing devices
SMART_READER_LITE
LIVE PREVIEW

Interference and Point-of-Care Testing Devices Nam K. Tran, PhD, - - PowerPoint PPT Presentation

Interference and Point-of-Care Testing Devices Nam K. Tran, PhD, HCLD, (ABB), FACB Associate Clinical Professor Director of Clinical Chemistry, Special Chemistry/Toxicology and POCT 1 Learning Objectives Identify common interferences


slide-1
SLIDE 1

1

Nam K. Tran, PhD, HCLD, (ABB), FACB Associate Clinical Professor Director of Clinical Chemistry, Special Chemistry/Toxicology and POCT

Interference and Point-of-Care Testing Devices

slide-2
SLIDE 2

2

Learning Objectives

  • Identify common interferences affecting POC

testing

  • Describe cases where interfering substances

affected patient care.

  • Describe solutions to mitigate the impact of

interfering substances on POC testing.

slide-3
SLIDE 3

3

POCT Device Formats

Definition: POCT is defined as testing at or near the site of patient care

slide-4
SLIDE 4

4

POCT Device Formats

Examples:

  • Disposable
  • Handheld
  • Portable
  • Transportable
  • Benchtop
  • Monitoring
slide-5
SLIDE 5

5

POCT Device Formats

Examples:

  • Disposable
  • Handheld
  • Portable
  • Transportable
  • Benchtop
  • Monitoring

Being FDA approved as a POCT device does not mean it is not susceptible to interfering substances!!!

slide-6
SLIDE 6

6

Pre-Analytical Total Testing Process: Lab testing occurs over three critical phases:

Total Testing Process: Difference Phases

slide-7
SLIDE 7

7

Pre-Analytical Analytical

Total Testing Process: Difference Phases

Total Testing Process: Lab testing occurs over three critical phases:

slide-8
SLIDE 8

8

Pre-Analytical Analytical Post-Analytical

Total Testing Process: Difference Phases

Total Testing Process: Lab testing occurs over three critical phases:

slide-9
SLIDE 9

9

Pre-Analytical Analytical Post-Analytical TREATMENT

Total Testing Process: Difference Phases

Total Testing Process: Lab testing occurs over three critical phases:

slide-10
SLIDE 10

10

Pre-Analytical Analytical Post-Analytical TREATMENT Errors in the Pre-Analytical Phase: Most frequent source of errors (up to 70%). Incorrect Patient preparation Sample collection Transportation Accessioning Processing

Total Testing Process: Sources of Error

Components

slide-11
SLIDE 11

11

Pre-Analytical Analytical Post-Analytical TREATMENT Errors in the Pre-Analytical Phase: Most frequent source of errors (up to 70%). Incorrect

Total Testing Process: Sources of Error

Patient preparation Sample collection Transportation Accessioning Processing Incorrect patient ID Mislabeling of specimens Hemolysis Wrong specimen type Improper specimen collection Interfering substances Components Sources of Error

slide-12
SLIDE 12

12

Pre-Analytical Analytical Post-Analytical TREATMENT Testing

Total Testing Process: Sources of Error

Patient preparation Sample collection Transportation Accessioning Processing Incorrect patient ID Mislabeling of specimens Hemolysis Wrong specimen type Improper specimen collection Interfering substances Errors in the Analytical Phase: Infrequent in laboratory tests, however may be higher in POCT due to non-lab trained personnel operating devices. Components Sources of Error

slide-13
SLIDE 13

13

Pre-Analytical Analytical Post-Analytical TREATMENT Testing

Total Testing Process: Sources of Error

Patient preparation Sample collection Transportation Accessioning Processing Incorrect patient ID Mislabeling of specimens Hemolysis Wrong specimen type Improper specimen collection Interfering substances QC/calibration Operator error Bad reagents Errors in the Analytical Phase: Infrequent in laboratory tests, however may be higher in POCT due to non-lab trained personnel operating devices. Components Sources of Error

slide-14
SLIDE 14

14

Pre-Analytical Analytical Post-Analytical TREATMENT Results interpretation Entry to LIS/EMR Contacting providers Sample archiving

Total Testing Process: Sources of Error

Testing Patient preparation Sample collection Transportation Accessioning Processing Incorrect patient ID Mislabeling of specimens Hemolysis Wrong specimen type Improper specimen collection Interfering substances QC/calibration Operator error Bad reagents Errors in the Post-Analytical Phase: Second most common among laboratory-based results. Components Sources of Error

slide-15
SLIDE 15

15

Pre-Analytical Analytical Post-Analytical TREATMENT Results interpretation Entry to LIS/EMR Contacting providers Sample archiving

