interference and point of care testing devices
play

Interference and Point-of-Care Testing Devices Nam K. Tran, PhD, - PowerPoint PPT Presentation

Interference and Point-of-Care Testing Devices Nam K. Tran, PhD, HCLD, (ABB), FACB Associate Clinical Professor Director of Clinical Chemistry, Special Chemistry/Toxicology and POCT 1 Learning Objectives Identify common interferences


  1. Continuous Glucose Monitors?  Similar sensor designs so susceptible to similar interferences (will vary based on manufacturer).  CGM based on interstitial fluid measurements and not plasma or whole blood.  Potential for many other sources of interferences.  CGM does not fall under CLIA and most devices compared against obsolete or poor reference methods such as the YSI.  Use WITH caution! 38

  2. I NTERFERENCES I N W HOLE BLOOD ANALYSI S 39

  3. I NTERFERENCES I N W HOLE BLOOD ANALYSI S Air Contamination Delayed Testing Hemodilution/Hemoconcentration Hemolysis 40

  4. I NTERFERENCES I N W HOLE BLOOD ANALYSI S Air Contamination Delayed Testing Hemodilution/Hemoconcentration Hemolysis 41

  5. Air Contam ination of Blood Specim ens Background: Anesthesia reports “impossible venous blood gas values” in one patient where end tidal CO2 was greater than the venous blood gas (VBG). 42

  6. Air Contam ination of Blood Specim ens Background: Anesthesia reports “impossible venous blood gas values” in one patient where end tidal CO2 was greater than the venous blood gas (VBG). POC Venous Blood Gas: pH = 7.54, pCO2 = 17.5, pO2 = 168.5 • POC VBG#2: pH = 7.56, pCO2 = 12.7, pO2 = 165.9 • • End tidal CO2 = 28 43

  7. Air Contam ination of Blood Specim ens Background: Anesthesia reports “impossible venous blood gas values” in one patient where end tidal CO2 was greater than the venous blood gas (VBG). POC Venous Blood Gas: pH = 7.54, pCO2 = 17.5, pO2 = 168.5 • POC VBG#2: pH = 7.56, pCO2 = 12.7, pO2 = 165.9 • • End tidal CO2 = 28 Lab Venous Blood Gas: pH 7.54, pCO2 = 19.2, pO2 = 161.5 • 44

  8. Air Contam ination of Blood Specim ens Background: Anesthesia reports “impossible venous blood gas values” in one patient where end tidal CO2 was greater than the venous blood gas (VBG). POC Venous Blood Gas: pH = 7.54, pCO2 = 17.5, pO2 = 168.5 • POC VBG#2: pH = 7.56, pCO2 = 12.7, pO2 = 165.9 • • End tidal CO2 = 28 Lab Venous Blood Gas: pH 7.54, pCO2 = 19.2, pO2 = 161.5 • Blood Gas Laboratory identified “air bubbles” in syringe 45

  9. Air Contam ination of Blood Specim ens Background: Anesthesia reports “impossible venous blood gas values” in one patient where end tidal CO2 was greater than the venous blood gas (VBG). POC Venous Blood Gas: pH = 7.54, pCO2 = 17.5, pO2 = 168.5 • POC VBG#2: pH = 7.56, pCO2 = 12.7, pO2 = 165.9 • • End tidal CO2 = 28 Lab Venous Blood Gas: pH 7.54, pCO2 = 19.2, pO2 = 161.5 • Air bubbles can quickly (<5 mins) cause the specimen to • equilibrate atmospheric air (1 atm = 760 mmHg = 0.21 x 760 = 150 mmHg for pO2!!!) 46

  10. I NTERFERENCES I N BLOOD GAS ANALYSI S Air Contamination Delayed Testing Hemodilution/Hemoconcentration Hemolysis 47

  11. Specimen Processing Delays and Lactate Pre-Analytical • Transportation delays Analysis should be performed within 20 to 30 minutes—Faster is better! 48

  12. Specimen Processing Delays and Lactate Pre-Analytical • Transportation delays Seymour CW, et al. BMC Research Notes 2011;4:169 49

  13. Specimen Processing Delays and Lactate Pre-Analytical If delays are expected, using a grey top • Transportation delays tube may be appropriate, however it may • Inadequate inhibition of take up to 15 minutes to achieve glycolysis inhibition! 50