Total Testing Process: Sources of Error

Testing Patient preparation Sample collection Transportation Accessioning Processing Incorrect patient ID Mislabeling of specimens Hemolysis Wrong specimen type Improper specimen collection Interfering substances QC/calibration Operator error Bad reagents Misinterpretation of results IT problems Errors in the Post-Analytical Phase: Second most common among laboratory-based results. Components Sources of Error

slide-16
SLIDE 16

16

Pre-Analytical Analytical Post-Analytical TREATMENT Results interpretation Entry to LIS/EMR Contacting providers Sample archiving

Total Testing Process: Sources of Error

Testing Patient preparation Sample collection Transportation Accessioning Processing Incorrect patient ID Mislabeling of specimens Hemolysis Wrong specimen type Improper specimen collection Interfering substances QC/calibration Operator error Bad reagents Misinterpretation of results IT problems Errors in the Post-Analytical Phase: Second most common among laboratory-based results. Components Sources of Error

What is the significance of testing error in POCT?

slide-17
SLIDE 17

17

Glucose Meter Paradigm to Highlight the Role of Testing Errors

slide-18
SLIDE 18

18

Glucose Meter Paradigm to Highlight the Role of Testing Errors

>12,000 glucose meter errors are reported to the FDA

each year – highest number of adverse events for any in vitro diagnostic device.

slide-19
SLIDE 19

19

Glucose Meter Paradigm to Highlight the Role of Testing Errors

>12,000 glucose meter errors are reported to the FDA

each year – highest number of adverse events for any in vitro diagnostic device.

12,672 serious injuries reported from 2004-2008 to the FDA.

slide-20
SLIDE 20

20

Glucose Meter Paradigm to Highlight the Role of Testing Errors

>12,000 glucose meter errors are reported to the FDA

each year – highest number of adverse events for any in vitro diagnostic device.

12,672 serious injuries reported from 2004-2008 to the FDA.

Most of these reported errors are due to erroneous results from interfering substances and operator error.

slide-21
SLIDE 21

21

Glucose Meter Paradigm to Highlight the Role of Testing Errors

>12,000 glucose meter errors are reported to the FDA

each year – highest number of adverse events for any in vitro diagnostic device.

12,672 serious injuries reported from 2004-2008 to the FDA.

Most of these reported errors are due to erroneous results from interfering substances and operator error.

slide-22
SLIDE 22

22

Glucose Meter Paradigm to Highlight the Role of Testing Errors

>12,000 glucose meter errors are reported to the FDA

each year – highest number of adverse events for any in vitro diagnostic device.

12,672 serious injuries reported from 2004-2008 to the FDA.

Most of these reported errors are due to erroneous results from interfering substances and operator error.

slide-23
SLIDE 23

23

Common Confounding Factors for Glucose Meters

Anemia and polycythemia causes falsely high or falsely low results respectively.

slide-24
SLIDE 24

24

  • 80
  • 60
  • 40
  • 20

20 40 60 10 20 30 40 50 60 70

BIAS (mg/dL) Hematocrit (%) y = -0.9465x + 30.951 R² = 0.4171

Hematocrit Effects on BGMS Measurements

Note: Bias = BGMS – Plasma Glucose Sample Size: 60 Hematocrit Range: 19 - 60% Glucose Range: 90 - 296 mg/dL

slide-25
SLIDE 25

25

Common Confounding Factors for Glucose Meters

slide-26
SLIDE 26

26

Common Confounding Factors for Glucose Meters

Oxidizing and reducing substances interfere with electrochemical sensors causing falsely high or low results.

slide-27
SLIDE 27

Drug Interferents (Oxidizing Substances)

Tang Z, et al. Am J Clin Pathol 2000;113:75-86

slide-28
SLIDE 28

28

The role of drug interferences in critical care BGMS accuracy Tran NK, et al. J Burn Care Res 2014;35:72-79

slide-29
SLIDE 29

50 100 150 200 250 300 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 32 34 36 Time Following Admission (Hours) Blood Glucose (mg/dL) Started Ascorbic Acid (66 mg/kg/hr) Stopped Ascorbic Acid

BGMS B***

History: Patient is a 21 y/o woman with 90% TBSA burns from MVA.

CASE EXAMPLE: ASCORBIC ACID INTERFERENCE

slide-30
SLIDE 30

50 100 150 200 250 300 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 32 34 36 Time Following Admission (Hours) Blood Glucose (mg/dL) Stopped Ascorbic Acid

BGMS B PLASMA GLUCOSE (Lab)

Started Ascorbic Acid (66 mg/kg/hr)

CASE EXAMPLE: ASCORBIC ACID INTERFERENCE

slide-31
SLIDE 31

50 100 150 200 250 300 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 32 34 36 Time Following Admission (Hours) Blood Glucose (mg/dL) Stopped Ascorbic Acid

BGMS A BGMS B*** PLASMA GLUCOSE (Lab)

Started Ascorbic Acid (66 mg/kg/hr)