  14. Specimen Processing Delays and Lactate Pre-Analytical • Transportation delays • Inadequate inhibition of glycolysis Astles R, et al. Clin Chem 1994;404:1327 51

  15. Specimen Processing Delays and Lactate Pre-Analytical If delays are expected, using a grey top • Transportation delays tube may be appropriate, however it may • Inadequate inhibition of take up to 15 minutes to achieve glycolysis inhibition! Astles R, et al. Clin Chem 1994;404:1327 52

  16. Specimen Processing Delays and Lactate Pre-Analytical False elevations of lactate could be • Transportation delays mitigated by placing samples on ice. Iced • Inadequate inhibition of samples exhibit similar results to those glycolysis tested immediately at up to 6 hours. • Specimens not placed on ice 53

  17. Specimen Processing Delays and Lactate Pre-Analytical • Transportation delays • Inadequate inhibition of glycolysis • Specimens not placed on ice Seymour CW, et al. BMC Research Notes 2011;4:169 54

  18. I NTERFERENCES I N W HOLE BLOOD ANALYSI S Air Contamination Delayed Testing Hemodilution/Hemoconcentration Hemolysis 55

  19. Contem porary Hem oglobinom etric Techniques • Spectrophotometric (Non-Cyanohemoglobin) Absorbance • Measurement of hemoglobin is based on the absorption spectra HHb O 2 Hb COHb MetHb • Oxy- and deoxyhemoglobin exhibit different absorption in the red to IR wavelengths. • Measurement based on Beer’s Law 500 550 600 650 700 nm (A = elc). • Some methods require lysis and reacting with non-cyanide-based reagents. White light Prism 56

  20. Contem porary Hem oglobinom etric Techniques Conductance (Impendance) Electrode Low Resistance High Resistance VS. • Red blood cell membranes are not conductive. 57

  21. Contem porary Hem oglobinom etric Techniques Conductance (Impendance) Electrode VS. • Red blood cell membranes are not conductive. Resistance ( Ω ) Hematocrit (%) 58

  22. Contem porary Hem oglobinom etric Techniques Conductance (Impendance) Electrode VS. • Red blood cell membranes are not conductive. • The number of red blood cells is proportional to the change in conductance and conforms to Ohm’s Law (V = IR) 59

  23. Contem porary Hem oglobinom etric Techniques Conductance (Impendance) Electrode VS. • Red blood cell membranes are not conductive. • The number of red blood cells is proportional to the change in conductance and conforms to Ohm’s Law (V = IR) • Conductance-based methods measure hematocrit. The hematocrit can then be used to calculate hemoglobin based on a conversion factor (estimated hemoglobin = hematocrit / 3.4)* 60

  24. Contem porary Hem oglobinom etric Techniques Conductance (Impendance) Electrode VS. 61

  25. Case Study 2 : Hem oconcentration Background: Patient with suspected Ebola Virus symptoms admitted for evaluation. Isolation protocols were in effect. 62

  26. Case Study 2 : Hem oconcentration Background: Patient with suspected Ebola Virus symptoms admitted for evaluation. Isolation protocols were in effect. 63

  27. Case Study 2 : Hem oconcentration Background: Patient with suspected Ebola Virus symptoms admitted for evaluation. Isolation protocols were in effect. A handheld blood gas chemistry analyzer served as the primary chemistry analyzer. 0853 hrs – Specimens collected for chemistry and CBC testing. 64

  28. Case Study 2 : Hem oconcentration Background: Patient with suspected Ebola Virus symptoms admitted for evaluation. Isolation protocols were in effect. A handheld blood gas chemistry analyzer served as the primary chemistry analyzer. 0853 hrs – Specimens collected for chemistry and CBC testing. 65

  29. Case Study 2 : Hem oconcentration Background: Patient with suspected Ebola Virus symptoms admitted for evaluation. Isolation protocols were in effect. A handheld blood gas chemistry analyzer served as the primary chemistry analyzer. 0853 hrs – Specimens collected for chemistry and CBC testing. 66

  30. Case Study 2 : Hem oconcentration Background: Patient with suspected Ebola Virus symptoms admitted for evaluation. Isolation protocols were in effect. A handheld blood gas chemistry analyzer served as the primary chemistry analyzer. 0853 hrs – Specimens collected for chemistry and CBC testing. Handheld Results Hct = 68% Hb = 21.9 g/dL 67

  31. Case Study 2 : Hem oconcentration Background: Patient with suspected Ebola Virus symptoms admitted for evaluation. Isolation protocols were in effect. A handheld blood gas chemistry analyzer served as the primary chemistry analyzer. 0853 hrs – Specimens collected for chemistry and CBC testing. CBC Results Handheld Results Hct = 41% Hct = 68% Hb = 13.2 g/dL Hb = 21.9 g/dL 68

  32. Case Study 2 : Hem oconcentration Background: Patient with suspected Ebola Virus symptoms admitted for evaluation. Isolation protocols were in effect. A handheld blood gas chemistry analyzer served as the primary chemistry analyzer. 0853 hrs – Specimens collected for chemistry and CBC testing. CBC Results Handheld Results Hct = 41% Hct = 68% Hb = 13.2 g/dL Hb = 21.9 g/dL RE-MIXING! Hct = 43% Hb = 13.8 g/dL 69

  33. Case Study 2 : Hem oconcentration Background: Patient with suspected Ebola Virus symptoms admitted for evaluation. Isolation protocols were in effect. A handheld blood gas chemistry analyzer served as the primary chemistry analyzer. 0853 hrs – Specimens collected for chemistry and CBC testing. CBC Results Handheld Results Hct = 41% Hct = 68% Hb = 13.2 g/dL Hb = 21.9 g/dL RE-MIXING! Inadequate mixing may result in Hct = 43% artificial changes in total Hb = 13.8 g/dL hemoglobin measurements. 70

  34. Contem porary Hem oglobinom etric Techniques Conductance (Impendence) = Plasma Protein Electrode High Resistance • Plasma protein content contributes to hematocrit measurements for conductance-based systems. 71

  35. Contem porary Hem oglobinom etric Techniques Conductance (Impendence) = Plasma Protein Electrode Low Resistance from low plasma protein concentration! • Plasma protein content contributes to hematocrit measurements for conductance-based systems. • Conductance-based systems assumes a relatively fixed protein concentration. Therefore, during hemodilution, hematocrit may be falsely lower and causing an underestimation of total hemoglobin. 72

  36. Contem porary Hem oglobinom etric Techniques Conductance (Impendence) = Plasma Protein Electrode • Plasma protein content contributes to hematocrit measurements for conductance-based systems. • Conductance-based systems assumes a relatively fixed protein concentration. Therefore, during hemodilution, hematocrit may be falsely lower and causing an underestimation of total hemoglobin. UCDMC Study: Comparison of a handheld blood gas analyzer using conductance-based • measurement of hemoglobin versus a benchtop blood gas analyzer using a spectrophotometric- based method for hemoglobinometry . 73

  37. Clinical I m pact of Hem odilution for Point-of- Care Hem oglobin Measurem ents • Sixty patients requiring cardiac surgery were evaluated. • Paired specimens were tested using a handheld POC analyzer and spectrophotometric methods through the core laboratory. • Mean (SD) bias was -1.4 (1.1) g/dL, P = 0.011. • Based on core laboratory results 12 patients would have received unnecessary transfusions. 74

  38. Clinical I m pact of Hem odilution for Point-of- Care Hem oglobin Measurem ents • Sixty patients requiring cardiac surgery were evaluated. • Paired specimens were tested using a handheld POC analyzer and spectrophotometric methods through the core laboratory. • Mean (SD) bias was -1.4 (1.1) g/dL, P = 0.011. • Based on core laboratory results 12 patients would have received unnecessary transfusions. 75

  39. Clinical I m pact of Hem odilution for Point-of- Care Hem oglobin Measurem ents • Sixty patients requiring cardiac surgery were evaluated. • Paired specimens were tested using a = $219 handheld POC analyzer and spectrophotometric methods through the core laboratory. • Mean (SD) bias was -1.4 (1.1) g/dL, P = 0.011. • Based on core laboratory results 12 patients would have received unnecessary $219 x 12 = $2,628 transfusions. POTENTIALLY WASTED Toner RW, et al. Appl Health Econ Health Policy 2011;9:29-37 76