CASE EXAMPLE: ASCORBIC ACID INTERFERENCE

slide-32
SLIDE 32

32

Common Confounding Factors for Glucose Meters

slide-33
SLIDE 33

33

Common Confounding Factors for Glucose Meters

Specimen temp alters biosensor enzyme kinetics. Hypotension/shock affect capillary specimens.

slide-34
SLIDE 34

34

Common Confounding Factors for Glucose Meters

Some glucose meters cannot differentiate between certain non- glucose sugars (e.g., maltose, galactose)

slide-35
SLIDE 35

35

Non-Glucose Sugar I nterferences

  • Icodextrin is a dialysis drug. It is metabolized by the body to maltose. In some

glucose biosensors, maltose is indistinguishable from glucose.

slide-36
SLIDE 36
slide-37
SLIDE 37

37

FDA MAUDE Database website: http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfmaude/ search.cfm, Accessed on August 20, 2014

BGMS A BGMS B BGMS C Timeframe 1997-14 2013-14 2007-11 Adverse Events (Deaths) 28 (13) 5 (0) 0 (0) Erroneous Results 557 168 15 Non-Clinical Event 387 59 21 TOTAL 1094 232 36

Maltose Related Deaths

slide-38
SLIDE 38

38

  • Similar sensor designs so susceptible

to similar interferences (will vary based on manufacturer).

  • CGM based on interstitial fluid

measurements and not plasma or whole blood.

  • Potential for many other sources of

interferences.

  • CGM does not fall under CLIA and

most devices compared against

  • bsolete or poor reference methods

such as the YSI.

  • Use WITH caution!

Continuous Glucose Monitors?

slide-39
SLIDE 39

39

I NTERFERENCES I N W HOLE BLOOD ANALYSI S

slide-40
SLIDE 40

40

I NTERFERENCES I N W HOLE BLOOD ANALYSI S Air Contamination Delayed Testing Hemolysis Hemodilution/Hemoconcentration

slide-41
SLIDE 41

41

I NTERFERENCES I N W HOLE BLOOD ANALYSI S Air Contamination Delayed Testing Hemodilution/Hemoconcentration Hemolysis

slide-42
SLIDE 42

42

Air Contam ination of Blood Specim ens

Background: Anesthesia reports “impossible venous blood gas values” in one patient where end tidal CO2 was greater than the venous blood gas (VBG).

slide-43
SLIDE 43

43

Air Contam ination of Blood Specim ens

Background: Anesthesia reports “impossible venous blood gas values” in one patient where end tidal CO2 was greater than the venous blood gas (VBG).

  • POC Venous Blood Gas: pH = 7.54, pCO2 = 17.5, pO2 = 168.5
  • POC VBG#2: pH = 7.56, pCO2 = 12.7, pO2 = 165.9
  • End tidal CO2 = 28
slide-44
SLIDE 44

44

Air Contam ination of Blood Specim ens

Background: Anesthesia reports “impossible venous blood gas values” in one patient where end tidal CO2 was greater than the venous blood gas (VBG).

  • POC Venous Blood Gas: pH = 7.54, pCO2 = 17.5, pO2 = 168.5
  • POC VBG#2: pH = 7.56, pCO2 = 12.7, pO2 = 165.9
  • End tidal CO2 = 28
  • Lab Venous Blood Gas: pH 7.54, pCO2 = 19.2, pO2 = 161.5
slide-45
SLIDE 45

45

Air Contam ination of Blood Specim ens

Blood Gas Laboratory identified “air bubbles” in syringe

Background: Anesthesia reports “impossible venous blood gas values” in one patient where end tidal CO2 was greater than the venous blood gas (VBG).

  • POC Venous Blood Gas: pH = 7.54, pCO2 = 17.5, pO2 = 168.5
  • POC VBG#2: pH = 7.56, pCO2 = 12.7, pO2 = 165.9
  • End tidal CO2 = 28
  • Lab Venous Blood Gas: pH 7.54, pCO2 = 19.2, pO2 = 161.5
slide-46
SLIDE 46

46

Air Contam ination of Blood Specim ens

Background: Anesthesia reports “impossible venous blood gas values” in one patient where end tidal CO2 was greater than the venous blood gas (VBG).

  • POC Venous Blood Gas: pH = 7.54, pCO2 = 17.5, pO2 = 168.5
  • POC VBG#2: pH = 7.56, pCO2 = 12.7, pO2 = 165.9
  • End tidal CO2 = 28
  • Lab Venous Blood Gas: pH 7.54, pCO2 = 19.2, pO2 = 161.5
  • Air bubbles can quickly (<5 mins) cause the specimen to

equilibrate atmospheric air (1 atm = 760 mmHg = 0.21 x 760 = 150 mmHg for pO2!!!)

slide-47
SLIDE 47

47

I NTERFERENCES I N BLOOD GAS ANALYSI S Air Contamination Delayed Testing Hemolysis Hemodilution/Hemoconcentration

slide-48
SLIDE 48

48

Pre-Analytical

  • Transportation delays

Analysis should be performed within 20 to 30 minutes—Faster is better!