  40. Case Study 3 : Hem odilution 77

  41. Case Study 3 : Hem odilution Background: FDA MAUDE database reports a case (03P76-25) of a neonatal patient with discrepant point-of-care (POC) hemoglobin values compared to the laboratory. The POC device used a conductance-based method of hemoglobin measurement, while the laboratory used a spectrophotometric method. 78

  42. Case Study 3 : Hem odilution Background: FDA MAUDE database reports a case (03P76-25) of a neonatal patient with discrepant point-of-care (POC) hemoglobin values compared to the laboratory. The POC device used a conductance-based method of hemoglobin measurement, while the laboratory used a spectrophotometric method. • POC device reported a hematocrit of 22%. Physician administered 7 mL of blood based on the POC result. 79

  43. Case Study 3 : Hem odilution Background: FDA MAUDE database reports a case (03P76-25) of a neonatal patient with discrepant point-of-care (POC) hemoglobin values compared to the laboratory. The POC device used a conductance-based method of hemoglobin measurement, while the laboratory used a spectrophotometric method. • POC device reported a hematocrit of 22%. Physician administered 7 mL of blood based on the POC result. • Transfusion was stopped halfway after the laboratory reported a hematocrit of 40% and hemoglobin of 11.7 g/dL. 80

  44. Case Study 3 : Hem odilution Background: FDA MAUDE database reports a case (03P76-25) of a neonatal patient with discrepant point-of-care (POC) hemoglobin values compared to the laboratory. The POC device used a conductance-based method of hemoglobin measurement, while the laboratory used a spectrophotometric method. • POC device reported a hematocrit of 22%. Physician administered 7 mL of blood based on the POC result. • Transfusion was stopped halfway after the laboratory reported a hematocrit of 40% and hemoglobin of 11.7 g/dL. • Post-transfusion POC and lab hematocrit values were 45 and 50% respectively. 81

  45. Analytical Perform ance of Optical vs. Conductance-Based Hem oglobinom etry 11 Device #1 Hb (g/dL) 10 y = 0.5092x + 4.0176 R² = 0.5253 9 8 7 6 5 4 4 5 6 7 8 9 10 11 12 Central Laboratory Hb (g/dL) 82

  46. Analytical Perform ance of Optical vs. Conductance-Based Hem oglobinom etry 11 Device #2 Hb (g/dL) 10 y = 0.5249x + 3.9443 R² = 0.5407 9 8 7 6 5 4 4 5 6 7 8 9 10 11 12 Central Laboratory Hb (g/dL) 83

  47. Analytical Perform ance of Optical vs. Conductance-Based Hem oglobinom etry 11 y = 0.9345x + 0.4057 R² = 0.9205 10 Device #3 Hb (g/dL) 9 8 7 6 5 4 4 5 6 7 8 9 10 11 Central Laboratory Hb (g/dL) 84

  48. Analytical Perform ance of Optical vs. Conductance-Based Hem oglobinom etry Notes: Reference Method = Beckman LH hematology analyzer 85

  49. Analytical Perform ance of Optical vs. Conductance-Based Hem oglobinom etry Median (IQR) Bias: 0.78 (0.78) g/dL P < 0.001 N = 50 Notes: Reference Method = Beckman LH hematology analyzer 86

  50. Analytical Perform ance of Optical vs. Conductance-Based Hem oglobinom etry Median (IQR) Bias: 0.73 (0.60) g/dL P < 0.001 N = 50 Notes: Reference Method = Beckman LH hematology analyzer 87

  51. Analytical Perform ance of Optical vs. Conductance-Based Hem oglobinom etry Median (IQR) Bias: 0.22 (0.20) g/dL P = 0.510 N = 50 Notes: Reference Method = Beckman LH hematology analyzer 88

  52. Analytical Perform ance of Optical vs. Conductance-Based Hem oglobinom etry 10 Serial Testing Performance at 7 and 8 g/dL • Serial testing revealed significant analytical 9.5 bias between spectrophotometry vs. Total Hemoglobin (g/dL) 9 conductance-based measurements. 8.5 8 7.5 Spectrophotometric-based Methods *** 7 6.5 6 1 2 3 4 5 Time Point Notes: *** P<0.001, Central Lab = Spectrophotometric Method, n = 20 patients 89