Specimen Processing Delays and Lactate

slide-49
SLIDE 49

49

Pre-Analytical

  • Transportation delays

Seymour CW, et al. BMC Research Notes 2011;4:169

Specimen Processing Delays and Lactate

slide-50
SLIDE 50

50

Pre-Analytical

  • Transportation delays
  • Inadequate inhibition of

glycolysis If delays are expected, using a grey top tube may be appropriate, however it may take up to 15 minutes to achieve inhibition!

Specimen Processing Delays and Lactate

slide-51
SLIDE 51

51

Pre-Analytical

  • Transportation delays
  • Inadequate inhibition of

glycolysis Astles R, et al. Clin Chem 1994;404:1327

Specimen Processing Delays and Lactate

slide-52
SLIDE 52

52

Pre-Analytical

  • Transportation delays
  • Inadequate inhibition of

glycolysis If delays are expected, using a grey top tube may be appropriate, however it may take up to 15 minutes to achieve inhibition! Astles R, et al. Clin Chem 1994;404:1327

Specimen Processing Delays and Lactate

slide-53
SLIDE 53

53

Pre-Analytical

  • Transportation delays
  • Inadequate inhibition of

glycolysis

  • Specimens not placed on ice

False elevations of lactate could be mitigated by placing samples on ice. Iced samples exhibit similar results to those tested immediately at up to 6 hours.

Specimen Processing Delays and Lactate

slide-54
SLIDE 54

54

Pre-Analytical

  • Transportation delays
  • Inadequate inhibition of

glycolysis

  • Specimens not placed on ice

Seymour CW, et al. BMC Research Notes 2011;4:169

Specimen Processing Delays and Lactate

slide-55
SLIDE 55

55

I NTERFERENCES I N W HOLE BLOOD ANALYSI S Air Contamination Delayed Testing Hemolysis Hemodilution/Hemoconcentration

slide-56
SLIDE 56

56

  • Spectrophotometric (Non-Cyanohemoglobin)

Contem porary Hem oglobinom etric Techniques

  • Measurement of hemoglobin is based on

the absorption spectra

  • Oxy- and deoxyhemoglobin exhibit

different absorption in the red to IR wavelengths.

  • Measurement based on Beer’s Law

(A = elc).

  • Some methods require lysis and reacting

with non-cyanide-based reagents.

Absorbance

HHb O2Hb COHb MetHb

Prism White light

500 550 600 650 700 nm

slide-57
SLIDE 57

57

Conductance (Impendance)

Contem porary Hem oglobinom etric Techniques

Electrode VS.

  • Red blood cell membranes are not conductive.

High Resistance Low Resistance

slide-58
SLIDE 58

58

Conductance (Impendance)

Contem porary Hem oglobinom etric Techniques

Electrode VS. Hematocrit (%) Resistance (Ω)

  • Red blood cell membranes are not conductive.
slide-59
SLIDE 59

59

Conductance (Impendance)

Contem porary Hem oglobinom etric Techniques

Electrode VS.

  • Red blood cell membranes are not conductive.
  • The number of red blood cells is proportional to the change in conductance and conforms to

Ohm’s Law (V = IR)

slide-60
SLIDE 60

60

Conductance (Impendance)

Contem porary Hem oglobinom etric Techniques

Electrode VS.

  • Red blood cell membranes are not conductive.
  • The number of red blood cells is proportional to the change in conductance and conforms to

Ohm’s Law (V = IR)

  • Conductance-based methods measure hematocrit. The hematocrit can then be used to calculate

hemoglobin based on a conversion factor (estimated hemoglobin = hematocrit / 3.4)*

slide-61
SLIDE 61

61

Conductance (Impendance)

Contem porary Hem oglobinom etric Techniques

Electrode VS.

slide-62
SLIDE 62

62

Case Study 2 : Hem oconcentration

Background: Patient with suspected Ebola Virus symptoms admitted for

  • evaluation. Isolation protocols were in effect.
slide-63
SLIDE 63

63

Case Study 2 : Hem oconcentration

Background: Patient with suspected Ebola Virus symptoms admitted for

  • evaluation. Isolation protocols were in effect.
slide-64
SLIDE 64

64

Case Study 2 : Hem oconcentration

Background: Patient with suspected Ebola Virus symptoms admitted for

  • evaluation. Isolation protocols were in effect. A handheld blood gas chemistry

analyzer served as the primary chemistry analyzer. 0853 hrs – Specimens collected for chemistry and CBC testing.

slide-65
SLIDE 65

65

Case Study 2 : Hem oconcentration

Background: Patient with suspected Ebola Virus symptoms admitted for

  • evaluation. Isolation protocols were in effect. A handheld blood gas chemistry

analyzer served as the primary chemistry analyzer. 0853 hrs – Specimens collected for chemistry and CBC testing.

slide-66
SLIDE 66

66

Case Study 2 : Hem oconcentration

Background: Patient with suspected Ebola Virus symptoms admitted for

  • evaluation. Isolation protocols were in effect. A handheld blood gas chemistry

analyzer served as the primary chemistry analyzer. 0853 hrs – Specimens collected for chemistry and CBC testing.

slide-67
SLIDE 67

67

Case Study 2 : Hem oconcentration

Background: Patient with suspected Ebola Virus symptoms admitted for

  • evaluation. Isolation protocols were in effect. A handheld blood gas chemistry

analyzer served as the primary chemistry analyzer. 0853 hrs – Specimens collected for chemistry and CBC testing.

Handheld Results Hct = 68% Hb = 21.9 g/dL

slide-68
SLIDE 68

68

Case Study 2 : Hem oconcentration

Background: Patient with suspected Ebola Virus symptoms admitted for

  • evaluation. Isolation protocols were in effect. A handheld blood gas chemistry

analyzer served as the primary chemistry analyzer. 0853 hrs – Specimens collected for chemistry and CBC testing.

Handheld Results Hct = 68% Hb = 21.9 g/dL CBC Results Hct = 41% Hb = 13.2 g/dL

slide-69
SLIDE 69

69

Case Study 2 : Hem oconcentration

Background: Patient with suspected Ebola Virus symptoms admitted for

  • evaluation. Isolation protocols were in effect. A handheld blood gas chemistry

analyzer served as the primary chemistry analyzer. 0853 hrs – Specimens collected for chemistry and CBC testing.

Hct = 43% Hb = 13.8 g/dL Handheld Results Hct = 68% Hb = 21.9 g/dL RE-MIXING! CBC Results Hct = 41% Hb = 13.2 g/dL

slide-70
SLIDE 70

70

Case Study 2 : Hem oconcentration

Background: Patient with suspected Ebola Virus symptoms admitted for

  • evaluation. Isolation protocols were in effect. A handheld blood gas chemistry

analyzer served as the primary chemistry analyzer. 0853 hrs – Specimens collected for chemistry and CBC testing.

Hct = 43% Hb = 13.8 g/dL Inadequate mixing may result in artificial changes in total hemoglobin measurements. Handheld Results Hct = 68% Hb = 21.9 g/dL RE-MIXING! CBC Results Hct = 41% Hb = 13.2 g/dL

slide-71
SLIDE 71

71

Electrode

Conductance (Impendence)

Contem porary Hem oglobinom etric Techniques

  • Plasma protein content contributes to hematocrit measurements for conductance-based systems.

= Plasma Protein High Resistance

slide-72
SLIDE 72

72

Electrode

Conductance (Impendence)

Contem porary Hem oglobinom etric Techniques

= Plasma Protein

  • Plasma protein content contributes to hematocrit measurements for conductance-based systems.
  • Conductance-based systems assumes a relatively fixed protein concentration. Therefore, during

hemodilution, hematocrit may be falsely lower and causing an underestimation of total hemoglobin.

Low Resistance from low plasma protein concentration!

slide-73
SLIDE 73

73

Electrode

Conductance (Impendence)

Contem porary Hem oglobinom etric Techniques

= Plasma Protein

  • Plasma protein content contributes to hematocrit measurements for conductance-based systems.
  • Conductance-based systems assumes a relatively fixed protein concentration. Therefore, during

hemodilution, hematocrit may be falsely lower and causing an underestimation of total hemoglobin.

  • UCDMC Study: Comparison of a handheld blood gas analyzer using conductance-based

measurement of hemoglobin versus a benchtop blood gas analyzer using a spectrophotometric- based method for hemoglobinometry.

slide-74
SLIDE 74

74

  • Sixty patients requiring cardiac surgery

were evaluated.

  • Paired specimens were tested using a

handheld POC analyzer and spectrophotometric methods through the core laboratory.

  • Mean (SD) bias was -1.4 (1.1) g/dL,

P = 0.011.

  • Based on core laboratory results 12

patients would have received unnecessary transfusions.

Clinical I m pact of Hem odilution for Point-of- Care Hem oglobin Measurem ents

slide-75
SLIDE 75

75

  • Sixty patients requiring cardiac surgery

were evaluated.

  • Paired specimens were tested using a

handheld POC analyzer and spectrophotometric methods through the core laboratory.

  • Mean (SD) bias was -1.4 (1.1) g/dL,

P = 0.011.

  • Based on core laboratory results 12

patients would have received unnecessary transfusions.

Clinical I m pact of Hem odilution for Point-of- Care Hem oglobin Measurem ents

slide-76
SLIDE 76

76

  • Sixty patients requiring cardiac surgery

were evaluated.

  • Paired specimens were tested using a

handheld POC analyzer and spectrophotometric methods through the core laboratory.