  53. Analytical Perform ance of Optical vs. Conductance-Based Hem oglobinom etry 10 Serial Testing Performance at 7 and 8 g/dL • Serial testing revealed significant analytical 9.5 bias between spectrophotometry vs. Total Hemoglobin (g/dL) 9 conductance-based measurements. 8.5 8 7.5 Spectrophotometric-based Methods *** 7 Conductance-based Methods 6.5 6 1 2 3 4 5 Time Point Notes: *** P<0.001, Central Lab = Spectrophotometric Method, n = 20 patients 90

  54. Analytical Perform ance of Optical vs. Conductance-Based Hem oglobinom etry 10 Serial Testing Performance at 7 and 8 g/dL • Serial testing revealed significant analytical 9.5 bias between spectrophotometry vs. Total Hemoglobin (g/dL) 9 conductance-based measurements. 8.5 • Conductance-based devices would have prompted unnecessary transfusions at time 8 point #5 for patients using the 7 g/dL cutoff. 7.5 *** 7 Unnecessary Transfusion Risk 6.5 6 1 2 3 4 5 Time Point Notes: *** P<0.001, Central Lab = Spectrophotometric Method, n = 20 patients 91

  55. Analytical Perform ance of Optical vs. Conductance-Based Hem oglobinom etry 10 Serial Testing Performance at 7 and 8 g/dL • Serial testing revealed significant analytical 9.5 bias between spectrophotometry vs. Total Hemoglobin (g/dL) 9 conductance-based measurements. 8.5 • Conductance-based devices would have prompted unnecessary transfusions at time 8 point #5 for patients using the 7 g/dL cutoff. 7.5 *** • All serial conductance measurements were 7 at risk for potential transfusions if the 8 g/dL cutoff was used. Unnecessary Transfusion Risk 6.5 6 1 2 3 4 5 Time Point Notes: *** P<0.001, Central Lab = Spectrophotometric Method, n = 20 patients 92

  56. Analytical Perform ance of Optical vs. Conductance-Based Hem oglobinom etry 10 Serial Testing Performance at 7 and 8 g/dL Central Laboratory Method • Serial testing revealed significant analytical 9.5 bias between spectrophotometry vs. Total Hemoglobin (g/dL) 9 conductance-based measurements. 8.5 • Conductance-based devices would have prompted unnecessary transfusions at time 8 point #5 for patients using the 7 g/dL cutoff. 7.5 *** • All serial conductance measurements were 7 at risk for potential transfusions if the 8 g/dL cutoff was used. Unnecessary Transfusion Risk 6.5 6 1 2 3 4 5 Time Point Notes: *** P<0.001, Central Lab = Spectrophotometric Method, n = 20 patients 93

  57. Manufacturer and User Facility Device Experience ( MAUDE) Database Sum m ary Device 1 Device 2 Device 3 Timeframe 2011-2016 2011-2016 2014-2016* Erroneous Results 8 0 0 Improper 5 0 0 Transfusions https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfmaude/results.cfm, Accessed on July 19, 2016 94

  58. I NTERFERENCES I N W HOLE BLOOD ANALYSI S Air Contamination Delayed Testing Hemodilution/Hemoconcentration Hemolysis 95

  59. I NTERFERENCES I N W HOLE BLOOD ANALYSI S Air Contamination Delayed Testing Pseudohyperkalemia Hemodilution/Hemoconcentration Hemolysis 96

  60. I NTERFERENCES I N W HOLE BLOOD ANALYSI S Air Contamination Delayed Testing Pseudohyperkalemia Hemodilution/Hemoconcentration “Pseudonormokalemia” Hemolysis 97

  61. I NTERFERENCES I N W HOLE BLOOD ANALYSI S Air Contamination Delayed Testing No current FDA approved integrated solutions for detecting hemolysis at the point-of-care Pseudohyperkalemia Hemodilution/Hemoconcentration “Pseudonormokalemia” Hemolysis 98

  62. Biotin: The “Snake Oil” of 2018? 99

Download Presentation
Download Policy: The content available on the website is offered to you 'AS IS' for your personal information and use only. It cannot be commercialized, licensed, or distributed on other websites without prior consent from the author. To download a presentation, simply click this link. If you encounter any difficulties during the download process, it's possible that the publisher has removed the file from their server.

Recommend


More recommend