  • Mean (SD) bias was -1.4 (1.1) g/dL,

P = 0.011.

  • Based on core laboratory results 12

patients would have received unnecessary transfusions.

Clinical I m pact of Hem odilution for Point-of- Care Hem oglobin Measurem ents = $219

$219 x 12 = $2,628 POTENTIALLY WASTED

Toner RW, et al. Appl Health Econ Health Policy 2011;9:29-37

slide-77
SLIDE 77

77

Case Study 3 : Hem odilution

slide-78
SLIDE 78

78

Case Study 3 : Hem odilution

Background: FDA MAUDE database reports a case (03P76-25) of a neonatal patient with discrepant point-of-care (POC) hemoglobin values compared to the

  • laboratory. The POC device used a conductance-based method of hemoglobin

measurement, while the laboratory used a spectrophotometric method.

slide-79
SLIDE 79

79

Case Study 3 : Hem odilution

Background: FDA MAUDE database reports a case (03P76-25) of a neonatal patient with discrepant point-of-care (POC) hemoglobin values compared to the

  • laboratory. The POC device used a conductance-based method of hemoglobin

measurement, while the laboratory used a spectrophotometric method.

  • POC device reported a hematocrit of 22%. Physician administered 7 mL of

blood based on the POC result.

slide-80
SLIDE 80

80

Case Study 3 : Hem odilution

Background: FDA MAUDE database reports a case (03P76-25) of a neonatal patient with discrepant point-of-care (POC) hemoglobin values compared to the

  • laboratory. The POC device used a conductance-based method of hemoglobin

measurement, while the laboratory used a spectrophotometric method.

  • POC device reported a hematocrit of 22%. Physician administered 7 mL of

blood based on the POC result.

  • Transfusion was stopped halfway after the laboratory reported a hematocrit of

40% and hemoglobin of 11.7 g/dL.

slide-81
SLIDE 81

81

Case Study 3 : Hem odilution

Background: FDA MAUDE database reports a case (03P76-25) of a neonatal patient with discrepant point-of-care (POC) hemoglobin values compared to the

  • laboratory. The POC device used a conductance-based method of hemoglobin

measurement, while the laboratory used a spectrophotometric method.

  • POC device reported a hematocrit of 22%. Physician administered 7 mL of

blood based on the POC result.

  • Transfusion was stopped halfway after the laboratory reported a hematocrit of

40% and hemoglobin of 11.7 g/dL.

  • Post-transfusion POC and lab hematocrit values were 45 and 50%

respectively.

slide-82
SLIDE 82

82

Analytical Perform ance of Optical vs. Conductance-Based Hem oglobinom etry

Device #1 Hb (g/dL) Central Laboratory Hb (g/dL)

y = 0.5092x + 4.0176 R² = 0.5253 4 5 6 7 8 9 10 11 4 5 6 7 8 9 10 11 12

slide-83
SLIDE 83

83

Analytical Perform ance of Optical vs. Conductance-Based Hem oglobinom etry

Device #2 Hb (g/dL) Central Laboratory Hb (g/dL)

y = 0.5249x + 3.9443 R² = 0.5407 4 5 6 7 8 9 10 11 4 5 6 7 8 9 10 11 12

slide-84
SLIDE 84

84

Analytical Perform ance of Optical vs. Conductance-Based Hem oglobinom etry

Device #3 Hb (g/dL)

y = 0.9345x + 0.4057 R² = 0.9205 4 5 6 7 8 9 10 11 4 5 6 7 8 9 10 11

Central Laboratory Hb (g/dL)

slide-85
SLIDE 85

85

Analytical Perform ance of Optical vs. Conductance-Based Hem oglobinom etry

Notes: Reference Method = Beckman LH hematology analyzer

slide-86
SLIDE 86

86

Analytical Perform ance of Optical vs. Conductance-Based Hem oglobinom etry

Notes: Reference Method = Beckman LH hematology analyzer Median (IQR) Bias: 0.78 (0.78) g/dL P < 0.001 N = 50

slide-87
SLIDE 87

87

Analytical Perform ance of Optical vs. Conductance-Based Hem oglobinom etry

Notes: Reference Method = Beckman LH hematology analyzer Median (IQR) Bias: 0.73 (0.60) g/dL P < 0.001 N = 50

slide-88
SLIDE 88

88

Analytical Perform ance of Optical vs. Conductance-Based Hem oglobinom etry

Notes: Reference Method = Beckman LH hematology analyzer Median (IQR) Bias: 0.22 (0.20) g/dL P = 0.510 N = 50

slide-89
SLIDE 89

89

Analytical Perform ance of Optical vs. Conductance-Based Hem oglobinom etry

Notes: *** P<0.001, Central Lab = Spectrophotometric Method, n = 20 patients 6 6.5 7 7.5 8 8.5 9 9.5 10 1 2 3 4 5 Total Hemoglobin (g/dL) Time Point Serial Testing Performance at 7 and 8 g/dL

  • Serial testing revealed significant analytical

bias between spectrophotometry vs. conductance-based measurements. *** Spectrophotometric-based Methods

slide-90
SLIDE 90

90

Analytical Perform ance of Optical vs. Conductance-Based Hem oglobinom etry

Notes: *** P<0.001, Central Lab = Spectrophotometric Method, n = 20 patients 6 6.5 7 7.5 8 8.5 9 9.5 10 1 2 3 4 5 Total Hemoglobin (g/dL) Time Point Serial Testing Performance at 7 and 8 g/dL

  • Serial testing revealed significant analytical

bias between spectrophotometry vs. conductance-based measurements. *** Spectrophotometric-based Methods Conductance-based Methods

slide-91
SLIDE 91

91

Analytical Perform ance of Optical vs. Conductance-Based Hem oglobinom etry

6 6.5 7 7.5 8 8.5 9 9.5 10 1 2 3 4 5 Total Hemoglobin (g/dL) Time Point Serial Testing Performance at 7 and 8 g/dL

  • Serial testing revealed significant analytical

bias between spectrophotometry vs. conductance-based measurements.

  • Conductance-based devices would have

prompted unnecessary transfusions at time point #5 for patients using the 7 g/dL cutoff. Unnecessary Transfusion Risk *** Notes: *** P<0.001, Central Lab = Spectrophotometric Method, n = 20 patients

slide-92
SLIDE 92

92

Analytical Perform ance of Optical vs. Conductance-Based Hem oglobinom etry

6 6.5 7 7.5 8 8.5 9 9.5 10 1 2 3 4 5 Total Hemoglobin (g/dL) Time Point Serial Testing Performance at 7 and 8 g/dL

  • Serial testing revealed significant analytical

bias between spectrophotometry vs. conductance-based measurements.

  • Conductance-based devices would have

prompted unnecessary transfusions at time point #5 for patients using the 7 g/dL cutoff.

  • All serial conductance measurements were

at risk for potential transfusions if the 8 g/dL cutoff was used. Unnecessary Transfusion Risk *** Notes: *** P<0.001, Central Lab = Spectrophotometric Method, n = 20 patients

slide-93
SLIDE 93

93

Analytical Perform ance of Optical vs. Conductance-Based Hem oglobinom etry

6 6.5 7 7.5 8 8.5 9 9.5 10 1 2 3 4 5

Central Laboratory Method

Total Hemoglobin (g/dL) Time Point Serial Testing Performance at 7 and 8 g/dL

  • Serial testing revealed significant analytical

bias between spectrophotometry vs. conductance-based measurements.

  • Conductance-based devices would have

prompted unnecessary transfusions at time point #5 for patients using the 7 g/dL cutoff.

  • All serial conductance measurements were

at risk for potential transfusions if the 8 g/dL cutoff was used. Unnecessary Transfusion Risk *** Notes: *** P<0.001, Central Lab = Spectrophotometric Method, n = 20 patients

slide-94
SLIDE 94

94

Manufacturer and User Facility Device Experience ( MAUDE) Database Sum m ary

Device 1 Device 2 Device 3 Timeframe 2011-2016 2011-2016 2014-2016* Erroneous Results 8 Improper Transfusions 5 https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfmaude/results.cfm, Accessed on July 19, 2016

slide-95
SLIDE 95

95

I NTERFERENCES I N W HOLE BLOOD ANALYSI S Air Contamination Delayed Testing Hemolysis Hemodilution/Hemoconcentration

slide-96
SLIDE 96

96

I NTERFERENCES I N W HOLE BLOOD ANALYSI S Air Contamination Delayed Testing Hemolysis Hemodilution/Hemoconcentration

Pseudohyperkalemia

slide-97
SLIDE 97

97

I NTERFERENCES I N W HOLE BLOOD ANALYSI S Air Contamination Delayed Testing Hemolysis Hemodilution/Hemoconcentration

Pseudohyperkalemia “Pseudonormokalemia”

slide-98
SLIDE 98

98

I NTERFERENCES I N W HOLE BLOOD ANALYSI S Air Contamination Delayed Testing Hemolysis Hemodilution/Hemoconcentration

Pseudohyperkalemia “Pseudonormokalemia”

No current FDA approved integrated solutions for detecting hemolysis at the point-of-care

slide-99
SLIDE 99

99

Biotin: The “Snake Oil” of 2018?

slide-100
SLIDE 100
slide-101
SLIDE 101

101

Biotin and Cardiac Troponin Testing

slide-102
SLIDE 102

102

Mumma B, et al. AACC Poster Presentation 2018

  • 1,443 Gen 5 troponin T samples tested (0-hour,

n = 797; 3-hour, n=646) from 850 patients.

Estimating the Probability of Biotin Interference

slide-103
SLIDE 103

103

Mumma B, et al. AACC Poster Presentation 2018

  • 1,443 Gen 5 troponin T samples tested (0-hour,

n = 797; 3-hour, n=646) from 850 patients.

  • Biotin not detectable in 471 (59%) and 399 (62%) 3-

hour samples.

Estimating the Probability of Biotin Interference

slide-104
SLIDE 104

104

  • 1,443 Gen 5 troponin T samples tested (0-hour,

n = 797; 3-hour, n=646) from 850 patients.

  • Biotin not detectable in 471 (59%) and 399 (62%) 3-

hour samples.

  • Only one 0-hour sample and one 3-hour sample

had biotin >20 ng/mL (0.13% [95% CI: 0-0.7%]).

Mumma B, et al. AACC Poster Presentation 2018

Estimating the Probability of Biotin Interference

slide-105
SLIDE 105

105

Mumma B, et al. AACC Poster Presentation 2018

Estimating the Probability of Biotin Interference

slide-106
SLIDE 106

106

Estimating the Probability of Biotin Interference

slide-107
SLIDE 107

107

Adult ED Patients with Unknown Biotin Status: 540

Average Plasma Biotin: 1.15 (0.97) ng/mL

Specimens collected as part of clinical validation

UC Davis Cardiac Troponin Patients

slide-108
SLIDE 108

108

Adult ED Patients with Unknown Biotin Status: 540

Average Plasma Biotin: 1.15 (0.97) ng/mL

20 40 60 80 100 120 140 160 Number of Patients Biotin Concentration (ng/mL)

Gen 5 TnT Biotin Interference Threshold is 20 ng/mL Biotin quantified by GC-TOF-MS

UC Davis Cardiac Troponin Patients

slide-109
SLIDE 109

109

Adult ED Patients with Unknown Biotin Status: 540

Average Plasma Biotin: 1.15 (0.97) ng/mL

20 40 60 80 100 120 140 160 Number of Patients Biotin Concentration (ng/mL)

Gen 5 TnT Biotin Interference Threshold is 20 ng/mL Biotin quantified by GC-TOF-MS BIOTIN IS LESS LIKELY TO BE A PROBLEM IN CARDIAC TROPONIN TESTING AND IS POPULATION SPECIFIC!

UC Davis Cardiac Troponin Patients

slide-110
SLIDE 110

110

Biotin and Urine Pregnancy Testing

slide-111
SLIDE 111

111

Biotin I nterference w ith Urine Pregnancy Tests

  • Recent studies show some point-of-

care urine pregnancy tests were affected by biotin.

  • Biotin is cleared by the kidneys.
  • In this study, the QuickVue urine

pregnancy test exhibited interference as low as 6 microgram/mL of urine biotin! Williams G, et al. Clin Biochem 2018;53:168-170

slide-112
SLIDE 112

112

Best POCT Practices for Mitigating I nterfering Substances

slide-113
SLIDE 113

113

  • Education: The laboratory must be the leader in educating providers and patients of potential

test interferences. Go to grand rounds, build partnerships, and provide multi-modality means to disseminate knowledge.

POCT Best Practices for I nterferences

slide-114
SLIDE 114

114

  • Education: The laboratory must be the leader in educating providers and patients of potential

test interferences. Go to grand rounds, build partnerships, and provide multi-modality means to disseminate knowledge.

POCT Best Practices for I nterferences

slide-115
SLIDE 115

115

  • Education: The laboratory must be the leader in educating providers and patients of potential

test interferences. Go to grand rounds, build partnerships, and provide multi-modality means to disseminate knowledge.

  • Surveillance: Know your population! Collect data and determine if your local population may be

be at risk for certain interferences (e.g., biotin, vitamin C, etc). MAUDE database is also helpful!

POCT Best Practices for I nterferences

slide-116
SLIDE 116

116

  • Education: The laboratory must be the leader in educating providers of potential test
  • interferences. Go to grand rounds, build partnerships, and provide multi-modality means to

disseminate knowledge.

  • Surveillance: Know your population! Collect data and determine if your local population may be

be at risk for certain interferences (e.g., biotin, vitamin C, etc). MAUDE database is also helpful!

  • Electronic Early-Warning Systems: Leverage electronic solutions. Ordering of susceptible

tests could flag both on the provider and laboratory side certain substances are identified.

POCT Best Practices for I nterferences

slide-117
SLIDE 117

117

Conclusions

  • Interfering substances are out there and

impact POC testing as much as traditional lab testing!

  • Interferences in common POC devices such

as glucose meters have resulted in injury and death.

  • Interferences in whole blood analysis have

resulted in inappropriate treatment decisions.

  • Medications and supplements may also affect

POC immunoassays such as urine pregnancy tests.

  • Education and awareness is critical to

minimizing errors associated with interfering substances.

slide-118
SLIDE 118

118

Questions?